An RF dosimeter for independent SAR measurement in MRI scanners. Issue 12 (15th November 2013)
- Record Type:
- Journal Article
- Title:
- An RF dosimeter for independent SAR measurement in MRI scanners. Issue 12 (15th November 2013)
- Main Title:
- An RF dosimeter for independent SAR measurement in MRI scanners
- Authors:
- Qian, Di
El‐Sharkawy, AbdEl‐Monem M.
Bottomley, Paul A.
Edelstein, William A. - Abstract:
- Abstract : Purpose: : The monitoring and management of radio frequency (RF) exposure is critical for ensuring magnetic resonance imaging (MRI) safety. Commercial MRI scanners can overestimate specific absorption rates (SAR) and improperly restrict clinical MRI scans or the application of new MRI sequences, while underestimation of SAR can lead to tissue heating and thermal injury. Accurate scanner‐independent RF dosimetry is essential for measuring actual exposure when SAR is critical for ensuring regulatory compliance and MRI safety, for establishing RF exposure while evaluating interventional leads and devices, and for routine MRI quality assessment by medical physicists. However, at present there are no scanner‐independent SAR dosimeters. Methods: : An SAR dosimeter with an RF transducer comprises two orthogonal, rectangular copper loops and a spherical MRI phantom. The transducer is placed in the magnet bore and calibrated to approximate the resistive loading of the scannerˈs whole‐body birdcage RF coil for human subjects in Philips, GE and Siemens 3 tesla (3T) MRI scanners. The transducer loop reactances are adjusted to minimize interference with the transmit RF field (B1 ) at the MRI frequency. Power from the RF transducer is sampled with a high dynamic range power monitor and recorded on a computer. The deposited power is calibrated and tested on eight different MRI scanners. Whole‐body absorbed power vs weight and body mass index (BMI) is measured directly on 26Abstract : Purpose: : The monitoring and management of radio frequency (RF) exposure is critical for ensuring magnetic resonance imaging (MRI) safety. Commercial MRI scanners can overestimate specific absorption rates (SAR) and improperly restrict clinical MRI scans or the application of new MRI sequences, while underestimation of SAR can lead to tissue heating and thermal injury. Accurate scanner‐independent RF dosimetry is essential for measuring actual exposure when SAR is critical for ensuring regulatory compliance and MRI safety, for establishing RF exposure while evaluating interventional leads and devices, and for routine MRI quality assessment by medical physicists. However, at present there are no scanner‐independent SAR dosimeters. Methods: : An SAR dosimeter with an RF transducer comprises two orthogonal, rectangular copper loops and a spherical MRI phantom. The transducer is placed in the magnet bore and calibrated to approximate the resistive loading of the scannerˈs whole‐body birdcage RF coil for human subjects in Philips, GE and Siemens 3 tesla (3T) MRI scanners. The transducer loop reactances are adjusted to minimize interference with the transmit RF field (B1 ) at the MRI frequency. Power from the RF transducer is sampled with a high dynamic range power monitor and recorded on a computer. The deposited power is calibrated and tested on eight different MRI scanners. Whole‐body absorbed power vs weight and body mass index (BMI) is measured directly on 26 subjects. Results: : A single linear calibration curve sufficed for RF dosimetry at 127.8 MHz on three different Philips and three GE 3T MRI scanners. An RF dosimeter operating at 123.2 MHz on two Siemens 3T scanners required a separate transducer and a slightly different calibration curve. Measurement accuracy was ∼3%. With the torso landmarked at the xiphoid, human adult whole‑body absorbed power varied approximately linearly with patient weight and BMI. This indicates that whole‐body torso SAR is on average independent of the imaging subject, albeit with fluctuations. Conclusions: : Our 3T RF dosimeter and transducers accurately measure RF exposure in body‐equivalent loads and provide scanner‐independent assessments of whole‐body RF power deposition for establishing safety compliance useful for MRI sequence and device testing. … (more)
- Is Part Of:
- Medical physics. Volume 40:Issue 12(2013)
- Journal:
- Medical physics
- Issue:
- Volume 40:Issue 12(2013)
- Issue Display:
- Volume 40, Issue 12 (2013)
- Year:
- 2013
- Volume:
- 40
- Issue:
- 12
- Issue Sort Value:
- 2013-0040-0012-0000
- Page Start:
- n/a
- Page End:
- n/a
- Publication Date:
- 2013-11-15
- Subjects:
- Instrumentation -- Clinical applications -- Biomedical instrumentation and transducers, including micro‐electro‐mechanical systems (MEMS) -- Dosimetry/exposure assessment -- Dosimetry/exposure assessment
biological effects of fields -- biomedical MRI -- biomedical transducers -- coils -- dosimetry -- phantoms -- safety
RF dosimetry -- MRI safety -- deposited power -- SAR -- RF exposure
Involving electronic [emr] or nuclear [nmr] magnetic resonance, e.g. magnetic resonance imaging -- Biological material, e.g. blood, urine; Haemocytometers -- Coils -- Coils; Windings; Conductive connections -- Scintigraphy
Magnetic resonance imaging -- Image scanners -- Transducers -- Calibration -- Radiofrequency power transmission -- Germanium -- Coils -- Medical radiation safety -- Medical imaging -- Dosimetry
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1118/1.4829527 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9334.xml