Asymmetric dose–volume optimization with smoothness control for rotating‐shield brachytherapy. Issue 11 (17th October 2014)
- Record Type:
- Journal Article
- Title:
- Asymmetric dose–volume optimization with smoothness control for rotating‐shield brachytherapy. Issue 11 (17th October 2014)
- Main Title:
- Asymmetric dose–volume optimization with smoothness control for rotating‐shield brachytherapy
- Authors:
- Liu, Yunlong
Flynn, Ryan T.
Kim, Yusung
Wu, Xiaodong - Abstract:
- Abstract : Purpose: It is important to reduce fluence map complexity in rotating‐shield brachytherapy (RSBT) inverse planning to improve delivery efficiency while maintaining plan quality. This study proposes an efficient and effective RSBT dose optimization method which enables to produce smooth fluence maps. Methods: Five cervical cancer patients each with a high‐risk clinical‐target‐volume (HR‐CTV) larger than 40 cm 3 were considered as the test cases. The RSBT source was a partially shielded electronic brachytherapy source (Xoft Axxent™). The anchor RSBT plans generated by the asymmetric dose–volume optimization with smoothness control (ADOS) method were compared against those produced by the dose–surface optimization (DSO) method and inverse‐planning with simulated annealing (IPSA). Either L 1 ‐norm or L 2 ‐norm was used to measure the smoothness of a fluence map in the proposed ADOS method as one weighted term of the objective function. Uniform dwell‐time scaling was applied to all plans such that HR‐CTV D 90 was maximized without violating the D 2cc tolerances of the rectum, bladder, and sigmoid colon. The quality of the anchor plans was measured with HR‐CTV D 90 of the anchor plans. Single‐shielded RSBT [(S‐RSBT), RSBT with single, fix sized delivery window] and dynamic‐sheilded RSBT [(D‐RSBT), RSBT with dynamically varying sized delivery window] delivery plans generated based on the anchor plans were also measured, with delivery time constraints of 10, 20, and 30Abstract : Purpose: It is important to reduce fluence map complexity in rotating‐shield brachytherapy (RSBT) inverse planning to improve delivery efficiency while maintaining plan quality. This study proposes an efficient and effective RSBT dose optimization method which enables to produce smooth fluence maps. Methods: Five cervical cancer patients each with a high‐risk clinical‐target‐volume (HR‐CTV) larger than 40 cm 3 were considered as the test cases. The RSBT source was a partially shielded electronic brachytherapy source (Xoft Axxent™). The anchor RSBT plans generated by the asymmetric dose–volume optimization with smoothness control (ADOS) method were compared against those produced by the dose–surface optimization (DSO) method and inverse‐planning with simulated annealing (IPSA). Either L 1 ‐norm or L 2 ‐norm was used to measure the smoothness of a fluence map in the proposed ADOS method as one weighted term of the objective function. Uniform dwell‐time scaling was applied to all plans such that HR‐CTV D 90 was maximized without violating the D 2cc tolerances of the rectum, bladder, and sigmoid colon. The quality of the anchor plans was measured with HR‐CTV D 90 of the anchor plans. Single‐shielded RSBT [(S‐RSBT), RSBT with single, fix sized delivery window] and dynamic‐sheilded RSBT [(D‐RSBT), RSBT with dynamically varying sized delivery window] delivery plans generated based on the anchor plans were also measured, with delivery time constraints of 10, 20, and 30 min/fraction (fx). Results: The average HR‐CTV D 90 values of the anchor plans achieved by the ADOS, DSO, and IPSA methods were 111.5, 94.2, and 107.4 Gy, respectively, where the weighting parameter β used in ADOS with L 2 ‐norm was set to be 100. By using S‐RSBT sequencing and 20 min/fx delivery time, the corresponding D 90 values were 88.8, 81.9, and 83.4 Gy; while using D‐RSBT sequencing with 20 min/fx delivery time, the corresponding D 90 values were 91.4, 88.3, and 78.9 Gy, respectively. The average optimization times for ADOS, DSO, and IPSA were, respectively, 77, 4, and 1800 s. By using L 1 ‐norm instead of L 2 ‐norm in the ADOS method, the optimization time was increased by 20 s, while the D 90 was reduced by 6.8 Gy on average. ADOS‐L1 was found to be more sensitive to the weighting parameter than ADOS‐L2. If β was increased to 10 000, the D 90 drop with ADOS‐L1 was 38 Gy, while the drop with ADOS‐L2 is 13 Gy. Conclusions: The ADOS method had a reasonable optimization time cost, while achieving comparable RSBT dose plans as the IPSA method, which is of much higher time complexity. Compared to the DSO and IPSA methods, ADOS is able to generate anchor plans which are more suitable for RSBT delivery while preserving the high quality of the original plans. Compared to ADOS‐L1, ADOS‐L2 is able to achieve better quality of anchor plans more efficiently. … (more)
- Is Part Of:
- Medical physics. Volume 41:Issue 11(2014)
- Journal:
- Medical physics
- Issue:
- Volume 41:Issue 11(2014)
- Issue Display:
- Volume 41, Issue 11 (2014)
- Year:
- 2014
- Volume:
- 41
- Issue:
- 11
- Issue Sort Value:
- 2014-0041-0011-0000
- Page Start:
- n/a
- Page End:
- n/a
- Publication Date:
- 2014-10-17
- Subjects:
- brachytherapy -- cancer -- dosimetry -- inverse problems -- medical control systems -- simulated annealing
Dose‐volume analysis -- Optimization -- Treatment planning -- Therapeutic applications, including brachytherapy -- Dosimetry/exposure assessment
Radiation therapy -- Scintigraphy
brachytherapy -- inverse planning -- fluence map smoothness -- optimization
Dosimetry -- Optimization -- Sequence analysis -- Brachytherapy -- Cancer -- Radiotherapy sources -- Radioactive sources -- Magnetic resonance imaging -- Medical treatment planning
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1118/1.4897617 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9330.xml