A system for EPID‐based real‐time treatment delivery verification during dynamic IMRT treatment. Issue 9 (9th August 2013)
- Record Type:
- Journal Article
- Title:
- A system for EPID‐based real‐time treatment delivery verification during dynamic IMRT treatment. Issue 9 (9th August 2013)
- Main Title:
- A system for EPID‐based real‐time treatment delivery verification during dynamic IMRT treatment
- Authors:
- Fuangrod, Todsaporn
Woodruff, Henry C.
van Uytven, Eric
McCurdy, Boyd M. C.
Kuncic, Zdenka
O'Connor, Daryl J.
Greer, Peter B. - Abstract:
- Abstract : Purpose: : To design and develop a real‐time electronic portal imaging device (EPID)‐based delivery verification system for dynamic intensity modulated radiation therapy (IMRT) which enables detection of gross treatment delivery errors before delivery of substantial radiation to the patient. Methods: : The system utilizes a comprehensive physics‐based model to generate a series of predicted transit EPID image frames as a reference dataset and compares these to measured EPID frames acquired during treatment. The two datasets are using MLC aperture comparison and cumulative signal checking techniques. The system operation in real‐time was simulated offline using previously acquired images for 19 IMRT patient deliveries with both frame‐by‐frame comparison and cumulative frame comparison. Simulated error case studies were used to demonstrate the system sensitivity and performance. Results: : The accuracy of the synchronization method was shown to agree within two control points which corresponds to approximately ∼1% of the total MU to be delivered for dynamic IMRT. The system achieved mean real‐time gamma results for frame‐by‐frame analysis of 86.6% and 89.0% for 3%, 3 mm and 4%, 4 mm criteria, respectively, and 97.9% and 98.6% for cumulative gamma analysis. The system can detect a 10% MU error using 3%, 3 mm criteria within approximately 10 s. The EPID‐based real‐time delivery verification system successfully detected simulated gross errors introduced into patientAbstract : Purpose: : To design and develop a real‐time electronic portal imaging device (EPID)‐based delivery verification system for dynamic intensity modulated radiation therapy (IMRT) which enables detection of gross treatment delivery errors before delivery of substantial radiation to the patient. Methods: : The system utilizes a comprehensive physics‐based model to generate a series of predicted transit EPID image frames as a reference dataset and compares these to measured EPID frames acquired during treatment. The two datasets are using MLC aperture comparison and cumulative signal checking techniques. The system operation in real‐time was simulated offline using previously acquired images for 19 IMRT patient deliveries with both frame‐by‐frame comparison and cumulative frame comparison. Simulated error case studies were used to demonstrate the system sensitivity and performance. Results: : The accuracy of the synchronization method was shown to agree within two control points which corresponds to approximately ∼1% of the total MU to be delivered for dynamic IMRT. The system achieved mean real‐time gamma results for frame‐by‐frame analysis of 86.6% and 89.0% for 3%, 3 mm and 4%, 4 mm criteria, respectively, and 97.9% and 98.6% for cumulative gamma analysis. The system can detect a 10% MU error using 3%, 3 mm criteria within approximately 10 s. The EPID‐based real‐time delivery verification system successfully detected simulated gross errors introduced into patient plan deliveries in near real‐time (within 0.1 s). Conclusions: : A real‐time radiation delivery verification system for dynamic IMRT has been demonstrated that is designed to prevent major mistreatments in modern radiation therapy. … (more)
- Is Part Of:
- Medical physics. Volume 40:Issue 9(2013)
- Journal:
- Medical physics
- Issue:
- Volume 40:Issue 9(2013)
- Issue Display:
- Volume 40, Issue 9 (2013)
- Year:
- 2013
- Volume:
- 40
- Issue:
- 9
- Issue Sort Value:
- 2013-0040-0009-0000
- Page Start:
- n/a
- Page End:
- n/a
- Publication Date:
- 2013-08-09
- Subjects:
- Dose‐volume analysis -- Dosimetry/exposure assessment -- Therapeutic applications, including brachytherapy
data acquisition -- dosimetry -- medical image processing -- radiation therapy -- sensitivity -- synchronisation
real‐time treatment verification -- EPID -- intensity‐modulated radiotherapy
Radiation therapy -- Digital computing or data processing equipment or methods, specially adapted for specific applications -- Image data processing or generation, in general
Medical imaging -- Image guided radiation therapy -- Multileaf collimators -- Real time information delivery -- Intensity modulated radiation therapy -- Dosimetry -- Radiation treatment -- Image detection systems -- Linear accelerators -- Drug delivery
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1118/1.4817484 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9325.xml