The impact of cine EPID image acquisition frame rate on markerless soft‐tissue tracking. Issue 6 (5th May 2014)
- Record Type:
- Journal Article
- Title:
- The impact of cine EPID image acquisition frame rate on markerless soft‐tissue tracking. Issue 6 (5th May 2014)
- Main Title:
- The impact of cine EPID image acquisition frame rate on markerless soft‐tissue tracking
- Authors:
- Yip, Stephen
Rottmann, Joerg
Berbeco, Ross - Abstract:
- Abstract : Purpose: Although reduction of the cine electronic portal imaging device (EPID) acquisition frame rate through multiple frame averaging may reduce hardware memory burden and decrease image noise, it can hinder the continuity of soft‐tissue motion leading to poor autotracking results. The impact of motion blurring and image noise on the tracking performance was investigated. Methods: Phantom and patient images were acquired at a frame rate of 12.87 Hz with an amorphous silicon portal imager (AS1000, Varian Medical Systems, Palo Alto, CA). The maximum frame rate of 12.87 Hz is imposed by the EPID. Low frame rate images were obtained by continuous frame averaging. A previously validated tracking algorithm was employed for autotracking. The difference between the programmed and autotracked positions of a Las Vegas phantom moving in the superior‐inferior direction defined the tracking error ( δ ). Motion blurring was assessed by measuring the area change of the circle with the greatest depth. Additionally, lung tumors on 1747 frames acquired at 11 field angles from four radiotherapy patients are manually and automatically tracked with varying frame averaging. δ was defined by the position difference of the two tracking methods. Image noise was defined as the standard deviation of the background intensity. Motion blurring and image noise are correlated with δ using Pearson correlation coefficient (R). Results: For both phantom and patient studies, the autotrackingAbstract : Purpose: Although reduction of the cine electronic portal imaging device (EPID) acquisition frame rate through multiple frame averaging may reduce hardware memory burden and decrease image noise, it can hinder the continuity of soft‐tissue motion leading to poor autotracking results. The impact of motion blurring and image noise on the tracking performance was investigated. Methods: Phantom and patient images were acquired at a frame rate of 12.87 Hz with an amorphous silicon portal imager (AS1000, Varian Medical Systems, Palo Alto, CA). The maximum frame rate of 12.87 Hz is imposed by the EPID. Low frame rate images were obtained by continuous frame averaging. A previously validated tracking algorithm was employed for autotracking. The difference between the programmed and autotracked positions of a Las Vegas phantom moving in the superior‐inferior direction defined the tracking error ( δ ). Motion blurring was assessed by measuring the area change of the circle with the greatest depth. Additionally, lung tumors on 1747 frames acquired at 11 field angles from four radiotherapy patients are manually and automatically tracked with varying frame averaging. δ was defined by the position difference of the two tracking methods. Image noise was defined as the standard deviation of the background intensity. Motion blurring and image noise are correlated with δ using Pearson correlation coefficient (R). Results: For both phantom and patient studies, the autotracking errors increased at frame rates lower than 4.29 Hz. Above 4.29 Hz, changes in errors were negligible with δ < 1.60 mm. Motion blurring and image noise were observed to increase and decrease with frame averaging, respectively. Motion blurring and tracking errors were significantly correlated for the phantom (R = 0.94) and patient studies (R = 0.72). Moderate to poor correlation was found between image noise and tracking error with R −0.58 and −0.19 for both studies, respectively. Conclusions: Cine EPID image acquisition at the frame rate of at least 4.29 Hz is recommended. Motion blurring in the images with frame rates below 4.29 Hz can significantly reduce the accuracy of autotracking. … (more)
- Is Part Of:
- Medical physics. Volume 41:Issue 6(2014)Part 1
- Journal:
- Medical physics
- Issue:
- Volume 41:Issue 6(2014)Part 1
- Issue Display:
- Volume 41, Issue 6, Part 1 (2014)
- Year:
- 2014
- Volume:
- 41
- Issue:
- 6
- Part:
- 1
- Issue Sort Value:
- 2014-0041-0006-0001
- Page Start:
- n/a
- Page End:
- n/a
- Publication Date:
- 2014-05-05
- Subjects:
- Therapeutic applications, including brachytherapy -- Noise
correlation methods -- image denoising -- image restoration -- medical image processing -- phantoms -- radiation therapy
EPID -- markerless tracking -- acquisition frame rate stereotactic body radiotherapy
Radiation therapy -- Digital computing or data processing equipment or methods, specially adapted for specific applications -- Image data processing or generation, in general -- Image enhancement or restoration, e.g. from bit‐mapped to bit‐mapped creating a similar image
Image guided radiation therapy -- Cancer -- Medical image contrast -- Digital image processing -- Lungs -- Dosimetry -- Medical image segmentation
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1118/1.4873322 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 8933.xml