Determination of the optimal tolerance for MLC positioning in sliding window and VMAT techniques. Issue 4 (26th March 2015)
- Record Type:
- Journal Article
- Title:
- Determination of the optimal tolerance for MLC positioning in sliding window and VMAT techniques. Issue 4 (26th March 2015)
- Main Title:
- Determination of the optimal tolerance for MLC positioning in sliding window and VMAT techniques
- Authors:
- Hernandez, V.
Abella, R.
Calvo, J. F.
Jurado‐Bruggemann, D.
Sancho, I.
Carrasco, P. - Abstract:
- Abstract : Purpose: Several authors have recommended a 2 mm tolerance for multileaf collimator (MLC) positioning in sliding window treatments. In volumetric modulated arc therapy (VMAT) treatments, however, the optimal tolerance for MLC positioning remains unknown. In this paper, the authors present the results of a multicenter study to determine the optimal tolerance for both techniques. Methods: The procedure used is based on dynalog file analysis. The study was carried out using seven Varian linear accelerators from five different centers. Dynalogs were collected from over 100 000 clinical treatments and in‐house software was used to compute the number of tolerance faults as a function of the user‐defined tolerance. Thus, the optimal value for this tolerance, defined as the lowest achievable value, was investigated. Results: Dynalog files accurately predict the number of tolerance faults as a function of the tolerance value, especially for low fault incidences. All MLCs behaved similarly and the Millennium120 and the HD120 models yielded comparable results. In sliding window techniques, the number of beams with an incidence of hold‐offs >1% rapidly decreases for a tolerance of 1.5 mm. In VMAT techniques, the number of tolerance faults sharply drops for tolerances around 2 mm. For a tolerance of 2.5 mm, less than 0.1% of the VMAT arcs presented tolerance faults. Conclusions: Dynalog analysis provides a feasible method for investigating the optimal tolerance for MLCAbstract : Purpose: Several authors have recommended a 2 mm tolerance for multileaf collimator (MLC) positioning in sliding window treatments. In volumetric modulated arc therapy (VMAT) treatments, however, the optimal tolerance for MLC positioning remains unknown. In this paper, the authors present the results of a multicenter study to determine the optimal tolerance for both techniques. Methods: The procedure used is based on dynalog file analysis. The study was carried out using seven Varian linear accelerators from five different centers. Dynalogs were collected from over 100 000 clinical treatments and in‐house software was used to compute the number of tolerance faults as a function of the user‐defined tolerance. Thus, the optimal value for this tolerance, defined as the lowest achievable value, was investigated. Results: Dynalog files accurately predict the number of tolerance faults as a function of the tolerance value, especially for low fault incidences. All MLCs behaved similarly and the Millennium120 and the HD120 models yielded comparable results. In sliding window techniques, the number of beams with an incidence of hold‐offs >1% rapidly decreases for a tolerance of 1.5 mm. In VMAT techniques, the number of tolerance faults sharply drops for tolerances around 2 mm. For a tolerance of 2.5 mm, less than 0.1% of the VMAT arcs presented tolerance faults. Conclusions: Dynalog analysis provides a feasible method for investigating the optimal tolerance for MLC positioning in dynamic fields. In sliding window treatments, the tolerance of 2 mm was found to be adequate, although it can be reduced to 1.5 mm. In VMAT treatments, the typically used 5 mm tolerance is excessively high. Instead, a tolerance of 2.5 mm is recommended. … (more)
- Is Part Of:
- Medical physics. Volume 42:Issue 4(2015)
- Journal:
- Medical physics
- Issue:
- Volume 42:Issue 4(2015)
- Issue Display:
- Volume 42, Issue 4 (2015)
- Year:
- 2015
- Volume:
- 42
- Issue:
- 4
- Issue Sort Value:
- 2015-0042-0004-0000
- Page Start:
- 1911
- Page End:
- 1916
- Publication Date:
- 2015-03-26
- Subjects:
- collimators -- radiation therapy
Therapeutic applications, including brachytherapy -- Collimators
Radiation therapy -- Using diaphragms, collimators
VMAT -- tolerance -- MLC positioning -- dynamic treatments
Multileaf collimators -- Fault dynamics -- Dosimetry -- Data analysis -- Intensity modulated radiation therapy -- Linear accelerators -- Computer software -- Control systems -- Drug delivery -- Medical treatment planning
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1118/1.4915541 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9329.xml