Comparison between the TRS-398 code of practice and the TG-51 dosimetry protocol for flattening filter free beams. (1st July 2016)
- Record Type:
- Journal Article
- Title:
- Comparison between the TRS-398 code of practice and the TG-51 dosimetry protocol for flattening filter free beams. (1st July 2016)
- Main Title:
- Comparison between the TRS-398 code of practice and the TG-51 dosimetry protocol for flattening filter free beams
- Authors:
- Lye, J E
Butler, D J
Oliver, C P
Alves, A
Lehmann, J
Gibbons, F P
Williams, I M - Abstract:
- Abstract: Dosimetry protocols for external beam radiotherapy currently in use, such as the IAEA TRS-398 and AAPM TG-51, were written for conventional linear accelerators. In these accelerators, a flattening filter is used to produce a beam which is uniform at water depths where the ionization chamber is used to measure the absorbed dose. Recently, clinical linacs have been implemented without the flattening filter, and published theoretical analysis suggested that with these beams a dosimetric error of order 0.6% could be expected for IAEA TRS-398, because the TPR20, 10 beam quality index does not accurately predict the stopping power ratio (water to air) for the softer flattening-filter-free (FFF) beam spectra. We measured doses on eleven FFF linacs at 6 MV and 10 MV using both dosimetry protocols and found average differences of 0.2% or less. The expected shift due to stopping powers was not observed. We present Monte Carlo k Q calculations which show a much smaller difference between FFF and flattened beams than originally predicted. These results are explained by the inclusion of the added backscatter plates and build-up filters used in modern clinical FFF linacs, compared to a Monte Carlo model of an FFF linac in which the flattening filter is removed and no additional build-up or backscatter plate is added.
- Is Part Of:
- Physics in medicine & biology. Volume 61:Number 14(2016:Jul.)
- Journal:
- Physics in medicine & biology
- Issue:
- Volume 61:Number 14(2016:Jul.)
- Issue Display:
- Volume 61, Issue 14 (2016)
- Year:
- 2016
- Volume:
- 61
- Issue:
- 14
- Issue Sort Value:
- 2016-0061-0014-0000
- Page Start:
- N362
- Page End:
- N372
- Publication Date:
- 2016-07-01
- Subjects:
- dosimetry -- flattening filter free -- Monte Carlo
Biophysics -- Periodicals
Medical physics -- Periodicals
610.153 - Journal URLs:
- http://ioppublishing.org/ ↗
http://iopscience.iop.org/0031-9155 ↗ - DOI:
- 10.1088/0031-9155/61/14/N362 ↗
- Languages:
- English
- ISSNs:
- 0031-9155
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 9189.xml