Reproducibility and Agreement of Anterior Segment Parameter Measurements Obtained Using the CASIA2 and Spectralis OCT2 Optical Coherence Tomography Devices. Issue 11 (November 2017)
- Record Type:
- Journal Article
- Title:
- Reproducibility and Agreement of Anterior Segment Parameter Measurements Obtained Using the CASIA2 and Spectralis OCT2 Optical Coherence Tomography Devices. Issue 11 (November 2017)
- Main Title:
- Reproducibility and Agreement of Anterior Segment Parameter Measurements Obtained Using the CASIA2 and Spectralis OCT2 Optical Coherence Tomography Devices
- Authors:
- Xu, Benjamin Y.
Mai, Derek D.
Penteado, Rafaella C.
Saunders, Luke
Weinreb, Robert N. - Abstract:
- Abstract : Purpose: To assess the reproducibility and agreement of measurement values obtained from the Tomey CASIA2 and Heidelberg Spectralis OCT2 anterior segment optical coherence tomographic devices. Methods: Twenty eyes from 10 subjects ranging from age 28 to 45 years with no history of eye conditions or intraocular surgery were included. Two scans were obtained with each device in a standardized dark room environment after a period of dark adaptation. One anterior segment optical coherence tomography image along the horizontal (temporal nasal) meridian was analyzed per eye and per scan. Lens vault, pupil diameter, anterior chamber width, angle opening distance, trabecular iris space area, and scleral spur angle were measured using manufacturer-provided image analysis programs. Intraclass correlation coefficient (ICC) values, coefficients of variation, and Bland-Altman plots were computed to assess the intradevice correlation and interdevice agreement of measurement values. Results: There was excellent intradevice reproducibility of measurement values for both the CASIA (ICC range, 0.86 to 0.99) and Spectralis (ICC range, 0.79 to 1.00). There was also excellent interdevice correlation of measurement values (ICC range, 0.78 to 0.93) for all parameters except anterior chamber width (ICC 0.20). Linear regression models and Bland-Altman plots showed that this relationship was strongest when measurement values were small. Conclusions: There is excellent intradeviceAbstract : Purpose: To assess the reproducibility and agreement of measurement values obtained from the Tomey CASIA2 and Heidelberg Spectralis OCT2 anterior segment optical coherence tomographic devices. Methods: Twenty eyes from 10 subjects ranging from age 28 to 45 years with no history of eye conditions or intraocular surgery were included. Two scans were obtained with each device in a standardized dark room environment after a period of dark adaptation. One anterior segment optical coherence tomography image along the horizontal (temporal nasal) meridian was analyzed per eye and per scan. Lens vault, pupil diameter, anterior chamber width, angle opening distance, trabecular iris space area, and scleral spur angle were measured using manufacturer-provided image analysis programs. Intraclass correlation coefficient (ICC) values, coefficients of variation, and Bland-Altman plots were computed to assess the intradevice correlation and interdevice agreement of measurement values. Results: There was excellent intradevice reproducibility of measurement values for both the CASIA (ICC range, 0.86 to 0.99) and Spectralis (ICC range, 0.79 to 1.00). There was also excellent interdevice correlation of measurement values (ICC range, 0.78 to 0.93) for all parameters except anterior chamber width (ICC 0.20). Linear regression models and Bland-Altman plots showed that this relationship was strongest when measurement values were small. Conclusions: There is excellent intradevice reproducibility and good interdevice agreement of anterior segment parameter measurement values for the CASIA2 and Spectralis OCT2. However, the measurements obtained with each device should not be considered interchangeable. … (more)
- Is Part Of:
- Journal of glaucoma. Volume 26:Issue 11(2017)
- Journal:
- Journal of glaucoma
- Issue:
- Volume 26:Issue 11(2017)
- Issue Display:
- Volume 26, Issue 11 (2017)
- Year:
- 2017
- Volume:
- 26
- Issue:
- 11
- Issue Sort Value:
- 2017-0026-0011-0000
- Page Start:
- Page End:
- Publication Date:
- 2017-11
- Subjects:
- anterior segment optical coherence tomography -- reproducibility -- comparison
Glaucoma -- Periodicals
617.741005 - Journal URLs:
- http://ovidsp.ovid.com/ovidweb.cgi?T=JS&NEWS=n&CSC=Y&PAGE=toc&D=yrovft&AN=00061198-000000000-00000 ↗
http://www.glaucomajournal.com ↗
http://journals.lww.com/glaucomajournal/pages/default.aspx ↗
http://journals.lww.com ↗ - DOI:
- 10.1097/IJG.0000000000000788 ↗
- Languages:
- English
- ISSNs:
- 1057-0829
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4996.230000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 8321.xml