Computer aided detection of surgical retained foreign object for prevention. Issue 3 (18th February 2015)
- Record Type:
- Journal Article
- Title:
- Computer aided detection of surgical retained foreign object for prevention. Issue 3 (18th February 2015)
- Main Title:
- Computer aided detection of surgical retained foreign object for prevention
- Authors:
- Hadjiiski, Lubomir
Marentis, Theodore C.
Chaudhury, Amrita R.
Rondon, Lucas
Chronis, Nikolaos
Chan, Heang‐Ping - Abstract:
- Abstract : Purpose: Surgical retained foreign objects (RFOs) have significant morbidity and mortality. They are associated with approximately $1.5 × 10 9 annually in preventable medical costs. The detection accuracy of radiographs for RFOs is a mediocre 59%. The authors address the RFO problem with two complementary technologies: a three‐dimensional (3D) gossypiboma micro tag, the μ Tag that improves the visibility of RFOs on radiographs, and a computer aided detection (CAD) system that detects the μ Tag. It is desirable for the CAD system to operate in a high specificity mode in the operating room (OR) and function as a first reader for the surgeon. This allows for fast point of care results and seamless workflow integration. The CAD system can also operate in a high sensitivity mode as a second reader for the radiologist to ensure the highest possible detection accuracy. Methods: The 3D geometry of the μ Tag produces a similar two dimensional (2D) depiction on radiographs regardless of its orientation in the human body and ensures accurate detection by a radiologist and the CAD. The authors created a data set of 1800 cadaver images with the 3D μ Tag and other common man‐made surgical objects positioned randomly. A total of 1061 cadaver images contained a single μ Tag and the remaining 739 were without μ Tag. A radiologist marked the location of the μ Tag using an in‐house developed graphical user interface. The data set was partitioned into three independent subsets: aAbstract : Purpose: Surgical retained foreign objects (RFOs) have significant morbidity and mortality. They are associated with approximately $1.5 × 10 9 annually in preventable medical costs. The detection accuracy of radiographs for RFOs is a mediocre 59%. The authors address the RFO problem with two complementary technologies: a three‐dimensional (3D) gossypiboma micro tag, the μ Tag that improves the visibility of RFOs on radiographs, and a computer aided detection (CAD) system that detects the μ Tag. It is desirable for the CAD system to operate in a high specificity mode in the operating room (OR) and function as a first reader for the surgeon. This allows for fast point of care results and seamless workflow integration. The CAD system can also operate in a high sensitivity mode as a second reader for the radiologist to ensure the highest possible detection accuracy. Methods: The 3D geometry of the μ Tag produces a similar two dimensional (2D) depiction on radiographs regardless of its orientation in the human body and ensures accurate detection by a radiologist and the CAD. The authors created a data set of 1800 cadaver images with the 3D μ Tag and other common man‐made surgical objects positioned randomly. A total of 1061 cadaver images contained a single μ Tag and the remaining 739 were without μ Tag. A radiologist marked the location of the μ Tag using an in‐house developed graphical user interface. The data set was partitioned into three independent subsets: a training set, a validation set, and a test set, consisting of 540, 560, and 700 images, respectively. A CAD system with modules that included preprocessing μ Tag enhancement, labeling, segmentation, feature analysis, classification, and detection was developed. The CAD system was developed using the training and the validation sets. Results: On the training set, the CAD achieved 81.5% sensitivity with 0.014 false positives (FPs) per image in a high specificity mode for the surgeons in the OR and 96.1% sensitivity with 0.81 FPs per image in a high sensitivity mode for the radiologists. On the independent test set, the CAD achieved 79.5% sensitivity with 0.003 FPs per image in a high specificity mode for the surgeons and 90.2% sensitivity with 0.23 FPs per image in a high sensitivity mode for the radiologists. Conclusions: To the best of the authors' knowledge, this is the first time a 3D μ Tag is used to produce a recognizable, substantially similar 2D projection on radiographs regardless of orientation in space. It is the first time a CAD system is used to search for man‐made objects over anatomic background. The CAD system for the μ Tags achieved reasonable performance in both the high specificity and the high sensitivity modes. … (more)
- Is Part Of:
- Medical physics. Volume 42:Issue 3(2015)
- Journal:
- Medical physics
- Issue:
- Volume 42:Issue 3(2015)
- Issue Display:
- Volume 42, Issue 3 (2015)
- Year:
- 2015
- Volume:
- 42
- Issue:
- 3
- Issue Sort Value:
- 2015-0042-0003-0000
- Page Start:
- 1213
- Page End:
- 1222
- Publication Date:
- 2015-02-18
- Subjects:
- diagnostic radiography -- medical image processing -- object detection
Radiography
Biological material, e.g. blood, urine; Haemocytometers -- Digital computing or data processing equipment or methods, specially adapted for specific applications -- Image data processing or generation, in general
computer aided detection (CAD) -- surgical retained foreign objects -- gossypiboma -- radiograph
Radiography -- Radiologists -- Image detection systems -- Artificial neural networks -- Three dimensional sensing -- Three dimensional image processing -- X‐ray detectors -- Image sensors -- Visibility
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1118/1.4907964 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9347.xml