Feasibility of a Modified Biopsy Needle for Irreversible Electroporation Ablation and Periprocedural Tissue Sampling. (December 2016)
- Record Type:
- Journal Article
- Title:
- Feasibility of a Modified Biopsy Needle for Irreversible Electroporation Ablation and Periprocedural Tissue Sampling. (December 2016)
- Main Title:
- Feasibility of a Modified Biopsy Needle for Irreversible Electroporation Ablation and Periprocedural Tissue Sampling
- Authors:
- Wimmer, Thomas
Srimathveeravalli, Govindarajan
Silk, Mikhail
Monette, Sebastien
Gutta, Narendra
Maybody, Majid
Erinjery, Joseph P.
Coleman, Jonathan A.
Solomon, Stephen B.
Sofocleous, Constantinos T. - Abstract:
- Objectives: To test the feasibility of modified biopsy needles as probes for irreversible electroporation ablation and periprocedural biopsy. Methods: Core biopsy needles of 16-G/9-cm were customized to serve as experimental ablation probes. Computed tomography-guided percutaneous irreversible electroporation was performed in in vivo porcine kidneys with pairs of experimental (n = 10) or standard probes (n = 10) using a single parameter set (1667 V/cm, ninety 100 µs pulses). Two biopsy samples were taken immediately following ablation using the experimental probes (n = 20). Ablation outcomes were compared using computed tomography, simulation, and histology. Biopsy and necropsy histology were compared. Results: Simulation-suggested ablations with experimental probes were smaller than that with standard electrodes (455.23 vs 543.16 mm 2 ), although both exhibited similar shape. Computed tomography (standard: 556 ± 61 mm 2, experimental: 515 ± 67 mm 2 ; P = .25) and histology (standard: 313 ± 77 mm 2, experimental: 275 ± 75 mm 2 ; P = .29) indicated ablations with experimental probes were not significantly different from the standard. Histopathology indicated similar morphological changes in both groups. Biopsies from the ablation zone yielded at least 1 core with sufficient tissue for analysis (11 of the 20). Conclusions: A combined probe for irreversible electroporation ablation and periprocedural tissue sampling from the ablation zone is feasible. Ablation outcomes areObjectives: To test the feasibility of modified biopsy needles as probes for irreversible electroporation ablation and periprocedural biopsy. Methods: Core biopsy needles of 16-G/9-cm were customized to serve as experimental ablation probes. Computed tomography-guided percutaneous irreversible electroporation was performed in in vivo porcine kidneys with pairs of experimental (n = 10) or standard probes (n = 10) using a single parameter set (1667 V/cm, ninety 100 µs pulses). Two biopsy samples were taken immediately following ablation using the experimental probes (n = 20). Ablation outcomes were compared using computed tomography, simulation, and histology. Biopsy and necropsy histology were compared. Results: Simulation-suggested ablations with experimental probes were smaller than that with standard electrodes (455.23 vs 543.16 mm 2 ), although both exhibited similar shape. Computed tomography (standard: 556 ± 61 mm 2, experimental: 515 ± 67 mm 2 ; P = .25) and histology (standard: 313 ± 77 mm 2, experimental: 275 ± 75 mm 2 ; P = .29) indicated ablations with experimental probes were not significantly different from the standard. Histopathology indicated similar morphological changes in both groups. Biopsies from the ablation zone yielded at least 1 core with sufficient tissue for analysis (11 of the 20). Conclusions: A combined probe for irreversible electroporation ablation and periprocedural tissue sampling from the ablation zone is feasible. Ablation outcomes are comparable to those of standard electrodes. … (more)
- Is Part Of:
- Technology in cancer research & treatment. Volume 15:Number 6(2016:Dec.)
- Journal:
- Technology in cancer research & treatment
- Issue:
- Volume 15:Number 6(2016:Dec.)
- Issue Display:
- Volume 15, Issue 6 (2016)
- Year:
- 2016
- Volume:
- 15
- Issue:
- 6
- Issue Sort Value:
- 2016-0015-0006-0000
- Page Start:
- 749
- Page End:
- 758
- Publication Date:
- 2016-12
- Subjects:
- ablation technique -- irreversible electroporation -- core needle biopsy -- computed tomography -- kidney
Oncology -- Periodicals
Cancer -- Diagnosis -- Periodicals
Cancer -- Treatment -- Technological innovations -- Periodicals
616.994 - Journal URLs:
- http://tct.sagepub.com/ ↗
http://www.tcrt.org ↗
http://www.sagepub.com ↗ - DOI:
- 10.1177/1533034615608739 ↗
- Languages:
- English
- ISSNs:
- 1533-0346
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 8652.xml