A novel curve fitting method for AV optimisation of biventricular pacemakers. (6th August 2015)
- Record Type:
- Journal Article
- Title:
- A novel curve fitting method for AV optimisation of biventricular pacemakers. (6th August 2015)
- Main Title:
- A novel curve fitting method for AV optimisation of biventricular pacemakers
- Authors:
- Dehbi, Hakim-Moulay
Jones, Siana
Sohaib, S M Afzal
Finegold, Judith A
Siggers, Jennifer H
Stegemann, Berthold
Whinnett, Zachary I
Francis, Darrel P - Abstract:
- Abstract: In this study, we designed and tested a new algorithm, which we call the 'restricted parabola', to identify the optimum atrioventricular (AV) delay in patients with biventricular pacemakers. This algorithm automatically restricts the hemodynamic data used for curve fitting to the parabolic zone in order to avoid inadvertently selecting an AV optimum that is too long. We used R, a programming language and software environment for statistical computing, to create an algorithm which applies multiple different cut-offs to partition curve fitting of a dataset into a parabolic and a plateau region and then selects the best cut-off using a least squares method. In 82 patients, AV delay was adjusted and beat-to-beat systolic blood pressure (SBP) was measured non-invasively using our multiple-repetition protocol. The novel algorithm was compared to fitting a parabola across the whole dataset to identify how many patients had a plateau region, and whether a higher hemodynamic response was achieved with one method. In 9/82 patients, the restricted parabola algorithm detected that the pattern was not parabolic at longer AV delays. For these patients, the optimal AV delay predicted by the restricted parabola algorithm increased SBP by 1.36 mmHg above that predicted by the conventional parabolic algorithm (95% confidence interval: 0.65 to 2.07 mmHg, p -value = 0.002). AV optima selected using our novel restricted parabola algorithm give a greater improvement in acuteAbstract: In this study, we designed and tested a new algorithm, which we call the 'restricted parabola', to identify the optimum atrioventricular (AV) delay in patients with biventricular pacemakers. This algorithm automatically restricts the hemodynamic data used for curve fitting to the parabolic zone in order to avoid inadvertently selecting an AV optimum that is too long. We used R, a programming language and software environment for statistical computing, to create an algorithm which applies multiple different cut-offs to partition curve fitting of a dataset into a parabolic and a plateau region and then selects the best cut-off using a least squares method. In 82 patients, AV delay was adjusted and beat-to-beat systolic blood pressure (SBP) was measured non-invasively using our multiple-repetition protocol. The novel algorithm was compared to fitting a parabola across the whole dataset to identify how many patients had a plateau region, and whether a higher hemodynamic response was achieved with one method. In 9/82 patients, the restricted parabola algorithm detected that the pattern was not parabolic at longer AV delays. For these patients, the optimal AV delay predicted by the restricted parabola algorithm increased SBP by 1.36 mmHg above that predicted by the conventional parabolic algorithm (95% confidence interval: 0.65 to 2.07 mmHg, p -value = 0.002). AV optima selected using our novel restricted parabola algorithm give a greater improvement in acute hemodynamics than fitting a parabola across all tested AV delays. Such an algorithm may assist the development of automated methods for biventricular pacemaker optimisation. … (more)
- Is Part Of:
- Physiological measurement. Volume 36:Number 9(2015:Sep.)
- Journal:
- Physiological measurement
- Issue:
- Volume 36:Number 9(2015:Sep.)
- Issue Display:
- Volume 36, Issue 9 (2015)
- Year:
- 2015
- Volume:
- 36
- Issue:
- 9
- Issue Sort Value:
- 2015-0036-0009-0000
- Page Start:
- 1889
- Page End:
- 1900
- Publication Date:
- 2015-08-06
- Subjects:
- AV optimisation -- biventricular pacemaker -- cardiac resynchronization therapy (CRT)
Physiology -- Measurement -- Periodicals
Patient monitoring -- Periodicals
612 - Journal URLs:
- http://ioppublishing.org/ ↗
http://iopscience.iop.org/0967-3334 ↗ - DOI:
- 10.1088/0967-3334/36/9/1889 ↗
- Languages:
- English
- ISSNs:
- 0967-3334
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 8540.xml