Technical Note: Display window setting: An important factor for detecting subtle but clinically relevant artifacts in daily CT quality control. Issue 12 (10th November 2016)
- Record Type:
- Journal Article
- Title:
- Technical Note: Display window setting: An important factor for detecting subtle but clinically relevant artifacts in daily CT quality control. Issue 12 (10th November 2016)
- Main Title:
- Technical Note: Display window setting: An important factor for detecting subtle but clinically relevant artifacts in daily CT quality control
- Authors:
- Long, Zaiyang
Bruesewitz, Michael R.
Sheedy, Emily N.
Powell, Michele A.
Kramer, Jacqualynn C.
Supalla, Randall R.
Colvin, Chance M.
Bechel, Jessica R.
Favazza, Christopher P.
Kofler, James M.
Leng, Shuai
McCollough, Cynthia H.
Yu, Lifeng - Abstract:
- Abstract : Purpose: This study aimed to investigate the influence of display window setting on technologist performance detecting subtle but clinically relevant artifacts in daily computed tomography (CT) quality control (dQC) images. Methods: Fifty three sets of dQC images were retrospectively selected, including 30 sets without artifacts, and 23 with subtle but clinically relevant artifacts. They were randomized and shown to six CT technologists (two new and four experienced). Each technologist reviewed all images in each of two sessions, one with a display window width (WW) of 100 HU, which is currently recommended by the American College of Radiology, and the other with a narrow WW of 40 HU, both at a window level of 0 HU. For each case, technologists rated the presence of image artifacts based on a five point scale. The area under the receiver operating characteristic curve (AUC) was used to evaluate the artifact detection performance. Results: At a WW of 100 HU, the AUC (95% confidence interval) was 0.658 (0.576, 0.740), 0.532 (0.429, 0.635), and 0.616 (0.543, 0.619) for the experienced, new, and all technologists, respectively. At a WW of 40 HU, the AUC was 0.768 (0.687, 0.850), 0.546 (0.433, 0.658), and 0.694 (0.619, 0.769), respectively. The performance significantly improved at WW of 40 HU for experienced technologists ( p = 0.009) and for all technologists ( p = 0.040). Conclusions: Use of a narrow display WW significantly improved technologists' performance inAbstract : Purpose: This study aimed to investigate the influence of display window setting on technologist performance detecting subtle but clinically relevant artifacts in daily computed tomography (CT) quality control (dQC) images. Methods: Fifty three sets of dQC images were retrospectively selected, including 30 sets without artifacts, and 23 with subtle but clinically relevant artifacts. They were randomized and shown to six CT technologists (two new and four experienced). Each technologist reviewed all images in each of two sessions, one with a display window width (WW) of 100 HU, which is currently recommended by the American College of Radiology, and the other with a narrow WW of 40 HU, both at a window level of 0 HU. For each case, technologists rated the presence of image artifacts based on a five point scale. The area under the receiver operating characteristic curve (AUC) was used to evaluate the artifact detection performance. Results: At a WW of 100 HU, the AUC (95% confidence interval) was 0.658 (0.576, 0.740), 0.532 (0.429, 0.635), and 0.616 (0.543, 0.619) for the experienced, new, and all technologists, respectively. At a WW of 40 HU, the AUC was 0.768 (0.687, 0.850), 0.546 (0.433, 0.658), and 0.694 (0.619, 0.769), respectively. The performance significantly improved at WW of 40 HU for experienced technologists ( p = 0.009) and for all technologists ( p = 0.040). Conclusions: Use of a narrow display WW significantly improved technologists' performance in dQC for detecting subtle but clinically relevant artifacts as compared to that using a 100 HU display WW. … (more)
- Is Part Of:
- Medical physics. Volume 43:Issue 12(2016)
- Journal:
- Medical physics
- Issue:
- Volume 43:Issue 12(2016)
- Issue Display:
- Volume 43, Issue 12 (2016)
- Year:
- 2016
- Volume:
- 43
- Issue:
- 12
- Issue Sort Value:
- 2016-0043-0012-0000
- Page Start:
- 6413
- Page End:
- 6417
- Publication Date:
- 2016-11-10
- Subjects:
- computerised tomography -- medical image processing -- quality control -- sensitivity analysis
Computed tomography
Computerised tomographs -- Biological material, e.g. blood, urine; Haemocytometers -- Digital computing or data processing equipment or methods, specially adapted for specific applications -- Image data processing or generation, in general
CT -- quality control -- artifacts
Computed tomography -- Medical image artifacts -- Image scanners -- Quality assurance -- Image detection systems -- Image sensors -- Pathology
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1118/1.4966698 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9344.xml