A simple, selective, and sensitive gas chromatography–mass spectrometry method for the analysis of five process‐related impurities in atenolol bulk drug and capsule formulations. Issue 15 (28th June 2017)
- Record Type:
- Journal Article
- Title:
- A simple, selective, and sensitive gas chromatography–mass spectrometry method for the analysis of five process‐related impurities in atenolol bulk drug and capsule formulations. Issue 15 (28th June 2017)
- Main Title:
- A simple, selective, and sensitive gas chromatography–mass spectrometry method for the analysis of five process‐related impurities in atenolol bulk drug and capsule formulations
- Authors:
- Reddy, Ambavaram Vijaya Bhaskar
Yusop, Zulkifli
Jaafar, Jafariah
Bin Aris, Azmi
Abdul Majid, Zaiton - Abstract:
- Abstract : An extremely sensitive and simple gas chromatography with mass spectrometry method was developed and completely validated for the analysis of five process‐related impurities, viz., 4‐hydroxy‐l ‐phenylglycine, 4‐hydroxyphenylacetonitrile, 4‐hydroxyphenylacetic acid, methyl‐4‐hydroxyphenylacetate, and 2‐[4‐{(2 RS )‐2‐hydroxy‐3‐[(1‐methylethyl)amino]propoxy}phenyl]acetonitrile, in atenolol. The separation of impurities was accomplished on a BPX‐5 column with dimensions of 50 m × 0.25 mm i.d. and 0.25 μm film thickness. The method validation was performed following International Conference on Harmonisation guidelines in which the method was capable to quantitate 4‐hydroxy‐l ‐phenylglycine, 4‐hydroxyphenylacetonitrile, and 4‐hydroxyphenylacetic acid at 0.3 ppm, and methyl‐4‐hydroxyphenylacetate and 2‐[4‐{(2 RS )‐2‐hydroxy‐3‐[(1‐methylethyl)amino]propoxy}phenyl]acetonitrile at 0.35 ppm with respect to 10 mg/mL of atenolol. The method was linear over the concentration range of 0.3–10 ppm for 4‐hydroxy‐l ‐phenylglycine, 4‐hydroxyphenylacetonitrile, and 4‐hydroxyphenylacetic acid, and 0.35–10 ppm for methyl‐4‐hydroxyphenylacetate and 2‐[4‐{(2 RS )‐2‐hydroxy‐3‐[(1‐methylethyl)amino]propoxy}phenyl]acetonitrile. The correlation coefficient in each case was found ≥0.998. The repeatability and recovery values were acceptable, and found between 89.38% and 105.60% for all five impurities under optimized operating conditions. The method developed here is simple, selective, andAbstract : An extremely sensitive and simple gas chromatography with mass spectrometry method was developed and completely validated for the analysis of five process‐related impurities, viz., 4‐hydroxy‐l ‐phenylglycine, 4‐hydroxyphenylacetonitrile, 4‐hydroxyphenylacetic acid, methyl‐4‐hydroxyphenylacetate, and 2‐[4‐{(2 RS )‐2‐hydroxy‐3‐[(1‐methylethyl)amino]propoxy}phenyl]acetonitrile, in atenolol. The separation of impurities was accomplished on a BPX‐5 column with dimensions of 50 m × 0.25 mm i.d. and 0.25 μm film thickness. The method validation was performed following International Conference on Harmonisation guidelines in which the method was capable to quantitate 4‐hydroxy‐l ‐phenylglycine, 4‐hydroxyphenylacetonitrile, and 4‐hydroxyphenylacetic acid at 0.3 ppm, and methyl‐4‐hydroxyphenylacetate and 2‐[4‐{(2 RS )‐2‐hydroxy‐3‐[(1‐methylethyl)amino]propoxy}phenyl]acetonitrile at 0.35 ppm with respect to 10 mg/mL of atenolol. The method was linear over the concentration range of 0.3–10 ppm for 4‐hydroxy‐l ‐phenylglycine, 4‐hydroxyphenylacetonitrile, and 4‐hydroxyphenylacetic acid, and 0.35–10 ppm for methyl‐4‐hydroxyphenylacetate and 2‐[4‐{(2 RS )‐2‐hydroxy‐3‐[(1‐methylethyl)amino]propoxy}phenyl]acetonitrile. The correlation coefficient in each case was found ≥0.998. The repeatability and recovery values were acceptable, and found between 89.38% and 105.60% for all five impurities under optimized operating conditions. The method developed here is simple, selective, and sensitive with apparently better resolution than the reported methods. Hence, the method is a straightforward and good quality control tool for the quantitation of selected impurities at trace concentrations in atenolol. … (more)
- Is Part Of:
- Journal of separation science. Volume 40:Issue 15(2017)
- Journal:
- Journal of separation science
- Issue:
- Volume 40:Issue 15(2017)
- Issue Display:
- Volume 40, Issue 15 (2017)
- Year:
- 2017
- Volume:
- 40
- Issue:
- 15
- Issue Sort Value:
- 2017-0040-0015-0000
- Page Start:
- 3086
- Page End:
- 3093
- Publication Date:
- 2017-06-28
- Subjects:
- beta‐blockers -- genotoxicity -- mass spectrometry -- method validation
Separation (Technology) -- Periodicals
Chromatographic analysis -- Periodicals
543.089 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)1615-9314 ↗
http://www.interscience.wiley.com/jpages/1615-9306 ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/jssc.201700252 ↗
- Languages:
- English
- ISSNs:
- 1615-9306
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5063.880000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9176.xml