A comprehensive investigation of the accuracy and reproducibility of a multitarget single isocenter VMAT radiosurgery technique. Issue 12 (26th November 2013)
- Record Type:
- Journal Article
- Title:
- A comprehensive investigation of the accuracy and reproducibility of a multitarget single isocenter VMAT radiosurgery technique. Issue 12 (26th November 2013)
- Main Title:
- A comprehensive investigation of the accuracy and reproducibility of a multitarget single isocenter VMAT radiosurgery technique
- Authors:
- Thomas, Andrew
Niebanck, Michael
Juang, Titania
Wang, Zhiheng
Oldham, Mark - Abstract:
- Abstract : Purpose: : Recent trends in stereotactic radiosurgery use multifocal volumetric modulated arc therapy (VMAT) plans to simultaneously treat several distinct targets. Conventional verification often involves low resolution measurements in a single plane, a cylinder, or intersecting planes of diodes or ion chambers. This work presents an investigation into the consistency and reproducibility of this new treatment technique using a comprehensive commissioned high‐resolution 3D dosimetry system (PRESAGE ® /Optical‐CT). Methods: : A complex VMAT plan consisting of a single isocenter but five separate targets was created in Eclipse for a head phantom containing a cylindrical PRESAGE ® dosimetry insert of 11 cm diameter and height. The plan contained five VMAT arcs delivering target doses from 12 to 20 Gy. The treatment was delivered to four dosimeters positioned in the head phantom and repeated four times, yielding four separate 3D dosimetry verifications. Each delivery was completely independent and was given after image guided radiation therapy (IGRT) positioning using Brainlab ExacTrac and cone beam computed tomography. A final delivery was given to a modified insert containing a pin‐point ion chamber enabling calibration of PRESAGE ® 3D data to dose. Dosimetric data were read out in an optical‐CT scanner. Consistency and reproducibility of the treatment technique (including IGRT setup) was investigated by comparing the dose distributions in the four inserts, and withAbstract : Purpose: : Recent trends in stereotactic radiosurgery use multifocal volumetric modulated arc therapy (VMAT) plans to simultaneously treat several distinct targets. Conventional verification often involves low resolution measurements in a single plane, a cylinder, or intersecting planes of diodes or ion chambers. This work presents an investigation into the consistency and reproducibility of this new treatment technique using a comprehensive commissioned high‐resolution 3D dosimetry system (PRESAGE ® /Optical‐CT). Methods: : A complex VMAT plan consisting of a single isocenter but five separate targets was created in Eclipse for a head phantom containing a cylindrical PRESAGE ® dosimetry insert of 11 cm diameter and height. The plan contained five VMAT arcs delivering target doses from 12 to 20 Gy. The treatment was delivered to four dosimeters positioned in the head phantom and repeated four times, yielding four separate 3D dosimetry verifications. Each delivery was completely independent and was given after image guided radiation therapy (IGRT) positioning using Brainlab ExacTrac and cone beam computed tomography. A final delivery was given to a modified insert containing a pin‐point ion chamber enabling calibration of PRESAGE ® 3D data to dose. Dosimetric data were read out in an optical‐CT scanner. Consistency and reproducibility of the treatment technique (including IGRT setup) was investigated by comparing the dose distributions in the four inserts, and with the predicted treatment planning system distribution. Results: : Dose distributions from the four dosimeters were registered and analyzed to determine the mean and standard deviation at all points throughout the dosimeters. A dose standard deviation of <3% was found from dosimeter to dosimeter. Global 3D gamma maps show that the predicted and measured dose matched well [3D gamma passing rate was 98.0% (3%, 2 mm)]. Conclusions: : The deliveries of the irradiation were found to be consistent and matched the treatment plan, demonstrating high accuracy and reproducibility of both the treatment machine and the IGRT procedure. The complexity of the treatment (multiple arcs) and dosimetry (multiple strong gradients) pose a substantial challenge for comprehensive verification. 3D dosimetry can be uniquely effective in this scenario. … (more)
- Is Part Of:
- Medical physics. Volume 40:Issue 12(2013)
- Journal:
- Medical physics
- Issue:
- Volume 40:Issue 12(2013)
- Issue Display:
- Volume 40, Issue 12 (2013)
- Year:
- 2013
- Volume:
- 40
- Issue:
- 12
- Issue Sort Value:
- 2013-0040-0012-0000
- Page Start:
- n/a
- Page End:
- n/a
- Publication Date:
- 2013-11-26
- Subjects:
- Stereotactic radiosurgery -- Conformal radiation treatment -- Dose‐volume analysis -- Optimization -- Dosimetry/exposure assessment
computerised tomography -- dosimeters -- dosimetry -- ionisation chambers -- phantoms -- radiation therapy -- surgery
3D dosimetry -- VMAT -- radiosurgery -- quality assurance -- PRESAGE®
Computerised tomographs -- Surgical instruments, devices or methods, e.g. tourniquets -- Radiation therapy -- Biological material, e.g. blood, urine; Haemocytometers -- Measurement of nuclear or x‐radiation -- Dosimeters -- Tubes for determining the presence, intensity, density or energy of radiation or particles -- Scintigraphy -- with scintillation detectors -- Ionisation chambers
Dosimetry -- Medical imaging -- Radiosurgery -- Radiation treatment -- Image guided radiation therapy -- Ionization chambers -- Cone beam computed tomography -- Image scanners -- Drug delivery
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1118/1.4829518 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9334.xml