A method for accurate zero calibration of asymmetric jaws in single‐isocenter half‐beam techniques. Issue 2 (10th January 2013)
- Record Type:
- Journal Article
- Title:
- A method for accurate zero calibration of asymmetric jaws in single‐isocenter half‐beam techniques. Issue 2 (10th January 2013)
- Main Title:
- A method for accurate zero calibration of asymmetric jaws in single‐isocenter half‐beam techniques
- Authors:
- Hernandez, V.
Sempau, J.
Abella, R.
Lopez, M.
Perez, M.
Artigues, M.
Arenas, M. - Abstract:
- Abstract : Purpose: : To present a practical method for calibrating the zero position of asymmetric jaws that provides higher accuracy at the central axis and improves dose homogeneity in the abutting region of half‐beams. Methods: : Junction doses were measured for each asymmetric jaw using the double‐exposure technique and electronic portal imaging devices. The junction dose was determined as a function of jaw position. The shift in the zero jaw position (or in its corresponding potentiometer readout) required to correct for the measured junction dose could thus be obtained. The jaw calibration was then modified to introduce the calculated shift and therefore achieve an accurate zero position in order to provide a relative junction dose that was as close to zero as possible. Results: : All the asymmetric jaws from four medical linear accelerators were calibrated with the new calibration procedure. Measured relative junction doses at gantry 0° were reduced from a maximum of ±40% to a maximum of ±8% for all the jaws in the four considered accelerators. These results were valid for 6 MV and 18 MV photon beams and for any combination of asymmetric jaws set to zero. The calibration was stable over a long period of time; therefore, the need for recalibrating is seldom necessary. Conclusions: : Accurate calibration of the zero position of the jaws is feasible in current medical linear accelerators. The proposed procedure is fast and it improves dose homogeneity at the junction ofAbstract : Purpose: : To present a practical method for calibrating the zero position of asymmetric jaws that provides higher accuracy at the central axis and improves dose homogeneity in the abutting region of half‐beams. Methods: : Junction doses were measured for each asymmetric jaw using the double‐exposure technique and electronic portal imaging devices. The junction dose was determined as a function of jaw position. The shift in the zero jaw position (or in its corresponding potentiometer readout) required to correct for the measured junction dose could thus be obtained. The jaw calibration was then modified to introduce the calculated shift and therefore achieve an accurate zero position in order to provide a relative junction dose that was as close to zero as possible. Results: : All the asymmetric jaws from four medical linear accelerators were calibrated with the new calibration procedure. Measured relative junction doses at gantry 0° were reduced from a maximum of ±40% to a maximum of ±8% for all the jaws in the four considered accelerators. These results were valid for 6 MV and 18 MV photon beams and for any combination of asymmetric jaws set to zero. The calibration was stable over a long period of time; therefore, the need for recalibrating is seldom necessary. Conclusions: : Accurate calibration of the zero position of the jaws is feasible in current medical linear accelerators. The proposed procedure is fast and it improves dose homogeneity at the junction of half‐beams, thus, allowing a more accurate and safer use of these techniques. … (more)
- Is Part Of:
- Medical physics. Volume 40:Issue 2(2013)
- Journal:
- Medical physics
- Issue:
- Volume 40:Issue 2(2013)
- Issue Display:
- Volume 40, Issue 2 (2013)
- Year:
- 2013
- Volume:
- 40
- Issue:
- 2
- Issue Sort Value:
- 2013-0040-0002-0000
- Page Start:
- n/a
- Page End:
- n/a
- Publication Date:
- 2013-01-10
- Subjects:
- Dose‐volume analysis -- Standards and calibration
calibration -- dosimetry -- linear accelerators -- potentiometers -- radiation therapy
jaw calibration -- asymmetric jaws -- field matching -- half‐beams -- junction dose
Radiation therapy -- Testing or calibrating of apparatus or arrangements provided for in groups G01D1/00 to G01D15/00 -- Calibrating of instruments or apparatus -- Adjustable resistors -- Linear accelerators
Calibration -- Dosimetry -- Linear accelerators -- Collimators -- Medical imaging -- Image guided radiation therapy -- Biomedical modeling -- Computer software -- Image sensors -- Photons
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1118/1.4773314 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9307.xml