Adjustable-Loop Femoral Cortical Suspensory Fixation for Patellar Tendon Anterior Cruciate Ligament Reconstruction: A Time Zero Biomechanical Comparison With Interference Screw Fixation. Issue 8 (July 2018)
- Record Type:
- Journal Article
- Title:
- Adjustable-Loop Femoral Cortical Suspensory Fixation for Patellar Tendon Anterior Cruciate Ligament Reconstruction: A Time Zero Biomechanical Comparison With Interference Screw Fixation. Issue 8 (July 2018)
- Main Title:
- Adjustable-Loop Femoral Cortical Suspensory Fixation for Patellar Tendon Anterior Cruciate Ligament Reconstruction: A Time Zero Biomechanical Comparison With Interference Screw Fixation
- Authors:
- Mickelson, Dayne T.
Lefebvre, Thomas
Gall, Ken
Riboh, Jonathan C. - Abstract:
- Background: Adjustable-loop cortical buttons for femoral fixation of bone-tendon-bone grafts have potential advantages over interference screw fixation; however, these devices have not been benchmarked biomechanically against interference screws. Purpose/Hypothesis: The purpose was to compare the time zero biomechanical properties of commercially available, adjustable-loop cortical button and metallic interference screws for femoral fixation of bone-tendon-bone grafts. It was hypothesized that no significant differences would be found in biomechanical properties between fixation techniques. Study Design: Controlled laboratory study. Methods: Adjustable-loop cortical buttons (n = 8) and metallic interference screws (n = 8) were used to fix matched pairs of human bone-tendon-bone allografts in porcine distal femurs. These constructs were preconditioned (10 N to 50 N at 1 Hz, 10 cycles), subjected to cyclic loading (50 N to 250 N at 1 Hz, 500 cycles), and then pulled to failure at 20 mm/min. Results: The loads to failure (mean ± SD, 700 ± 256 N vs 688 ± 215 N, P = .92) and linear stiffnesses (219 ± 48 N/mm vs 218 ± 49 N/mm, P = .97) for the adjustable-loop cortical button and metallic interference screws, respectively, were not significantly different. Cyclic displacement was higher in the adjustable-loop cortical button group (2.1 ± 0.6 mm vs 1.3 ± 0.4 mm, P = .01). The mechanism of failure was different between groups, with bone block slippage occurring most commonly in theBackground: Adjustable-loop cortical buttons for femoral fixation of bone-tendon-bone grafts have potential advantages over interference screw fixation; however, these devices have not been benchmarked biomechanically against interference screws. Purpose/Hypothesis: The purpose was to compare the time zero biomechanical properties of commercially available, adjustable-loop cortical button and metallic interference screws for femoral fixation of bone-tendon-bone grafts. It was hypothesized that no significant differences would be found in biomechanical properties between fixation techniques. Study Design: Controlled laboratory study. Methods: Adjustable-loop cortical buttons (n = 8) and metallic interference screws (n = 8) were used to fix matched pairs of human bone-tendon-bone allografts in porcine distal femurs. These constructs were preconditioned (10 N to 50 N at 1 Hz, 10 cycles), subjected to cyclic loading (50 N to 250 N at 1 Hz, 500 cycles), and then pulled to failure at 20 mm/min. Results: The loads to failure (mean ± SD, 700 ± 256 N vs 688 ± 215 N, P = .92) and linear stiffnesses (219 ± 48 N/mm vs 218 ± 49 N/mm, P = .97) for the adjustable-loop cortical button and metallic interference screws, respectively, were not significantly different. Cyclic displacement was higher in the adjustable-loop cortical button group (2.1 ± 0.6 mm vs 1.3 ± 0.4 mm, P = .01). The mechanism of failure was different between groups, with bone block slippage occurring most commonly in the interference screw group (n = 5) and fracture of the bone block through the suture hole occurring most commonly in the adjustable-loop cortical button group (n = 6). Conclusion: Adjustable-loop cortical buttons and interference screws have similar time zero failure loads, although cyclic displacement was higher with the adjustable-loop cortical buttons. The mean difference in displacement was less than 1 mm compared with the interference screw. Clinical Relevance: Adjustable-loop cortical buttons may be an acceptable alternative to an interference screw for femoral fixation of bone-tendon-bone grafts in anterior cruciate ligament reconstruction. The clinical relevance of the observed differences in cyclic displacement is unknown and should be evaluated in future studies. … (more)
- Is Part Of:
- American journal of sports medicine. Volume 46:Issue 8(2018)
- Journal:
- American journal of sports medicine
- Issue:
- Volume 46:Issue 8(2018)
- Issue Display:
- Volume 46, Issue 8 (2018)
- Year:
- 2018
- Volume:
- 46
- Issue:
- 8
- Issue Sort Value:
- 2018-0046-0008-0000
- Page Start:
- 1857
- Page End:
- 1862
- Publication Date:
- 2018-07
- Subjects:
- ACL reconstruction -- bone-tendon-bone graft -- suspensory fixation -- adjustable loop -- interference screw -- biomechanics
Sports medicine -- Periodicals
Sports injuries -- Periodicals
Orthopedic surgery -- Periodicals
617.102705 - Journal URLs:
- http://www.mdconsult.com/public/search?search_type=journal&j_sort=pub_date&j_date_range=1995-current&j_issn=0363-5465 ↗
http://ajs.sagepub.com ↗
http://www.ajsm.org ↗
http://www.sagepub.com ↗
http://firstsearch.oclc.org ↗ - DOI:
- 10.1177/0363546518771365 ↗
- Languages:
- English
- ISSNs:
- 0363-5465
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 8687.xml