Characterisation of a plastic scintillation detector to be used in a multicentre stereotactic radiosurgery dosimetry audit. (November 2017)
- Record Type:
- Journal Article
- Title:
- Characterisation of a plastic scintillation detector to be used in a multicentre stereotactic radiosurgery dosimetry audit. (November 2017)
- Main Title:
- Characterisation of a plastic scintillation detector to be used in a multicentre stereotactic radiosurgery dosimetry audit
- Authors:
- Dimitriadis, A.
Patallo, I. Silvestre
Billas, I.
Duane, S.
Nisbet, A.
Clark, C.H. - Abstract:
- Abstract: Scintillation detectors are considered highly suitable for dosimetric measurement of small fields in radiotherapy due to their near-tissue equivalence and their small size. A commercially available scintillation detector, the Exradin W1 (Standard Imaging, Middleton, USA), has been previously characterised by two independent studies (Beierholm et al., 2014; Carrasco et al., 2015a, 2015b ) but the results from these publications differed in some aspects (e.g. energy dependence, long term stability). The respective authors highlighted the need for more studies to be published (Beierholm et al., 2015; Carrasco et al., 2015a, 2015b ). In this work, the Exradin W1 was characterised in terms of dose response, dependence on dose rate, energy, temperature and angle of irradiation, and long-term stability. The observed dose linearity, short-term repeatability and temperature dependence were in good agreement with previously published data. Appropriate corrections should therefore be applied, where possible, in order to achieve measurements with low-uncertainty. The angular dependence was characterised along both the symmetrical and polar axis of the detector for the first time in this work and a dose variation of up to 1% was observed. The response of the detector was observed to decrease at a rate of approximately 1.6% kGy −1 for the first 5 kGy delivered, and then stabilised to 0.2% kGy −1 in the subsequent 20 kGy. The main goal of this work was to assess the suitabilityAbstract: Scintillation detectors are considered highly suitable for dosimetric measurement of small fields in radiotherapy due to their near-tissue equivalence and their small size. A commercially available scintillation detector, the Exradin W1 (Standard Imaging, Middleton, USA), has been previously characterised by two independent studies (Beierholm et al., 2014; Carrasco et al., 2015a, 2015b ) but the results from these publications differed in some aspects (e.g. energy dependence, long term stability). The respective authors highlighted the need for more studies to be published (Beierholm et al., 2015; Carrasco et al., 2015a, 2015b ). In this work, the Exradin W1 was characterised in terms of dose response, dependence on dose rate, energy, temperature and angle of irradiation, and long-term stability. The observed dose linearity, short-term repeatability and temperature dependence were in good agreement with previously published data. Appropriate corrections should therefore be applied, where possible, in order to achieve measurements with low-uncertainty. The angular dependence was characterised along both the symmetrical and polar axis of the detector for the first time in this work and a dose variation of up to 1% was observed. The response of the detector was observed to decrease at a rate of approximately 1.6% kGy −1 for the first 5 kGy delivered, and then stabilised to 0.2% kGy −1 in the subsequent 20 kGy. The main goal of this work was to assess the suitability of the Exradin W1 for use in dose verification measurements for stereotactic radiosurgery. The results obtained confirm that the detector is suitable for use in such situations. The detector is now utilised in a multi-centre stereotactic radiosurgery dosimetric audit, with the application of appropriate correction factors. Highlights: A commercially available plastic scintillation detector was characterised. The detector showed suitable characteristics for use in a radiosurgery audit. An uncertainty budget was developed and the detector is now being used in a national UK stereotactic radiosurgery audit. … (more)
- Is Part Of:
- Radiation physics and chemistry. Volume 140(2017)
- Journal:
- Radiation physics and chemistry
- Issue:
- Volume 140(2017)
- Issue Display:
- Volume 140, Issue 2017 (2017)
- Year:
- 2017
- Volume:
- 140
- Issue:
- 2017
- Issue Sort Value:
- 2017-0140-2017-0000
- Page Start:
- 373
- Page End:
- 378
- Publication Date:
- 2017-11
- Subjects:
- Scintillation detector -- Dosimetry -- Radiotherapy -- Radiosurgery -- Audit
Radiation chemistry -- Periodicals
Radiometry -- Periodicals
Radiation -- Periodicals
Chimie sous rayonnement -- Périodiques
539.2 - Journal URLs:
- http://www.sciencedirect.com/science/journal/0969806X ↗
http://www.elsevier.com/journals ↗
http://www.journals.elsevier.com/radiation-physics-and-chemistry/ ↗ - DOI:
- 10.1016/j.radphyschem.2017.02.023 ↗
- Languages:
- English
- ISSNs:
- 0969-806X
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 7227.984000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 8345.xml