Development of a phantom to validate high‐dose‐rate brachytherapy treatment planning systems with heterogeneous algorithms. Issue 4 (16th March 2015)
- Record Type:
- Journal Article
- Title:
- Development of a phantom to validate high‐dose‐rate brachytherapy treatment planning systems with heterogeneous algorithms. Issue 4 (16th March 2015)
- Main Title:
- Development of a phantom to validate high‐dose‐rate brachytherapy treatment planning systems with heterogeneous algorithms
- Authors:
- Moura, Eduardo S.
Micka, John A.
Hammer, Cliff G.
Culberson, Wesley S.
DeWerd, Larry A.
Rostelato, Maria Elisa C. M.
Zeituni, Carlos A. - Abstract:
- Abstract : Purpose: This work presents the development of a phantom to verify the treatment planning system (TPS) algorithms used for high‐dose‐rate (HDR) brachytherapy. It is designed to measure the relative dose in a heterogeneous media. The experimental details used, simulation methods, and comparisons with a commercial TPS are also provided. Methods: To simulate heterogeneous conditions, four materials were used: Virtual Water™ (VM), BR50/50™, cork, and aluminum. The materials were arranged in 11 heterogeneity configurations. Three dosimeters were used to measure the relative response from a HDR 192 Ir source: TLD‐100™, Gafchromic ® EBT3 film, and an Exradin™ A1SL ionization chamber. To compare the results from the experimental measurements, the various configurations were modeled in thepenelope /penEasy Monte Carlo code. Images of each setup geometry were acquired from a CT scanner and imported into BrachyVision™ TPS software, which includes a grid‐based Boltzmann solver Acuros™. The results of the measurements performed in the heterogeneous setups were normalized to the dose values measured in the homogeneous Virtual Water™ setup and the respective differences due to the heterogeneities were considered. Additionally, dose values calculated based on the American Association of Physicists in Medicine‐Task Group 43 formalism were compared to dose values calculated with the Acuros™ algorithm in the phantom. Calculated doses were compared at the same points, whereAbstract : Purpose: This work presents the development of a phantom to verify the treatment planning system (TPS) algorithms used for high‐dose‐rate (HDR) brachytherapy. It is designed to measure the relative dose in a heterogeneous media. The experimental details used, simulation methods, and comparisons with a commercial TPS are also provided. Methods: To simulate heterogeneous conditions, four materials were used: Virtual Water™ (VM), BR50/50™, cork, and aluminum. The materials were arranged in 11 heterogeneity configurations. Three dosimeters were used to measure the relative response from a HDR 192 Ir source: TLD‐100™, Gafchromic ® EBT3 film, and an Exradin™ A1SL ionization chamber. To compare the results from the experimental measurements, the various configurations were modeled in thepenelope /penEasy Monte Carlo code. Images of each setup geometry were acquired from a CT scanner and imported into BrachyVision™ TPS software, which includes a grid‐based Boltzmann solver Acuros™. The results of the measurements performed in the heterogeneous setups were normalized to the dose values measured in the homogeneous Virtual Water™ setup and the respective differences due to the heterogeneities were considered. Additionally, dose values calculated based on the American Association of Physicists in Medicine‐Task Group 43 formalism were compared to dose values calculated with the Acuros™ algorithm in the phantom. Calculated doses were compared at the same points, where measurements have been performed. Results: Differences in the relative response as high as 11.5% were found from the homogeneous setup when the heterogeneous materials were inserted into the experimental phantom. The aluminum and cork materials produced larger differences than the plastic materials, with the BR50/50™ material producing results similar to the Virtual Water™ results. Our experimental methods agree with thepenelope /penEasy simulations for most setups and dosimeters. The TPS relative differences with the Acuros™ algorithm were similar in both experimental and simulated setups. The discrepancy between the BrachyVision™, Acuros™, and TG‐43 dose responses in the phantom described by this work exceeded 12% for certain setups. Conclusions: The results derived from the phantom measurements show good agreement with the simulations and TPS calculations, using Acuros™ algorithm. Differences in the dose responses were evident in the experimental results when heterogeneous materials were introduced. These measurements prove the usefulness of the heterogeneous phantom for verification of HDR treatment planning systems based on model‐based dose calculation algorithms. … (more)
- Is Part Of:
- Medical physics. Volume 42:Issue 4(2015)
- Journal:
- Medical physics
- Issue:
- Volume 42:Issue 4(2015)
- Issue Display:
- Volume 42, Issue 4 (2015)
- Year:
- 2015
- Volume:
- 42
- Issue:
- 4
- Issue Sort Value:
- 2015-0042-0004-0000
- Page Start:
- 1566
- Page End:
- 1574
- Publication Date:
- 2015-03-16
- Subjects:
- aluminium -- brachytherapy -- dosimetry -- ionisation chambers -- phantoms -- thermoluminescent dosimeters
Therapeutic applications, including brachytherapy -- Dosimetry/exposure assessment -- Gas‐filled counters: ionization chambers, proportional, and avalanche counters
Radiation therapy -- Measurement of nuclear or x‐radiation -- Tubes for determining the presence, intensity, density or energy of radiation or particles -- Measuring radiation intensity -- Scintigraphy -- with scintillation detectors -- Ionisation chambers
brachytherapy -- 192Ir -- treatment planning system -- heterogeneities
Dosimetry -- Thermoluminescent dosimeters -- Monte Carlo methods -- Brachytherapy -- Infrared sources -- Aluminium -- Medical imaging -- Tissues -- Backscattering -- Medical treatment planning
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1118/1.4914390 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9312.xml