Measurement of neutron dose equivalent outside and inside of the treatment vault of GRID therapy. Issue 9 (5th August 2013)
- Record Type:
- Journal Article
- Title:
- Measurement of neutron dose equivalent outside and inside of the treatment vault of GRID therapy. Issue 9 (5th August 2013)
- Main Title:
- Measurement of neutron dose equivalent outside and inside of the treatment vault of GRID therapy
- Authors:
- Wang, Xudong
Charlton, Michael A.
Esquivel, Carlos
Eng, Tony Y.
Li, Ying
Papanikolaou, Nikos - Abstract:
- Abstract : Purpose : : To evaluate the neutron and photon dose equivalent rates at the treatment vault entrance (Hn, D and HG ), and to study the secondary radiation to the patient in GRID therapy. The radiation activation on the grid was studied. Methods: : A Varian Clinac 23EX accelerator was working at 18 MV mode with a grid manufactured by .decimal, Inc. The Hn, D and HG were measured using an Andersson–Braun neutron REM meter, and a Geiger Müller counter. The radiation activation on the grid was measured after the irradiation with an ion chamber γ ‐ray survey meter. The secondary radiation dose equivalent to patient was evaluated by etched track detectors and OSL detectors on a RANDO ® phantom. Results : : Within the measurement uncertainty, there is no significant difference between the Hn, D and HG with and without a grid. However, the neutron dose equivalent to the patient with the grid is, on average, 35.3% lower than that without the grid when using the same field size and the same amount of monitor unit. The photon dose equivalent to the patient with the grid is, on average, 44.9% lower. The measured average half‐life of the radiation activation in the grid is 12.0 (±0.9) min. The activation can be categorized into a fast decay component and a slow decay component with half‐lives of 3.4 (±1.6) min and 15.3 (±4.0) min, respectively. There was no detectable radioactive contamination found on the surface of the grid through a wipe test. Conclusions : : This workAbstract : Purpose : : To evaluate the neutron and photon dose equivalent rates at the treatment vault entrance (Hn, D and HG ), and to study the secondary radiation to the patient in GRID therapy. The radiation activation on the grid was studied. Methods: : A Varian Clinac 23EX accelerator was working at 18 MV mode with a grid manufactured by .decimal, Inc. The Hn, D and HG were measured using an Andersson–Braun neutron REM meter, and a Geiger Müller counter. The radiation activation on the grid was measured after the irradiation with an ion chamber γ ‐ray survey meter. The secondary radiation dose equivalent to patient was evaluated by etched track detectors and OSL detectors on a RANDO ® phantom. Results : : Within the measurement uncertainty, there is no significant difference between the Hn, D and HG with and without a grid. However, the neutron dose equivalent to the patient with the grid is, on average, 35.3% lower than that without the grid when using the same field size and the same amount of monitor unit. The photon dose equivalent to the patient with the grid is, on average, 44.9% lower. The measured average half‐life of the radiation activation in the grid is 12.0 (±0.9) min. The activation can be categorized into a fast decay component and a slow decay component with half‐lives of 3.4 (±1.6) min and 15.3 (±4.0) min, respectively. There was no detectable radioactive contamination found on the surface of the grid through a wipe test. Conclusions : : This work indicates that there is no significant change of the Hn, D and HG in GRID therapy, compared with a conventional external beam therapy. However, the neutron and scattered photon dose equivalent to the patient decrease dramatically with the grid and can be clinical irrelevant. Meanwhile, the users of a grid should be aware of the possible high dose to the radiation worker from the radiation activation on the surface of the grid. A delay in handling the grid after the beam delivery is suggested. … (more)
- Is Part Of:
- Medical physics. Volume 40:Issue 9(2013)
- Journal:
- Medical physics
- Issue:
- Volume 40:Issue 9(2013)
- Issue Display:
- Volume 40, Issue 9 (2013)
- Year:
- 2013
- Volume:
- 40
- Issue:
- 9
- Issue Sort Value:
- 2013-0040-0009-0000
- Page Start:
- n/a
- Page End:
- n/a
- Publication Date:
- 2013-08-05
- Subjects:
- Dosimetry/exposure assessment -- Therapeutic applications, including brachytherapy
dosimetry -- Geiger counters -- linear accelerators -- measurement uncertainty -- phantoms -- radiation therapy
GRID therapy -- photoneutrons
Radiation therapy -- Linear accelerators
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1118/1.4816653 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9304.xml