A multicenter dosimetry study to evaluate the imaging dose from Elekta XVI and Varian OBI kV-CBCT systems to cardiovascular implantable electronic devices (CIEDs). (November 2018)
- Record Type:
- Journal Article
- Title:
- A multicenter dosimetry study to evaluate the imaging dose from Elekta XVI and Varian OBI kV-CBCT systems to cardiovascular implantable electronic devices (CIEDs). (November 2018)
- Main Title:
- A multicenter dosimetry study to evaluate the imaging dose from Elekta XVI and Varian OBI kV-CBCT systems to cardiovascular implantable electronic devices (CIEDs)
- Authors:
- Aslian, Hossein
Delana, Anna
Kaiser, Stefano Ren
Moretti, Eugenia
Foti, Claudio
Bregant, Paola
de Denaro, Mario
Longo, Francesco
Severgnini, Mara - Abstract:
- Highlights: The potential contribution of imaging dose from kV-CBCT to the CIED was evaluated. Three dosimeters were considered in a multi-center study, XRQA2, MOSFET and TLDs. Detailed results of imaging doses to CIED from different CBCT-protocols were found. An approach for more accurate risk classification of CIED patients was suggested. Abstract: The increasing use of daily CBCT in radiotherapy has raised concerns about the additional dose delivered to the patient, and it can also become a concern issue for those patients with cardiovascular implantable electronic devices (CIEDs) (Pacemaker [PM] and Implantable Cardioverter Defibrillator [ICD]). Although guidelines highly recommend that the cumulative dose received by CIEDs should be kept as low as possible, and a safe threshold based on patient risk classification needs to be respected, this additional imaging dose is not usually considered. Four centers with different dosimetry systems and different CBCT imaging protocols participated in this multicenter study to investigate the imaging dose to the CIEDs from Elekta XVI and Varian OBI kV-CBCT systems. It was found that although imaging doses received by CIEDs outside the CBCT field are negligible, special attention should be paid to this value when CIEDs are inside the field because the daily use of CBCT can sometimes contribute considerably to the total dose received by a CIED.
- Is Part Of:
- Physica medica. Volume 55(2018)
- Journal:
- Physica medica
- Issue:
- Volume 55(2018)
- Issue Display:
- Volume 55, Issue 2018 (2018)
- Year:
- 2018
- Volume:
- 55
- Issue:
- 2018
- Issue Sort Value:
- 2018-0055-2018-0000
- Page Start:
- 40
- Page End:
- 46
- Publication Date:
- 2018-11
- Subjects:
- Imaging dose -- Cardiovascular implantable electronic devices -- kV-Cone Beam CT System -- Gafchromic XRQA2 film -- MOSFET -- TLD
Medical physics -- Periodicals
Biophysics -- Periodicals
Biophysics -- Periodicals
Imagerie médicale -- Périodiques
Radiothérapie -- Périodiques
Rayons X -- Sécurité -- Mesures -- Périodiques
Physique -- Périodiques
Médecine -- Périodiques
610.153 - Journal URLs:
- http://www.sciencedirect.com/science/journal/11201797 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/11201797 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/11201797 ↗
http://www.elsevier.com/journals ↗
http://www.physicamedica.com ↗ - DOI:
- 10.1016/j.ejmp.2018.10.013 ↗
- Languages:
- English
- ISSNs:
- 1120-1797
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6475.070000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 8750.xml