HDR 192Ir source speed measurements using a high speed video camera. Issue 1 (30th December 2014)
- Record Type:
- Journal Article
- Title:
- HDR 192Ir source speed measurements using a high speed video camera. Issue 1 (30th December 2014)
- Main Title:
- HDR 192Ir source speed measurements using a high speed video camera
- Authors:
- Fonseca, Gabriel P.
Viana, Rodrigo S. S.
Podesta, Mark
Rubo, Rodrigo A.
de Sales, Camila P.
Reniers, Brigitte
Yoriyaz, Hélio
Verhaegen, Frank - Abstract:
- Abstract : Purpose: The dose delivered with a HDR 192 Ir afterloader can be separated into a dwell component, and a transit component resulting from the source movement. The transit component is directly dependent on the source speed profile and it is the goal of this study to measure accurate source speed profiles. Methods: A high speed video camera was used to record the movement of a 192 Ir source (Nucletron, an Elekta company, Stockholm, Sweden) for interdwell distances of 0.25–5 cm with dwell times of 0.1, 1, and 2 s. Transit dose distributions were calculated using a Monte Carlo code simulating the source movement. Results: The source stops at each dwell position oscillating around the desired position for a duration up to (0.026 ± 0.005) s. The source speed profile shows variations between 0 and 81 cm/s with average speed of ∼33 cm/s for most of the interdwell distances. The source stops for up to (0.005 ± 0.001) s at nonprogrammed positions in between two programmed dwell positions. The dwell time correction applied by the manufacturer compensates the transit dose between the dwell positions leading to a maximum overdose of 41 mGy for the considered cases and assuming an air‐kerma strength of 48 000 U. The transit dose component is not uniformly distributed leading to over and underdoses, which is within 1.4% for commonly prescribed doses (3–10 Gy). Conclusions: The source maintains its speed even for the short interdwell distances. Dose variations due to the transitAbstract : Purpose: The dose delivered with a HDR 192 Ir afterloader can be separated into a dwell component, and a transit component resulting from the source movement. The transit component is directly dependent on the source speed profile and it is the goal of this study to measure accurate source speed profiles. Methods: A high speed video camera was used to record the movement of a 192 Ir source (Nucletron, an Elekta company, Stockholm, Sweden) for interdwell distances of 0.25–5 cm with dwell times of 0.1, 1, and 2 s. Transit dose distributions were calculated using a Monte Carlo code simulating the source movement. Results: The source stops at each dwell position oscillating around the desired position for a duration up to (0.026 ± 0.005) s. The source speed profile shows variations between 0 and 81 cm/s with average speed of ∼33 cm/s for most of the interdwell distances. The source stops for up to (0.005 ± 0.001) s at nonprogrammed positions in between two programmed dwell positions. The dwell time correction applied by the manufacturer compensates the transit dose between the dwell positions leading to a maximum overdose of 41 mGy for the considered cases and assuming an air‐kerma strength of 48 000 U. The transit dose component is not uniformly distributed leading to over and underdoses, which is within 1.4% for commonly prescribed doses (3–10 Gy). Conclusions: The source maintains its speed even for the short interdwell distances. Dose variations due to the transit dose component are much lower than the prescribed treatment doses for brachytherapy, although transit dose component should be evaluated individually for clinical cases. … (more)
- Is Part Of:
- Medical physics. Volume 42:Issue 1(2015)
- Journal:
- Medical physics
- Issue:
- Volume 42:Issue 1(2015)
- Issue Display:
- Volume 42, Issue 1 (2015)
- Year:
- 2015
- Volume:
- 42
- Issue:
- 1
- Issue Sort Value:
- 2015-0042-0001-0000
- Page Start:
- 412
- Page End:
- 415
- Publication Date:
- 2014-12-30
- Subjects:
- dosimetry -- Monte Carlo methods -- video cameras
Dosimetry/exposure assessment -- Display and recording equipment, oscilloscopes, TV cameras, etc. -- Monte Carlo simulations -- Distribution theory and Monte Carlo studies
Television signal recording -- Scintigraphy
brachytherapy -- Ir‐192 -- source speed -- transit dose
Dosimetry -- Infrared sources -- Cameras -- Brachytherapy -- Infrared detectors -- Optical fiber testing -- Monte Carlo methods -- Ionization chambers -- Time measurement -- Oscilloscopes
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1118/1.4903286 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9326.xml