Clinical yield of magnetoencephalography distributed source imaging in epilepsy: A comparison with equivalent current dipole method. Issue 1 (11th October 2017)
- Record Type:
- Journal Article
- Title:
- Clinical yield of magnetoencephalography distributed source imaging in epilepsy: A comparison with equivalent current dipole method. Issue 1 (11th October 2017)
- Main Title:
- Clinical yield of magnetoencephalography distributed source imaging in epilepsy: A comparison with equivalent current dipole method
- Authors:
- Pellegrino, Giovanni
Hedrich, Tanguy
Chowdhury, Rasheda Arman
Hall, Jeffery A.
Dubeau, Francois
Lina, Jean‐Marc
Kobayashi, Eliane
Grova, Christophe - Abstract:
- Abstract: Objective: Source localization of interictal epileptic discharges (IEDs) is clinically useful in the presurgical workup of epilepsy patients. It is usually obtained by equivalent current dipole (ECD) which localizes a point source and is the only inverse solution approved by clinical guidelines. In contrast, magnetic source imaging using distributed methods (dMSI) provides maps of the location and the extent of the generators, but its yield has not been clinically validated. We systematically compared ECD versus dMSI performed using coherent Maximum Entropy on the Mean (cMEM), a method sensitive to the spatial extent of the generators. Methods: 340 source localizations of IEDs derived from 49 focal epilepsy patients with foci well‐defined through intracranial EEG, MRI lesions, and surgery were analyzed. The comparison was based on the assessment of the sublobar concordance with the focus and of the distance between the source and the focus. Results: dMSI sublobar concordance was significantly higher than ECD (81% vs 69%, P < 0.001), especially for extratemporal lobe sources (dMSI = 84%; ECD = 67%, P < 0.001) and for seizure free patients (dMSI = 83%; ECD = 70%, P < 0.001). The median distance from the focus was 4.88 mm for ECD and 3.44 mm for dMSI ( P < 0.001). ECD dipoles were often wrongly localized in deep brain regions. Conclusions: dMSI using cMEM exhibited better accuracy. dMSI also offered the advantage of recovering more realistic maps of the generator,Abstract: Objective: Source localization of interictal epileptic discharges (IEDs) is clinically useful in the presurgical workup of epilepsy patients. It is usually obtained by equivalent current dipole (ECD) which localizes a point source and is the only inverse solution approved by clinical guidelines. In contrast, magnetic source imaging using distributed methods (dMSI) provides maps of the location and the extent of the generators, but its yield has not been clinically validated. We systematically compared ECD versus dMSI performed using coherent Maximum Entropy on the Mean (cMEM), a method sensitive to the spatial extent of the generators. Methods: 340 source localizations of IEDs derived from 49 focal epilepsy patients with foci well‐defined through intracranial EEG, MRI lesions, and surgery were analyzed. The comparison was based on the assessment of the sublobar concordance with the focus and of the distance between the source and the focus. Results: dMSI sublobar concordance was significantly higher than ECD (81% vs 69%, P < 0.001), especially for extratemporal lobe sources (dMSI = 84%; ECD = 67%, P < 0.001) and for seizure free patients (dMSI = 83%; ECD = 70%, P < 0.001). The median distance from the focus was 4.88 mm for ECD and 3.44 mm for dMSI ( P < 0.001). ECD dipoles were often wrongly localized in deep brain regions. Conclusions: dMSI using cMEM exhibited better accuracy. dMSI also offered the advantage of recovering more realistic maps of the generator, which could be exploited for neuronavigation aimed at targeting invasive EEG and surgical resection. Therefore, dMSI may be preferred to ECD in clinical practice. Hum Brain Mapp 39:218–231, 2018 . ©2017 Wiley Periodicals, Inc. … (more)
- Is Part Of:
- Human brain mapping. Volume 39:Issue 1(2018)
- Journal:
- Human brain mapping
- Issue:
- Volume 39:Issue 1(2018)
- Issue Display:
- Volume 39, Issue 1 (2018)
- Year:
- 2018
- Volume:
- 39
- Issue:
- 1
- Issue Sort Value:
- 2018-0039-0001-0000
- Page Start:
- 218
- Page End:
- 231
- Publication Date:
- 2017-10-11
- Subjects:
- interictal epileptiform discharges -- source localization -- MEG -- presurgical evaluation -- dipole -- magnetic source imaging -- surgery -- distributed source -- spike
Brain mapping -- Periodicals
611.81 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)1097-0193 ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/hbm.23837 ↗
- Languages:
- English
- ISSNs:
- 1065-9471
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4336.031000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 8803.xml