Independent dosimetric assessment of the model EP917 episcleral brachytherapy plaque. Issue 9 (13th August 2014)
- Record Type:
- Journal Article
- Title:
- Independent dosimetric assessment of the model EP917 episcleral brachytherapy plaque. Issue 9 (13th August 2014)
- Main Title:
- Independent dosimetric assessment of the model EP917 episcleral brachytherapy plaque
- Authors:
- Aryal, Prakash
Molloy, Janelle A.
Rivard, Mark J. - Abstract:
- Abstract : Purpose: : To investigate the influence of slot design on dose distributions and dose‐volume histograms (DVHs) for the model EP917 plaque for episcleral brachytherapy. Methods: : Dimensions and orientations of the slots were measured for three model EP917 plaques and compared to data in the Plaque Simulator (PS) treatment planning software (version 5.7.6). These independently determined coordinates were incorporated into the MCNP Monte Carlo simulation environment to obtain dose from the plaques in a water environment and in a clinical environment with ocular structures. A tumor volume was simulated as 5 mm in apical height and 11 mm in basal diameter. Variations in plaque mass density and composition; slot length, width, and depth; seed positioning; and Ag‐marker rod positioning were simulated to examine their influence on plaque central axis (CAX) and planar dose distributions, and DVHs. Results: : Seed shifts in a single slot toward the eye and shifts of the 125 I‐coated Ag rod within the capsule had the greatest impact on CAX dose distribution. A shift of 0.0994 mm toward the eye increased dose by 14%, 9%, 4.3%, and 2.7% at 1, 2, 5, and 10 mm, respectively, from the inner sclera. When examining the fully‐modeled plaque in the ocular geometry, the largest dose variations were caused by shifting the Ag rods toward the sclera and shifting the seeds away from the globe when the slots were made 0.51 mm deeper, causing +34.3% and −69.4% dose changes to the outerAbstract : Purpose: : To investigate the influence of slot design on dose distributions and dose‐volume histograms (DVHs) for the model EP917 plaque for episcleral brachytherapy. Methods: : Dimensions and orientations of the slots were measured for three model EP917 plaques and compared to data in the Plaque Simulator (PS) treatment planning software (version 5.7.6). These independently determined coordinates were incorporated into the MCNP Monte Carlo simulation environment to obtain dose from the plaques in a water environment and in a clinical environment with ocular structures. A tumor volume was simulated as 5 mm in apical height and 11 mm in basal diameter. Variations in plaque mass density and composition; slot length, width, and depth; seed positioning; and Ag‐marker rod positioning were simulated to examine their influence on plaque central axis (CAX) and planar dose distributions, and DVHs. Results: : Seed shifts in a single slot toward the eye and shifts of the 125 I‐coated Ag rod within the capsule had the greatest impact on CAX dose distribution. A shift of 0.0994 mm toward the eye increased dose by 14%, 9%, 4.3%, and 2.7% at 1, 2, 5, and 10 mm, respectively, from the inner sclera. When examining the fully‐modeled plaque in the ocular geometry, the largest dose variations were caused by shifting the Ag rods toward the sclera and shifting the seeds away from the globe when the slots were made 0.51 mm deeper, causing +34.3% and −69.4% dose changes to the outer sclera, respectively. At points along the CAX, dose from the full plaque geometry using the measured slot design was 2.4% ± 1.1% higher than the manufacturer‐provided slot design and 2.2% ± 2.3% higher than the homogeneous calculation of PS treatment planning results. The ratio of D 10 values for the measured slot design to the D 10 values for the manufacturer‐provided slot design was higher by 9%, 10%, and 19% for the tumor, inner sclera, and outer sclera, respectively. In comparison to the measured slot design, a theoretical plaque having narrower and deeper slots delivered 30%, 37%, and 62% lower D 10 doses to the tumor, inner sclera, and outer sclera, respectively. Conclusions: : While the measured positions of the slots on the model EP917 plaque were in close agreement (<0.7 mm) with the PS values, small differences in the slot shape caused substantial differences in dose distributions and DVH metrics. Increasing slot depth by 0.1 mm decreased outer scleral dose by 20%, yet shifting the Ag rods in the seeds toward the globe by 0.1 mm increased outer scleral dose by 35%. The clinical medical physicist is advised to measure these types of plaques upon acceptance testing before clinical use to inspect slot shape and position for comparison with data used for treatment planning purposes. … (more)
- Is Part Of:
- Medical physics. Volume 41:Issue 9(2014)
- Journal:
- Medical physics
- Issue:
- Volume 41:Issue 9(2014)
- Issue Display:
- Volume 41, Issue 9 (2014)
- Year:
- 2014
- Volume:
- 41
- Issue:
- 9
- Issue Sort Value:
- 2014-0041-0009-0000
- Page Start:
- n/a
- Page End:
- n/a
- Publication Date:
- 2014-08-13
- Subjects:
- Dosimetry/exposure assessment -- Therapeutic applications, including brachytherapy -- Monte Carlo methods
biomedical equipment -- brachytherapy -- cancer -- dosimetry -- eye -- Monte Carlo methods
eye plaque brachytherapy -- Monte Carlo -- dosimetry -- 125I
Radiation therapy -- Scintigraphy
Dosimetry -- Cancer -- Brachytherapy -- Monte Carlo methods -- Photons -- Gold -- Medical treatment planning -- Computer software -- Collimation -- Medical physicists
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1118/1.4892603 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9333.xml