Clinical validation of a real-time applicator position monitoring system for gynecologic intracavitary brachytherapy. (27th July 2016)
- Record Type:
- Journal Article
- Title:
- Clinical validation of a real-time applicator position monitoring system for gynecologic intracavitary brachytherapy. (27th July 2016)
- Main Title:
- Clinical validation of a real-time applicator position monitoring system for gynecologic intracavitary brachytherapy
- Authors:
- Bou-Zeid, Wassim
Bauer, Christian
Kim, Yusung
Beichel, Reinhard R
Sun, Wenqing
Waldron, Timothy
Xia, Junyi - Abstract:
- Abstract: Purpose: To validate the clinical feasibility and efficacy of a real-time applicator position monitoring system (RAPS) through a phantom study and a prospective clinical trial. Methods and materials: The RAPS measures the brachytherapy applicator displacement in real-time by computing the relative displacement between two infrared reflective targets, one attached to the applicator and the other to the patient's skin. A phantom study was performed to compare RAPS measurements with the ground truth. Six cervical cancer patients were enrolled in the clinical trial using MRI-based high-dose-rate brachytherapy with a Tandem-and-Ovoids applicator. The results from the RAPS are compared with the clinical method. Results: In the phantom study, an average difference between RAPS measurements and known displacements was 0.02 ± 0.01 mm in the superior-inferior direction, 0.02 ± 0.02 mm in the lateral direction, and 0.11 ± 0.06 mm in the anterior-posterior direction. In the clinical trial, the absolute difference in applicator displacement between the RAPS and the clinical method was 1.46 ± 1.13 mm. In all patient cases, a maximum applicator displacement of 6.66 mm (2.0 ± 1.5 mm) was observed using the RAPS. Conclusions: This work demonstrates the clinical efficacy of RAPS to measure applicator displacement.
- Is Part Of:
- Biomedical physics & engineering express. Volume 2:Number 4(2016)
- Journal:
- Biomedical physics & engineering express
- Issue:
- Volume 2:Number 4(2016)
- Issue Display:
- Volume 2, Issue 4 (2016)
- Year:
- 2016
- Volume:
- 2
- Issue:
- 4
- Issue Sort Value:
- 2016-0002-0004-0000
- Page Start:
- Page End:
- Publication Date:
- 2016-07-27
- Subjects:
- brachytherapy -- cervical cancer -- real-time tracking -- x-ray image
Medical physics -- Periodicals
Biophysics -- Periodicals
Biomedical engineering -- Periodicals
Medical sciences -- Periodicals
610.153 - Journal URLs:
- http://iopscience.iop.org/2057-1976/ ↗
http://www.iop.org/ ↗ - DOI:
- 10.1088/2057-1976/2/4/045008 ↗
- Languages:
- English
- ISSNs:
- 2057-1976
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 8438.xml