Impact of plan parameters on the dosimetric accuracy of volumetric modulated arc therapy. Issue 7 (18th June 2013)
- Record Type:
- Journal Article
- Title:
- Impact of plan parameters on the dosimetric accuracy of volumetric modulated arc therapy. Issue 7 (18th June 2013)
- Main Title:
- Impact of plan parameters on the dosimetric accuracy of volumetric modulated arc therapy
- Authors:
- Masi, Laura
Doro, Raffaela
Favuzza, Virginia
Cipressi, Samantha
Livi, Lorenzo - Abstract:
- Abstract : Purpose: : To evaluate the effect of plan parameters on volumetric modulated arc therapy (VMAT) dosimetric accuracy, together with the possibility of scoring plan complexity. Methods: : 142 clinical VMAT plans initially optimized using a 4° control point (CP) separation were evaluated. All plans were delivered by a 6 MV Linac to a biplanar diode array for patient‐specific quality assurance (QA). Local Γ index analysis (3%, 3 mm and 2%, 2 mm) enabled the comparison between delivered and calculated dose. The following parameters were considered for each plan: average leaf travel (LT), modulation complexity score applied to VMAT (MCSv), MU value, and a multiplicative combination of LT and MCSv (LTMCS). Pearson's correlation analysis was performed between Γ passing rates and each parameter. The effects of CP angular separation on VMAT dosimetric accuracy were also analyzed by focusing on plans with high LT values. Forty out of 142 plans with LT above 350 mm were further optimized using a finer angle spacing (3° or 2°) and Γ analysis was performed. The average Γ passing rates obtained at 4° and at 3°/2° sampling were compared. A further correlation analysis between all parameters and the Γ pass‐rates was performed on 142 plans, but including the newly optimized 40 plans (CP every 3° or 2°) in place of the old ones (CP every 4°). Results: : A moderate significant ( p < 0.05) correlation between each examined parameter and Γ passing rates was observed for the originalAbstract : Purpose: : To evaluate the effect of plan parameters on volumetric modulated arc therapy (VMAT) dosimetric accuracy, together with the possibility of scoring plan complexity. Methods: : 142 clinical VMAT plans initially optimized using a 4° control point (CP) separation were evaluated. All plans were delivered by a 6 MV Linac to a biplanar diode array for patient‐specific quality assurance (QA). Local Γ index analysis (3%, 3 mm and 2%, 2 mm) enabled the comparison between delivered and calculated dose. The following parameters were considered for each plan: average leaf travel (LT), modulation complexity score applied to VMAT (MCSv), MU value, and a multiplicative combination of LT and MCSv (LTMCS). Pearson's correlation analysis was performed between Γ passing rates and each parameter. The effects of CP angular separation on VMAT dosimetric accuracy were also analyzed by focusing on plans with high LT values. Forty out of 142 plans with LT above 350 mm were further optimized using a finer angle spacing (3° or 2°) and Γ analysis was performed. The average Γ passing rates obtained at 4° and at 3°/2° sampling were compared. A further correlation analysis between all parameters and the Γ pass‐rates was performed on 142 plans, but including the newly optimized 40 plans (CP every 3° or 2°) in place of the old ones (CP every 4°). Results: : A moderate significant ( p < 0.05) correlation between each examined parameter and Γ passing rates was observed for the original 142 plans at 4° CP discretization. A negative correlation was found for LT with Pearson's r absolute values above 0.6, suggesting that a lower dosimetric accuracy may be expected for higher LT values when a 4° CP sampling is used. A positive correlation was observed for MCSv and LTMCS with r values above 0.5. In order to score plan complexity, threshold values of LTMCS were defined. The average Γ passing rates were significantly higher for the plans created using the finer CP spacing (3°/2°) compared to the plans optimized using the standard 4° spacing (Student t ‐test p < 0.05). The correlation between LT and passing rates was strongly diminished when plans with finer angular separations were considered, yielding Pearson's r absolute values below 0.45. Conclusions: : At 4° CP sampling, LT, MCSv, and LTMCS were found to be significantly correlated with VMAT dosimetric accuracy, expressed as Γ pass‐rates. These parameters were found to be possible candidates for scoring plan complexity using threshold values. A finer CP separation (3°/2°) led to a significant increase in dosimetric accuracy for plans with high leaf travel values, and to a decrease in correlation between LT and Γ passing rates. These results indicated that the influence of LT on VMAT dosimetric accuracy can be controlled by reducing CP separation. CP spacing for all plans requiring large leaf motion should not exceed 3°. The reported data were integrated to optimize our clinical workflow for plan creation, optimization, selection among rival plans, and patient‐specific QA of VMAT treatments. … (more)
- Is Part Of:
- Medical physics. Volume 40:Issue 7(2013)
- Journal:
- Medical physics
- Issue:
- Volume 40:Issue 7(2013)
- Issue Display:
- Volume 40, Issue 7 (2013)
- Year:
- 2013
- Volume:
- 40
- Issue:
- 7
- Issue Sort Value:
- 2013-0040-0007-0000
- Page Start:
- n/a
- Page End:
- n/a
- Publication Date:
- 2013-06-18
- Subjects:
- Dosimetry/exposure assessment -- Therapeutic applications, including brachytherapy -- Optimization -- Quality assurance in radiotherapy
dosimetry -- radiation therapy
VMAT -- planning parameters
Radiation therapy
Intensity modulated radiation therapy -- Optical microcavities -- Cancer -- Medical treatment planning -- Multileaf collimators -- Data analysis -- Quality assurance -- Therapeutics -- Linear accelerators -- Anatomy
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1118/1.4810969 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9312.xml