Comparative Ex Vivo Evaluation of Two Electronic Percussive Testing Devices Measuring the Stability of Dental Implants. Issue 12 (1st December 2014)
- Record Type:
- Journal Article
- Title:
- Comparative Ex Vivo Evaluation of Two Electronic Percussive Testing Devices Measuring the Stability of Dental Implants. Issue 12 (1st December 2014)
- Main Title:
- Comparative Ex Vivo Evaluation of Two Electronic Percussive Testing Devices Measuring the Stability of Dental Implants
- Authors:
- Geckili, Onur
Bilhan, Hakan
Cilingir, Altug
Bilmenoglu, Caglar
Ates, Gokcen
Urgun, Aliye Ceren
Bural, Canan - Abstract:
- Abstract : Background: A comparative ex vivo study was performed to determine electronic percussive test values (PTVs) measured by cabled and wireless electronic percussive testing (EPT) devices and to evaluate the intra‐ and interobserver reliability of the wireless EPT device. Methods: Forty implants were inserted into the vertebrae and forty into the pelvis of a steer, a safe distance apart. The implants were all 4.3 mm wide and 13 mm long, from the same manufacturer. PTV of each implant was measured by four different examiners, using both EPT devices, and compared. Additionally, the intra‐ and interobserver reliability of the wireless EPT device was evaluated. Results: Statistically significant differences ( P <0.05) were observed between PTVs made by the two EPT devices. PTVs measured by the wireless EPT device were significantly higher than the cabled EPT device ( P <0.05), indicating lower implant stability. The intraobserver reliability of the wireless EPT device was evaluated as excellent for the measurements in type II bone and good‐to‐excellent in type IV bone; interobserver reliability was evaluated as fair‐to‐good in both bone types. Conclusion: The wireless EPT device gives PTVs higher than the cabled EPT device, indicating lower implant stability, and its inter‐ and intraobserver reliability is good and acceptable.
- Is Part Of:
- Journal of periodontology. Volume 85:Issue 12(2014)
- Journal:
- Journal of periodontology
- Issue:
- Volume 85:Issue 12(2014)
- Issue Display:
- Volume 85, Issue 12 (2014)
- Year:
- 2014
- Volume:
- 85
- Issue:
- 12
- Issue Sort Value:
- 2014-0085-0012-0000
- Page Start:
- 1786
- Page End:
- 1791
- Publication Date:
- 2014-12-01
- Subjects:
- Dental implants -- dentistry -- jaw, edentulous -- osseointegration -- reproducibility of results -- research design
Periodontics -- Periodicals
617.632 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1902/(ISSN)1943-3670 ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1902/jop.2014.140206 ↗
- Languages:
- English
- ISSNs:
- 0022-3492
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5030.700000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9112.xml