Infant Robotic Cleft Palate Surgery: A Feasibility Assessment Using a Realistic Cleft Palate Simulator. Issue 2 (February 2017)
- Record Type:
- Journal Article
- Title:
- Infant Robotic Cleft Palate Surgery: A Feasibility Assessment Using a Realistic Cleft Palate Simulator. Issue 2 (February 2017)
- Main Title:
- Infant Robotic Cleft Palate Surgery
- Authors:
- Podolsky, Dale J.
Fisher, David M.
Wong Riff, Karen W. Y.
Looi, Thomas
Drake, James M.
Forrest, Christopher R. - Abstract:
- Abstract : Background: A surgical robot offers enhanced precision, visualization, and access and the potential to improve outcomes in cleft palate surgery. The goal of this study was to investigate the feasibility of using the da Vinci robot for cleft palate repair in infants using a cleft palate simulator test bed. Methods: A high-fidelity cleft palate simulator was developed that allows performance of a robotic cleft palate repair procedure. A complete cleft palate repair was performed with the da Vinci Si with 5-mm instruments and the da Vinci Xi with 8-mm instruments. The advantages of the robotic approach were assessed in comparison with using standard instruments. For each system, arm repositioning, collisions, instrument and endoscope excursion, wrist orientation, and vision were compared for 12 steps of the repair. Results: The cleft palate simulator provided a reproducible platform for testing robotic cleft palate surgery. The advantages of the robotic approach were the ability to articulate a miniature wrist intraorally with superior visualization, increased ambidexterity, and improved ergonomics compared with using standard instruments. Cleft palate repair with the Xi was superior to the Si with respect to arm repositioning, instrument collisions and excursion, and wrist orientation. However, Xi performance remained suboptimal because of the larger instruments. Conclusions: Robotic cleft palate repair using the da Vinci system offers advantages compared with theAbstract : Background: A surgical robot offers enhanced precision, visualization, and access and the potential to improve outcomes in cleft palate surgery. The goal of this study was to investigate the feasibility of using the da Vinci robot for cleft palate repair in infants using a cleft palate simulator test bed. Methods: A high-fidelity cleft palate simulator was developed that allows performance of a robotic cleft palate repair procedure. A complete cleft palate repair was performed with the da Vinci Si with 5-mm instruments and the da Vinci Xi with 8-mm instruments. The advantages of the robotic approach were assessed in comparison with using standard instruments. For each system, arm repositioning, collisions, instrument and endoscope excursion, wrist orientation, and vision were compared for 12 steps of the repair. Results: The cleft palate simulator provided a reproducible platform for testing robotic cleft palate surgery. The advantages of the robotic approach were the ability to articulate a miniature wrist intraorally with superior visualization, increased ambidexterity, and improved ergonomics compared with using standard instruments. Cleft palate repair with the Xi was superior to the Si with respect to arm repositioning, instrument collisions and excursion, and wrist orientation. However, Xi performance remained suboptimal because of the larger instruments. Conclusions: Robotic cleft palate repair using the da Vinci system offers advantages compared with the traditional approach. Cleft palate repair is more feasible with the Xi and 8-mm instruments. However, performance is limited by the instrumentation, which requires modification to ensure safety and efficacy. CLINICAL QUESTION/LEVEL OF EVIDENCE: Therapeutic, V. Abstract : Supplemental Digital Content is available in the text. … (more)
- Is Part Of:
- Plastic and reconstructive surgery. Volume 139:Issue 2(2017:Feb.)
- Journal:
- Plastic and reconstructive surgery
- Issue:
- Volume 139:Issue 2(2017:Feb.)
- Issue Display:
- Volume 139, Issue 2 (2017)
- Year:
- 2017
- Volume:
- 139
- Issue:
- 2
- Issue Sort Value:
- 2017-0139-0002-0000
- Page Start:
- Page End:
- Publication Date:
- 2017-02
- Subjects:
- Surgery, Plastic -- Periodicals
617.95205 - Journal URLs:
- http://journals.lww.com ↗
- DOI:
- 10.1097/PRS.0000000000003010 ↗
- Languages:
- English
- ISSNs:
- 0032-1052
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6528.924000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 8698.xml