How Much Stereoscopic Effect Does Laparoscopy Need? Controlled, Prospective Randomized Trial on Surgical Task Efficiency in Standardized Phantom Tasks. (October 2018)
- Record Type:
- Journal Article
- Title:
- How Much Stereoscopic Effect Does Laparoscopy Need? Controlled, Prospective Randomized Trial on Surgical Task Efficiency in Standardized Phantom Tasks. (October 2018)
- Main Title:
- How Much Stereoscopic Effect Does Laparoscopy Need? Controlled, Prospective Randomized Trial on Surgical Task Efficiency in Standardized Phantom Tasks
- Authors:
- Kunert, Wolfgang
Auer, Thomas
Storz, Pirmin
Braun, Manuel
Kirschniak, Andreas
Falch, Claudius - Abstract:
- Background. To regain 2-eyed vision in laparoscopy, dual-channeled optics have been introduced. With this optics design, the distance between the 2 front lenses defines how much stereoscopic effect is seen. This study quantifies the impact of an enhanced and a reduced stereo effect on surgical task efficiency. Methods. A prospective single-blinded study was performed with 20 laparoscopic novices in an inanimate experimental setting. A standard bichannelled stereo system was used to perform a suturing and knotting task. The working distance and the task size were scaled to vary the stereo effect and, thereby, simulate hypothetic stereo optics with enhanced and reduced optical bases. The task performances were timed, and the number of trials for stitching out was counted. The participants finally filled out a questionnaire to collect subjective impressions. Results. The increase of the stereo effect by 50% caused no objective improvement in laparoscopic knotting compared with typical 3D (control group with stereo basis of 4.5 mm). But ergonomic disadvantages (headache) were subjectively reported in 1 of 20 cases in the questionnaire. The reduction of the stereo effect by one-third led to a significantly longer average execution time. There was no significant dependence found between stereo effect and number of stich-out trials, stitching precision, or knotting quality. Conclusions. Considering laparoscopy, it does not seem advisable to enhance the stereo effect because ofBackground. To regain 2-eyed vision in laparoscopy, dual-channeled optics have been introduced. With this optics design, the distance between the 2 front lenses defines how much stereoscopic effect is seen. This study quantifies the impact of an enhanced and a reduced stereo effect on surgical task efficiency. Methods. A prospective single-blinded study was performed with 20 laparoscopic novices in an inanimate experimental setting. A standard bichannelled stereo system was used to perform a suturing and knotting task. The working distance and the task size were scaled to vary the stereo effect and, thereby, simulate hypothetic stereo optics with enhanced and reduced optical bases. The task performances were timed, and the number of trials for stitching out was counted. The participants finally filled out a questionnaire to collect subjective impressions. Results. The increase of the stereo effect by 50% caused no objective improvement in laparoscopic knotting compared with typical 3D (control group with stereo basis of 4.5 mm). But ergonomic disadvantages (headache) were subjectively reported in 1 of 20 cases in the questionnaire. The reduction of the stereo effect by one-third led to a significantly longer average execution time. There was no significant dependence found between stereo effect and number of stich-out trials, stitching precision, or knotting quality. Conclusions. Considering laparoscopy, it does not seem advisable to enhance the stereo effect because of ergonomic problems. Otherwise, a miniaturization of the 3D scope (5 mm version) is problematic because its benefit mostly shrinks with the reduced stereo effect. … (more)
- Is Part Of:
- Surgical innovation. Volume 25:Number 5(2018)
- Journal:
- Surgical innovation
- Issue:
- Volume 25:Number 5(2018)
- Issue Display:
- Volume 25, Issue 5 (2018)
- Year:
- 2018
- Volume:
- 25
- Issue:
- 5
- Issue Sort Value:
- 2018-0025-0005-0000
- Page Start:
- 515
- Page End:
- 524
- Publication Date:
- 2018-10
- Subjects:
- biomedical engineering -- ergonomics and/or human factors study -- robotic surgery -- laparoscopic surgery -- three-dimensional imaging -- stereoscopy -- stereoscopic basis -- randomized controlled trial -- experimental setup
Surgery, Operative -- Periodicals
Endoscopic surgery -- Periodicals
Laparoscopic surgery -- Periodicals
Surgical Procedures, Operative -- Periodicals
Surgical Procedures, Minimally Invasive -- Periodicals
Diffusion of Innovation -- Periodicals
Chirurgie opératoire -- Périodiques
Chirurgie endoscopique -- Périodiques
Chirurgie laparoscopique -- Périodiques
617.91 - Journal URLs:
- http://journals.sagepub.com/home/sri ↗
http://sri.sagepub.com/ ↗
http://www.sagepub.com/journalsProdDesc.nav?prodId=Journal201793 ↗
http://www.sagepublications.com/ ↗ - DOI:
- 10.1177/1553350618784801 ↗
- Languages:
- English
- ISSNs:
- 1553-3506
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 8819.xml