Predictive factors of lead failure in patients implanted with cardiac devices. (15th November 2015)
- Record Type:
- Journal Article
- Title:
- Predictive factors of lead failure in patients implanted with cardiac devices. (15th November 2015)
- Main Title:
- Predictive factors of lead failure in patients implanted with cardiac devices
- Authors:
- Aizawa, Yoshiyasu
Negishi, Masachika
Kashimura, Shin
Nakajima, Kazuaki
Kunitomi, Akira
Katsumata, Yoshinori
Nishiyama, Takahiko
Kimura, Takehiro
Nishiyama, Nobuhiro
Fukumoto, Kotaro
Tanimoto, Yoko
Kohsaka, Shun
Takatsuki, Seiji
Fukuda, Keiichi - Abstract:
- Abstract: Introduction: Lead failures (LFs) are one of the most common complications in patients implanted with cardiovascular implantable electronic devices. LFs often cause serious secondary complications such as inappropriate ICD shocks or asystole. This study aimed to identify the clinical factors associated with the occurrence of LFs. Methods: A total of 735 consecutive device implantations (mean age 67 ± 15 years, males 64%) performed at a single university hospital setting from 1997 to 2014 were included. The implanted devices consisted of 421 pacemakers, 250 implantable cardioverter defibrillators (ICD), 9 cardiac resynchronization therapy pacemakers (CRT-P), and 55 CRT defibrillators (CRT-D). The primary endpoint was the development of an LF. Results: During a mean duration of 5.8 ± 4.3 years, 38 LFs developed in 31 patients (mean age 56 ± 14 years). LFs included 32 ICD (7 Sprint Fidelis, 2 Riata), and 6 pacing leads. Nine patients received inappropriate ICD shocks and 1 had syncope due to an LF. All patients underwent lead reinsertions with device replacements. Eight patients required opposite site implantations due to venous occlusions. The predictive factors of LFs were the age, male sex, taller body length, ICD vs. pacemaker, lesser lead number, extra-thoracic puncture of the axillary vein vs. a cut-down of the cephalic vein, use of recalled leads and patients with idiopathic ventricular fibrillation (IVF) and Brugada syndrome (BrS). Conclusion: LFs occurredAbstract: Introduction: Lead failures (LFs) are one of the most common complications in patients implanted with cardiovascular implantable electronic devices. LFs often cause serious secondary complications such as inappropriate ICD shocks or asystole. This study aimed to identify the clinical factors associated with the occurrence of LFs. Methods: A total of 735 consecutive device implantations (mean age 67 ± 15 years, males 64%) performed at a single university hospital setting from 1997 to 2014 were included. The implanted devices consisted of 421 pacemakers, 250 implantable cardioverter defibrillators (ICD), 9 cardiac resynchronization therapy pacemakers (CRT-P), and 55 CRT defibrillators (CRT-D). The primary endpoint was the development of an LF. Results: During a mean duration of 5.8 ± 4.3 years, 38 LFs developed in 31 patients (mean age 56 ± 14 years). LFs included 32 ICD (7 Sprint Fidelis, 2 Riata), and 6 pacing leads. Nine patients received inappropriate ICD shocks and 1 had syncope due to an LF. All patients underwent lead reinsertions with device replacements. Eight patients required opposite site implantations due to venous occlusions. The predictive factors of LFs were the age, male sex, taller body length, ICD vs. pacemaker, lesser lead number, extra-thoracic puncture of the axillary vein vs. a cut-down of the cephalic vein, use of recalled leads and patients with idiopathic ventricular fibrillation (IVF) and Brugada syndrome (BrS). Conclusion: LFs occurred mainly with ICD leads. A lesser age, the puncture method, lead model, and diagnosis of IVF/BrS were associated with the development of LFs. … (more)
- Is Part Of:
- International journal of cardiology. Volume 199(2015)
- Journal:
- International journal of cardiology
- Issue:
- Volume 199(2015)
- Issue Display:
- Volume 199, Issue 2015 (2015)
- Year:
- 2015
- Volume:
- 199
- Issue:
- 2015
- Issue Sort Value:
- 2015-0199-2015-0000
- Page Start:
- 277
- Page End:
- 281
- Publication Date:
- 2015-11-15
- Subjects:
- Pacemaker -- Implantable cardioverter defibrillator -- Lead failure -- Fidelis -- Riata -- Brugada syndrome -- Idiopathic ventricular fibrillation
Cardiology -- Periodicals
Electronic journals
616.12 - Journal URLs:
- http://www.clinicalkey.com/dura/browse/journalIssue/01675273 ↗
http://www.sciencedirect.com/science/journal/01675273 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.ijcard.2015.07.055 ↗
- Languages:
- English
- ISSNs:
- 0167-5273
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4542.158000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 8783.xml