A combined experimental and finite element approach to analyse the fretting mechanism of the head–stem taper junction in total hip replacement. (September 2017)
- Record Type:
- Journal Article
- Title:
- A combined experimental and finite element approach to analyse the fretting mechanism of the head–stem taper junction in total hip replacement. (September 2017)
- Main Title:
- A combined experimental and finite element approach to analyse the fretting mechanism of the head–stem taper junction in total hip replacement
- Authors:
- Bitter, Thom
Khan, Imran
Marriott, Tim
Lovelady, Elaine
Verdonschot, Nico
Janssen, Dennis - Abstract:
- Fretting corrosion at the taper interface of modular hip implants has been implicated as a possible cause of implant failure. This study was set up to gain more insight in the taper mechanics that lead to fretting corrosion. The objectives of this study therefore were (1) to select experimental loading conditions to reproduce clinically relevant fretting corrosion features observed in retrieved components, (2) to develop a finite element model consistent with the fretting experiments and (3) to apply more complicated loading conditions of activities of daily living to the finite element model to study the taper mechanics. The experiments showed similar wear patterns on the taper surface as observed in retrievals. The finite element wear score based on Archard's law did not correlate well with the amount of material loss measured in the experiments. However, similar patterns were observed between the simulated micromotions and the experimental wear measurements. Although the finite element model could not be validated, the loading conditions based on activities of daily living demonstrate the importance of assembly load on the wear potential. These findings suggest that finite element models that do not incorporate geometry updates to account for wear loss may not be appropriate to predict wear volumes of taper connections.
- Is Part Of:
- Proceedings of the Institution of Mechanical Engineers. Volume 231:Number 9(2017:Sep.)
- Journal:
- Proceedings of the Institution of Mechanical Engineers
- Issue:
- Volume 231:Number 9(2017:Sep.)
- Issue Display:
- Volume 231, Issue 9 (2017)
- Year:
- 2017
- Volume:
- 231
- Issue:
- 9
- Issue Sort Value:
- 2017-0231-0009-0000
- Page Start:
- 862
- Page End:
- 870
- Publication Date:
- 2017-09
- Subjects:
- Fretting -- finite element modelling -- total hip arthroplasty -- taper junction -- wear testing -- wear modelling
Biomedical engineering -- Periodicals
Medical instruments and apparatus -- Periodicals
610.28 - Journal URLs:
- http://pih.sagepub.com/ ↗
http://journals.pepublishing.com/content/119779 ↗
http://www.uk.sagepub.com/home.nav ↗ - DOI:
- 10.1177/0954411917713774 ↗
- Languages:
- English
- ISSNs:
- 0954-4119
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 8529.xml