Design and experimental testing of air slab caps which convert commercial electron diodes into dual purpose, correction‐free diodes for small field dosimetry. Issue 10 (12th September 2014)
- Record Type:
- Journal Article
- Title:
- Design and experimental testing of air slab caps which convert commercial electron diodes into dual purpose, correction‐free diodes for small field dosimetry. Issue 10 (12th September 2014)
- Main Title:
- Design and experimental testing of air slab caps which convert commercial electron diodes into dual purpose, correction‐free diodes for small field dosimetry
- Authors:
- Charles, P. H.
Cranmer‐Sargison, G.
Thwaites, D. I.
Kairn, T.
Crowe, S. B.
Pedrazzini, G.
Aland, T.
Kenny, J.
Langton, C. M.
Trapp, J. V. - Abstract:
- Abstract : Purpose: : Two diodes which do not require correction factors for small field relative output measurements are designed and validated using experimental methodology. This was achieved by adding an air layer above the active volume of the diode detectors, which canceled out the increase in response of the diodes in small fields relative to standard field sizes. Methods: : Due to the increased density of silicon and other components within a diode, additional electrons are created. In very small fields, a very small air gap acts as an effective filter of electrons with a high angle of incidence. The aim was to design a diode that balanced these perturbations to give a response similar to a water‐only geometry. Three thicknesses of air were placed at the proximal end of a PTW 60017 electron diode (PTWe) using an adjustable "air cap". A set of output ratios ( OR Det f clin ) for square field sizes of side length down to 5 mm was measured using each air thickness and compared to OR Det f clin measured using an IBA stereotactic field diode (SFD). k Q clin, Q msr f clin, f msr was transferred from the SFD to the PTWe diode and plotted as a function of air gap thickness for each field size. This enabled the optimal air gap thickness to be obtained by observing which thickness of air was required such that k Q clin, Q msr f clin, f msr was equal to 1.00 at all field sizes. A similar procedure was used to find the optimal air thickness required to make a modified SunAbstract : Purpose: : Two diodes which do not require correction factors for small field relative output measurements are designed and validated using experimental methodology. This was achieved by adding an air layer above the active volume of the diode detectors, which canceled out the increase in response of the diodes in small fields relative to standard field sizes. Methods: : Due to the increased density of silicon and other components within a diode, additional electrons are created. In very small fields, a very small air gap acts as an effective filter of electrons with a high angle of incidence. The aim was to design a diode that balanced these perturbations to give a response similar to a water‐only geometry. Three thicknesses of air were placed at the proximal end of a PTW 60017 electron diode (PTWe) using an adjustable "air cap". A set of output ratios ( OR Det f clin ) for square field sizes of side length down to 5 mm was measured using each air thickness and compared to OR Det f clin measured using an IBA stereotactic field diode (SFD). k Q clin, Q msr f clin, f msr was transferred from the SFD to the PTWe diode and plotted as a function of air gap thickness for each field size. This enabled the optimal air gap thickness to be obtained by observing which thickness of air was required such that k Q clin, Q msr f clin, f msr was equal to 1.00 at all field sizes. A similar procedure was used to find the optimal air thickness required to make a modified Sun Nuclear EDGE detector (EDGEe) which is "correction‐free" in small field relative dosimetry. In addition, the feasibility of experimentally transferring k Q clin, Q msr f clin, f msr values from the SFD to unknown diodes was tested by comparing the experimentally transferred k Q clin, Q msr f clin, f msr values for unmodified PTWe and EDGEe diodes to Monte Carlo simulated values. Results: : 1.0 mm of air was required to make the PTWe diode correction‐free. This modified diode (PTWeair ) produced output factors equivalent to those in water at all field sizes (5–50 mm). The optimal air thickness required for the EDGEe diode was found to be 0.6 mm. The modified diode (EDGEeair ) produced output factors equivalent to those in water, except at field sizes of 8 and 10 mm where it measured approximately 2% greater than the relative dose to water. The experimentally calculated k Q clin, Q msr f clin, f msr for both the PTWe and the EDGEe diodes (without air) matched Monte Carlo simulated results, thus proving that it is feasible to transfer k Q clin, Q msr f clin, f msr from one commercially available detector to another using experimental methods and the recommended experimental setup. Conclusions: : It is possible to create a diode which does not require corrections for small field output factor measurements. This has been performed and verified experimentally. The ability of a detector to be "correction‐free" depends strongly on its design and composition. A nonwater‐equivalent detector can only be "correction‐free" if competing perturbations of the beam cancel out at all field sizes. This should not be confused with true water equivalency of a detector. … (more)
- Is Part Of:
- Medical physics. Volume 41:Issue 10(2014)
- Journal:
- Medical physics
- Issue:
- Volume 41:Issue 10(2014)
- Issue Display:
- Volume 41, Issue 10 (2014)
- Year:
- 2014
- Volume:
- 41
- Issue:
- 10
- Issue Sort Value:
- 2014-0041-0010-0000
- Page Start:
- n/a
- Page End:
- n/a
- Publication Date:
- 2014-09-12
- Subjects:
- dosimetry -- elemental semiconductors -- Monte Carlo methods -- perturbation theory -- semiconductor diodes -- silicon
Dosimetry/exposure assessment -- Junction diodes
Scintigraphy -- Diodes
small field -- dosimetry -- diode -- correction‐free -- air -- output factors -- nonequilibrium
Field size -- Monte Carlo methods -- Silicon -- Water vapor -- Dosimetry -- Silicon detectors -- Particle beam detectors -- Error analysis -- Alloys -- Linear accelerators
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1118/1.4894728 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9350.xml