Aortic valve perforation following Amplatzer vascular plug device closure of 'pulsatile' Fontan. (December 2017)
- Record Type:
- Journal Article
- Title:
- Aortic valve perforation following Amplatzer vascular plug device closure of 'pulsatile' Fontan. (December 2017)
- Main Title:
- Aortic valve perforation following Amplatzer vascular plug device closure of 'pulsatile' Fontan
- Authors:
- Madan, Nitin
Singh, Harinder R.
Mumtaz, Muhammad A.
Gowda, Srinath - Abstract:
- Abstract: Persistent antegrade pulmonary blood flow can interfere with optimal functioning of a Fontan circulation in patients with single ventricle physiology. Typically this is addressed surgically at the time of the Fontan procedure with oversewing of the main pulmonary artery, however different variations to this approach have been described. In patients with residual antegrade pulmonary blood flow after a Fontan procedure, "pulsatile" flow into the Fontan system can cause unfavorable hemodynamics leading to various complications. Successful transcatheter device closure of the pulmonary valve in both symptomatic and asymptomatic patients has been described. We report a patient where inadvertent residual antegrade pulmonary blood flow was closed with Amplatzer vascular plug II device with resultant aortic valve perforation noted later in the course requiring surgical removal of the device, aortic valve repair and transection of the main pulmonary artery. Highlights: Persistent antegrade pulmonary blood flow interferes with optimal function of Fontan. Successful transcatheter closure of 'pulsatile' Fontan has been previously described. Aortic valve perforation following Amplatzer vascular plug closure is reported here. The necessity for proper patient selection prior to transcatheter closure is highlighted.
- Is Part Of:
- Progress in pediatric cardiology. Volume 47(2017)
- Journal:
- Progress in pediatric cardiology
- Issue:
- Volume 47(2017)
- Issue Display:
- Volume 47, Issue 2017 (2017)
- Year:
- 2017
- Volume:
- 47
- Issue:
- 2017
- Issue Sort Value:
- 2017-0047-2017-0000
- Page Start:
- 84
- Page End:
- 86
- Publication Date:
- 2017-12
- Subjects:
- Fontan -- Fontan complications -- Device closure -- Valve perforation
Pediatric cardiology -- Periodicals
Cardiovascular Diseases -- Periodicals
Infant
Child
Cardiologie pédiatrique -- Périodiques
618.9212005 - Journal URLs:
- http://www.sciencedirect.com/science/journal/10589813 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/10589813 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/10589813 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.ppedcard.2017.05.005 ↗
- Languages:
- English
- ISSNs:
- 1058-9813
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6872.440000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9198.xml