Dynamic trajectory‐based couch motion for improvement of radiation therapy trajectories in cranial SRT. Issue 5 (15th April 2015)
- Record Type:
- Journal Article
- Title:
- Dynamic trajectory‐based couch motion for improvement of radiation therapy trajectories in cranial SRT. Issue 5 (15th April 2015)
- Main Title:
- Dynamic trajectory‐based couch motion for improvement of radiation therapy trajectories in cranial SRT
- Authors:
- Lee MacDonald, R.
Thomas, Christopher G. - Abstract:
- Abstract : Purpose: To investigate potential improvement in external beam stereotactic radiation therapy plan quality for cranial cases using an optimized dynamic gantry and patient support couch motion trajectory, which could minimize exposure to sensitive healthy tissue. Methods: Anonymized patient anatomy and treatment plans of cranial cancer patients were used to quantify the geometric overlap between planning target volumes and organs‐at‐risk (OARs) based on their two‐dimensional projection from source to a plane at isocenter as a function of gantry and couch angle. Published dose constraints were then used as weighting factors for the OARs to generate a map of couch‐gantry coordinate space, indicating degree of overlap at each point in space. A couch‐gantry collision space was generated by direct measurement on a linear accelerator and couch using an anthropomorphic solid‐water phantom. A dynamic, fully customizable algorithm was written to generate a navigable ideal trajectory for the patient specific couch‐gantry space. The advanced algorithm can be used to balance the implementation of absolute minimum values of overlap with the clinical practicality of large‐scale couch motion and delivery time. Optimized cranial cancer treatment trajectories were compared to conventional treatment trajectories. Results: Comparison of optimized treatment trajectories with conventional treatment trajectories indicated an average decrease in mean dose to the OARs of 19% and anAbstract : Purpose: To investigate potential improvement in external beam stereotactic radiation therapy plan quality for cranial cases using an optimized dynamic gantry and patient support couch motion trajectory, which could minimize exposure to sensitive healthy tissue. Methods: Anonymized patient anatomy and treatment plans of cranial cancer patients were used to quantify the geometric overlap between planning target volumes and organs‐at‐risk (OARs) based on their two‐dimensional projection from source to a plane at isocenter as a function of gantry and couch angle. Published dose constraints were then used as weighting factors for the OARs to generate a map of couch‐gantry coordinate space, indicating degree of overlap at each point in space. A couch‐gantry collision space was generated by direct measurement on a linear accelerator and couch using an anthropomorphic solid‐water phantom. A dynamic, fully customizable algorithm was written to generate a navigable ideal trajectory for the patient specific couch‐gantry space. The advanced algorithm can be used to balance the implementation of absolute minimum values of overlap with the clinical practicality of large‐scale couch motion and delivery time. Optimized cranial cancer treatment trajectories were compared to conventional treatment trajectories. Results: Comparison of optimized treatment trajectories with conventional treatment trajectories indicated an average decrease in mean dose to the OARs of 19% and an average decrease in maximum dose to the OARs of 12%. Degradation was seen for homogeneity index (6.14% ± 0.67%–5.48% ± 0.76%) and conformation number (0.82 ± 0.02–0.79 ± 0.02), but neither was statistically significant. Removal of OAR constraints from volumetric modulated arc therapy optimization reveals that reduction in dose to OARs is almost exclusively due to the optimized trajectory and not the OAR constraints. Conclusions: The authors' study indicated that simultaneous couch and gantry motion during radiation therapy to minimize the geometrical overlap in the beams‐eye‐view of target volumes and the organs‐at‐risk can have an appreciable dose reduction to organs‐at‐risk. … (more)
- Is Part Of:
- Medical physics. Volume 42:Issue 5(2015)
- Journal:
- Medical physics
- Issue:
- Volume 42:Issue 5(2015)
- Issue Display:
- Volume 42, Issue 5 (2015)
- Year:
- 2015
- Volume:
- 42
- Issue:
- 5
- Issue Sort Value:
- 2015-0042-0005-0000
- Page Start:
- 2317
- Page End:
- 2325
- Publication Date:
- 2015-04-15
- Subjects:
- biological tissues -- cancer -- dosimetry -- linear accelerators -- phantoms -- radiation therapy -- statistical analysis
Therapeutic applications, including brachytherapy -- Cancer -- Dosimetry/exposure assessment -- Probability theory, stochastic processes, and statistics
Radiation therapy -- Linear accelerators -- Scintigraphy
dynamic couch motion -- SRT -- cranial -- trajectory -- algorithm
Dosimetry -- Kinematics -- Cancer -- Tissues -- Radiation treatment -- Optimization -- Interpolation -- Linear accelerators
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1118/1.4917165 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9314.xml