Analysis of dose verification results for 924 intensity‐modulated radiation therapy plans. Issue Volume 2:Issues 4(2018) (19th December 2018)
- Record Type:
- Journal Article
- Title:
- Analysis of dose verification results for 924 intensity‐modulated radiation therapy plans. Issue Volume 2:Issues 4(2018) (19th December 2018)
- Main Title:
- Analysis of dose verification results for 924 intensity‐modulated radiation therapy plans
- Authors:
- Wu, Shizhang
Chen, Jinhu
Li, Zhenjiang
Qiu, Qingtao
Wang, Xingli
Li, Chengqiang
Yin, Yong - Abstract:
- Abstract: Objective: Intensity‐modulated radiation therapy (IMRT) plays an increasingly important role in clinical applications, and dose verification is particularly crucial for each patient. In the present study, we retrospectively analyzed the results of dose verification in 924 actual IMRT plans, including the relationship between gamma pass rates and the location of lesions, and the total number of monitor units, and the maximum area of the collected dose. Methods: All the 924 IMRT plans implemented in a Varian Trilogy accelerator between 1 January 2014 and 31 March 2016 at Shandong Cancer Hospital were acquired. The Varian Eclipse planning system was used for all treatment planning and verification. Then, actual implemented plans were transplanted into a water phantom. Subsequently, the derived fluence was compared with the actual measured fluence by gamma analysis with the gamma criteria (3%/3 mm). Results: A total of 924 IMRT plans were categorized into six groups, including the brain, head and neck, chest, abdomen, pelvis, and breast. The gamma pass rate average was >98% for 902 IMRT plans, whereas 22 plans did not pass the first time. A correlation was observed between the treatment site and gamma pass rate ( P = 0.017). Meanwhile, a negative correlation was observed between the gamma pass rate and the total number of monitor units ( P < 0.001), and the largest area of the acquisition dose (P < 0.001), respectively. Varian Trilogy accelerator IMRT QA data has aAbstract: Objective: Intensity‐modulated radiation therapy (IMRT) plays an increasingly important role in clinical applications, and dose verification is particularly crucial for each patient. In the present study, we retrospectively analyzed the results of dose verification in 924 actual IMRT plans, including the relationship between gamma pass rates and the location of lesions, and the total number of monitor units, and the maximum area of the collected dose. Methods: All the 924 IMRT plans implemented in a Varian Trilogy accelerator between 1 January 2014 and 31 March 2016 at Shandong Cancer Hospital were acquired. The Varian Eclipse planning system was used for all treatment planning and verification. Then, actual implemented plans were transplanted into a water phantom. Subsequently, the derived fluence was compared with the actual measured fluence by gamma analysis with the gamma criteria (3%/3 mm). Results: A total of 924 IMRT plans were categorized into six groups, including the brain, head and neck, chest, abdomen, pelvis, and breast. The gamma pass rate average was >98% for 902 IMRT plans, whereas 22 plans did not pass the first time. A correlation was observed between the treatment site and gamma pass rate ( P = 0.017). Meanwhile, a negative correlation was observed between the gamma pass rate and the total number of monitor units ( P < 0.001), and the largest area of the acquisition dose (P < 0.001), respectively. Varian Trilogy accelerator IMRT QA data has a stable pass rate with a 100 confidence limit (CL) value of 95.19. Conclusion: There was a correlation between the pass rate and the treatment site, the total number of monitor units, and the maximum area of collected dose. When using Mapcheck for patient plan dose verification, the gamma pass rate is very high, and only a few of them need to be analyzed. … (more)
- Is Part Of:
- Precision radiation oncology. Volume 2:Issues 4(2018)
- Journal:
- Precision radiation oncology
- Issue:
- Volume 2:Issues 4(2018)
- Issue Display:
- Volume 2, Issue 4 (2018)
- Year:
- 2018
- Volume:
- 2
- Issue:
- 4
- Issue Sort Value:
- 2018-0002-0004-0000
- Page Start:
- 125
- Page End:
- 130
- Publication Date:
- 2018-12-19
- Subjects:
- dose verification -- gamma pass rate -- Mapcheck
- Journal URLs:
- http://onlinelibrary.wiley.com/ ↗
- DOI:
- 10.1002/pro6.58 ↗
- Languages:
- English
- ISSNs:
- 2398-7324
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9180.xml