Interlaboratory variability in the measurement of direct oral anticoagulants: results from the external quality assessment scheme. (29th January 2018)
- Record Type:
- Journal Article
- Title:
- Interlaboratory variability in the measurement of direct oral anticoagulants: results from the external quality assessment scheme. (29th January 2018)
- Main Title:
- Interlaboratory variability in the measurement of direct oral anticoagulants: results from the external quality assessment scheme
- Authors:
- Tripodi, A.
Chantarangkul, V.
Legnani, C.
Testa, S.
Tosetto, A. - Abstract:
- Abstract : Essentials Tests for direct oral anticoagulants (DOACs) are not widely applied. These tests are perceived to be difficult to run and subjected to large between‐lab variation. We carried out proficiency testing surveys for DOAC testing in Italy. Interlab variability was small and similar to that of the international normalised ratio. Summary: Background: Tests for direct oral anticoagulants (DOACs) are not widely available. The perception that they are difficult to perform and are subject to large between‐laboratory variation makes their implementation difficult. Aims: We carried out proficiency‐testing surveys for DOACs within the activity of the external quality‐assessment scheme of the Italian Federation of Thrombosis Centers. Design: Participants were provided with coded freeze‐dried plasmas without or with graded concentrations of the three main DOACs, and asked to measure prothrombin time (PT), activated partial thromboplastin time (APTT), thrombin time and DOAC concentrations with dedicated tests. The results were centralized for statistical analysis. Results and conclusions: All participants ( n = 235) reported results for PT and APTT, and approximately one‐third reported results for DOAC concentration. PT and APTT showed variable responsiveness to DOACs: PT was more responsive to rivaroxaban than to dabigatran or apixaban. APTT was more responsive to dabigatran than to rivaroxaban or apixaban. The thrombin time ratio (test/normal) was close to unity forAbstract : Essentials Tests for direct oral anticoagulants (DOACs) are not widely applied. These tests are perceived to be difficult to run and subjected to large between‐lab variation. We carried out proficiency testing surveys for DOAC testing in Italy. Interlab variability was small and similar to that of the international normalised ratio. Summary: Background: Tests for direct oral anticoagulants (DOACs) are not widely available. The perception that they are difficult to perform and are subject to large between‐laboratory variation makes their implementation difficult. Aims: We carried out proficiency‐testing surveys for DOACs within the activity of the external quality‐assessment scheme of the Italian Federation of Thrombosis Centers. Design: Participants were provided with coded freeze‐dried plasmas without or with graded concentrations of the three main DOACs, and asked to measure prothrombin time (PT), activated partial thromboplastin time (APTT), thrombin time and DOAC concentrations with dedicated tests. The results were centralized for statistical analysis. Results and conclusions: All participants ( n = 235) reported results for PT and APTT, and approximately one‐third reported results for DOAC concentration. PT and APTT showed variable responsiveness to DOACs: PT was more responsive to rivaroxaban than to dabigatran or apixaban. APTT was more responsive to dabigatran than to rivaroxaban or apixaban. The thrombin time ratio (test/normal) was close to unity for plasmas without dabigatran, and was high (i.e. 7.6‐fold or 15.4‐fold longer than the plasma free from the drug) for plasmas containing dabigatran at low (i.e. 42 ng mL −1 ) or high (i.e. 182 ng mL −1 ) concentration. Dedicated tests were responsive to the respective drugs, and their interlaboratory variability was relatively small (overall coefficients of variation of 8.7%, 8.4% or 10.3% for dabigatran, rivaroxaban and apixaban, respectively) and was comparable to that observed within the same survey for the International Normalized Ratio (i.e. 11.4%). In conclusion, tests for DOAC measurement performed reasonably well in a national quality‐control scheme. Regulatory authorities should urgently issue recommendations on their use, and clinical laboratories should make them available. … (more)
- Is Part Of:
- Journal of thrombosis and haemostasis. Volume 16:Number 3(2018)
- Journal:
- Journal of thrombosis and haemostasis
- Issue:
- Volume 16:Number 3(2018)
- Issue Display:
- Volume 16, Issue 3 (2018)
- Year:
- 2018
- Volume:
- 16
- Issue:
- 3
- Issue Sort Value:
- 2018-0016-0003-0000
- Page Start:
- 565
- Page End:
- 570
- Publication Date:
- 2018-01-29
- Subjects:
- anticoagulants -- partial thromboplastin time -- prothrombin time -- quality control -- thrombin time
Thrombosis -- Periodicals
Hemostasis -- Periodicals
Blood coagulation disorders -- Periodicals
616.1 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1538-7836 ↗
http://www.blackwellpublishing.com/journals/jth ↗
https://www.sciencedirect.com/journal/journal-of-thrombosis-and-haemostasis ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1111/jth.13949 ↗
- Languages:
- English
- ISSNs:
- 1538-7933
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5069.345000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 8993.xml