Quality control for multiple breath washout tests in multicentre bronchiectasis studies: Experiences from the BRONCH-UK clinimetrics study. (December 2018)
- Record Type:
- Journal Article
- Title:
- Quality control for multiple breath washout tests in multicentre bronchiectasis studies: Experiences from the BRONCH-UK clinimetrics study. (December 2018)
- Main Title:
- Quality control for multiple breath washout tests in multicentre bronchiectasis studies: Experiences from the BRONCH-UK clinimetrics study
- Authors:
- O'Neill, Katherine
Lakshmipathy, Gokul R.
Ferguson, Kathryn
Cosgrove, Denise
Hill, A.T.
Loebinger, Michael R.
Carroll, Mary
Chalmers, J.D.
Gatheral, Timothy
Johnson, Chris
DeSoyza, Anthony
Hurst, John R.
Bradbury, Ian
Elborn, J.S.
Bradley, Judy M. - Abstract:
- Abstract: Introduction: Multiple Breath Washout (MBW) to measure Lung Clearance Index (LCI) is increasingly being used as a secondary endpoint in multicentre bronchiectasis studies. LCI data quality control or "over-reading" is resource intensive and the impact is unclear. Objectives: To assess the proportion of MBW tests deemed unacceptable with over-reading, and to assess the change in LCI (number of turnovers), LCI coefficient of variation (CV%) and tidal volume (VT) CV% results after over-reading. Methods: Data were analysed from 250 MBW tests (from 98 adult bronchiectasis patients) collected as part of the Bronch-UK Clinimetrics study in 5 UK centres. Each MBW test was over-read centrally using pre-defined criteria. MBW tests with <2 technically valid and repeatable trials were deemed unacceptable to include in analysis. In accepted tests, values for LCI, LCI CV% and VT CV% before and after over-reading, were compared. Results: Insufficient data was collected in 10/250 tests. With over-reading, 30/240 (12%) were deemed unacceptable to include in analysis. In those accepted tests, overall the change in LCI, LCI CV% and VT CV% with over-reading was not statistically significant. When MBW new sites were compared to MBW expert sites, the change in LCI with over-reading was significantly greater in MBW new sites (p = 0.047). Data suggests that over-reading could be important up to at least 12 months post initiation of MBW activity. Conclusion: MBW over-reading was importantAbstract: Introduction: Multiple Breath Washout (MBW) to measure Lung Clearance Index (LCI) is increasingly being used as a secondary endpoint in multicentre bronchiectasis studies. LCI data quality control or "over-reading" is resource intensive and the impact is unclear. Objectives: To assess the proportion of MBW tests deemed unacceptable with over-reading, and to assess the change in LCI (number of turnovers), LCI coefficient of variation (CV%) and tidal volume (VT) CV% results after over-reading. Methods: Data were analysed from 250 MBW tests (from 98 adult bronchiectasis patients) collected as part of the Bronch-UK Clinimetrics study in 5 UK centres. Each MBW test was over-read centrally using pre-defined criteria. MBW tests with <2 technically valid and repeatable trials were deemed unacceptable to include in analysis. In accepted tests, values for LCI, LCI CV% and VT CV% before and after over-reading, were compared. Results: Insufficient data was collected in 10/250 tests. With over-reading, 30/240 (12%) were deemed unacceptable to include in analysis. In those accepted tests, overall the change in LCI, LCI CV% and VT CV% with over-reading was not statistically significant. When MBW new sites were compared to MBW expert sites, the change in LCI with over-reading was significantly greater in MBW new sites (p = 0.047). Data suggests that over-reading could be important up to at least 12 months post initiation of MBW activity. Conclusion: MBW over-reading was important in this study as 12% of tests were considered unacceptable. Over-reading improved test result accuracy in sites new to MBW. Highlights: Lung Clearance Index is emerging as a potential endpoint in bronchiectasis. Data quality control or "over-reading" avoids the inclusion of invalid data. MBW experience influences result accuracy. … (more)
- Is Part Of:
- Respiratory medicine. Volume 145(2018)
- Journal:
- Respiratory medicine
- Issue:
- Volume 145(2018)
- Issue Display:
- Volume 145, Issue 2018 (2018)
- Year:
- 2018
- Volume:
- 145
- Issue:
- 2018
- Issue Sort Value:
- 2018-0145-2018-0000
- Page Start:
- 206
- Page End:
- 211
- Publication Date:
- 2018-12
- Subjects:
- Bronchiectasis -- Multiple breath washout -- Lung clearance index -- Over-reading -- Quality control
Chest -- Diseases -- Periodicals
Chest -- Diseases -- Great Britain -- Periodicals
Respiratory organs -- Diseases -- Periodicals
Respiratory Tract Diseases -- Periodicals
Appareil respiratoire -- Maladies -- Périodiques
Thorax -- Maladies -- Périodiques
Appareil respiratoire -- Maladies -- Traitement -- Périodiques
Electronic journals
616.2 - Journal URLs:
- http://www.sciencedirect.com/science/journal/09546111 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/09546111 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/09546111 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.rmed.2018.10.030 ↗
- Languages:
- English
- ISSNs:
- 0954-6111
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 7777.661900
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 8931.xml