Ultra-low dose quantitative CT myocardial perfusion imaging with sparse-view dynamic acquisition and image reconstruction: A feasibility study. (1st March 2018)
- Record Type:
- Journal Article
- Title:
- Ultra-low dose quantitative CT myocardial perfusion imaging with sparse-view dynamic acquisition and image reconstruction: A feasibility study. (1st March 2018)
- Main Title:
- Ultra-low dose quantitative CT myocardial perfusion imaging with sparse-view dynamic acquisition and image reconstruction: A feasibility study
- Authors:
- Enjilela, Esmaeil
Lee, Ting-Yim
Hsieh, Jiang
Wisenberg, Gerald
Teefy, Patrick
Yadegari, Andrew
Bagur, Rodrigo
Islam, Ali
Branch, Kelley
So, Aaron - Abstract:
- Abstract: Purpose: We implemented and validated a compressed sensing (CS) based algorithm for reconstructing dynamic contrast-enhanced (DCE) CT images of the heart from sparsely sampled X-ray projections. Methods: DCE CT imaging of the heart was performed on five normal and ischemic pigs after contrast injection. DCE images were reconstructed with filtered backprojection (FBP) and CS from all projections (984-view) and 1/3 of all projections (328-view), and with CS from 1/4 of all projections (246-view). Myocardial perfusion (MP) measurements with each protocol were compared to those with the reference 984-view FBP protocol. Results: Both the 984-view CS and 328-view CS protocols were in good agreements with the reference protocol. The Pearson correlation coefficients of 984-view CS and 328-view CS determined from linear regression analyses were 0.98 and 0.99 respectively. The corresponding mean biases of MP measurement determined from Bland-Altman analyses were 2.7 and 1.2 ml/min/100 g. When only 328 projections were used for image reconstruction, CS was more accurate than FBP for MP measurement with respect to 984-view FBP. However, CS failed to generate MP maps comparable to those with 984-view FBP when only 246 projections were used for image reconstruction. Conclusion: DCE heart images reconstructed from one-third of a full projection set with CS were minimally affected by aliasing artifacts, leading to accurate MP measurements with the effective dose reduced to justAbstract: Purpose: We implemented and validated a compressed sensing (CS) based algorithm for reconstructing dynamic contrast-enhanced (DCE) CT images of the heart from sparsely sampled X-ray projections. Methods: DCE CT imaging of the heart was performed on five normal and ischemic pigs after contrast injection. DCE images were reconstructed with filtered backprojection (FBP) and CS from all projections (984-view) and 1/3 of all projections (328-view), and with CS from 1/4 of all projections (246-view). Myocardial perfusion (MP) measurements with each protocol were compared to those with the reference 984-view FBP protocol. Results: Both the 984-view CS and 328-view CS protocols were in good agreements with the reference protocol. The Pearson correlation coefficients of 984-view CS and 328-view CS determined from linear regression analyses were 0.98 and 0.99 respectively. The corresponding mean biases of MP measurement determined from Bland-Altman analyses were 2.7 and 1.2 ml/min/100 g. When only 328 projections were used for image reconstruction, CS was more accurate than FBP for MP measurement with respect to 984-view FBP. However, CS failed to generate MP maps comparable to those with 984-view FBP when only 246 projections were used for image reconstruction. Conclusion: DCE heart images reconstructed from one-third of a full projection set with CS were minimally affected by aliasing artifacts, leading to accurate MP measurements with the effective dose reduced to just 33% of conventional full-view FBP method. The proposed CS sparse-view image reconstruction method could facilitate the implementation of sparse-view dynamic acquisition for ultra-low dose CT MP imaging. … (more)
- Is Part Of:
- International journal of cardiology. Volume 254(2018)
- Journal:
- International journal of cardiology
- Issue:
- Volume 254(2018)
- Issue Display:
- Volume 254, Issue 2018 (2018)
- Year:
- 2018
- Volume:
- 254
- Issue:
- 2018
- Issue Sort Value:
- 2018-0254-2018-0000
- Page Start:
- 272
- Page End:
- 281
- Publication Date:
- 2018-03-01
- Subjects:
- Quantitative CT myocardial perfusion imaging -- Sparse-view image reconstruction -- Compressed sensing -- Total variation minimization -- Radiation dose reduction
Cardiology -- Periodicals
Electronic journals
616.12 - Journal URLs:
- http://www.clinicalkey.com/dura/browse/journalIssue/01675273 ↗
http://www.sciencedirect.com/science/journal/01675273 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.ijcard.2017.11.030 ↗
- Languages:
- English
- ISSNs:
- 0167-5273
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4542.158000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 9100.xml