Higher UHMWPE wear-rate in cementless compared with cemented cups with the Saturne® Dual-Mobility acetabular system. Issue 2 (March 2018)
- Record Type:
- Journal Article
- Title:
- Higher UHMWPE wear-rate in cementless compared with cemented cups with the Saturne® Dual-Mobility acetabular system. Issue 2 (March 2018)
- Main Title:
- Higher UHMWPE wear-rate in cementless compared with cemented cups with the Saturne® Dual-Mobility acetabular system
- Authors:
- Tabori-Jensen, Steffan
Frølich, Christina
Hansen, Torben B.
Bøvling, Søren
Homilius, Morten
Stilling, Maiken - Abstract:
- Introduction: Dual mobility (DM) total hip arthroplasty (THA) may reduce dislocation risk, but might increase the risk of high polyethylene (PE) wear due to double wearing surfaces. Methods: 127 patients (97 female) with 129 hips operated with THA after displaced femoral neck fracture FNF between 2005 and 2011, were seen for a cross-sectional clinical follow-up. Acetabular components were Saturne® DM cups with 28mm chrome-cobalt heads in UHMWPE. Cementless cups (n = 73) were hydroxyapatite coated. Radiographs were obtained for analysis of cup placement, 2D polyethylene wear and wear-rate (PolyWare 3D), and further radiological evaluation. Activity measurements included Timed Up and Go test (TUG) and walking distance from Harris Hip Score (HHS). Results: At a mean follow-up of 2.83 (1.0-7.7) years the mean wear was 0.82 mm (range 0.17-4.51, SD 0.50), and the wear-rate was 0.37 mm (range 0.06-1.90, SD 0.29). Wear-rate of 0.43 mm/year (SD 0.30) in cementless cups was higher (p = 0.004) than 0.30 mm/year (SD 0.27) in cemented cups. Mean age at time of surgery was 75.1 years (range 30-95). There was no correlation between age at time of surgery and wear (p = 0.56). There was no correlation between cup inclination and wear-rate (p = 0.35). TUG was mean 13.4 seconds (range 4.5-30.1) and correlated with wear rate (p = 0.03). Conclusions: At short term follow-up, the mean wear-rate in old and low demand patients was high, correlated to activity, and was above the generally acceptedIntroduction: Dual mobility (DM) total hip arthroplasty (THA) may reduce dislocation risk, but might increase the risk of high polyethylene (PE) wear due to double wearing surfaces. Methods: 127 patients (97 female) with 129 hips operated with THA after displaced femoral neck fracture FNF between 2005 and 2011, were seen for a cross-sectional clinical follow-up. Acetabular components were Saturne® DM cups with 28mm chrome-cobalt heads in UHMWPE. Cementless cups (n = 73) were hydroxyapatite coated. Radiographs were obtained for analysis of cup placement, 2D polyethylene wear and wear-rate (PolyWare 3D), and further radiological evaluation. Activity measurements included Timed Up and Go test (TUG) and walking distance from Harris Hip Score (HHS). Results: At a mean follow-up of 2.83 (1.0-7.7) years the mean wear was 0.82 mm (range 0.17-4.51, SD 0.50), and the wear-rate was 0.37 mm (range 0.06-1.90, SD 0.29). Wear-rate of 0.43 mm/year (SD 0.30) in cementless cups was higher (p = 0.004) than 0.30 mm/year (SD 0.27) in cemented cups. Mean age at time of surgery was 75.1 years (range 30-95). There was no correlation between age at time of surgery and wear (p = 0.56). There was no correlation between cup inclination and wear-rate (p = 0.35). TUG was mean 13.4 seconds (range 4.5-30.1) and correlated with wear rate (p = 0.03). Conclusions: At short term follow-up, the mean wear-rate in old and low demand patients was high, correlated to activity, and was above the generally accepted osteolysis threshold (0.1 mm/yr.). Cementless HA-coated cups had higher wear-rate than cemented cups. … (more)
- Is Part Of:
- Hip international. Volume 28:Issue 2(2018)
- Journal:
- Hip international
- Issue:
- Volume 28:Issue 2(2018)
- Issue Display:
- Volume 28, Issue 2 (2018)
- Year:
- 2018
- Volume:
- 28
- Issue:
- 2
- Issue Sort Value:
- 2018-0028-0002-0000
- Page Start:
- 125
- Page End:
- 132
- Publication Date:
- 2018-03
- Subjects:
- Dual mobility cup -- Femoral neck fracture -- Hip arthroplasty -- Hydroxyapatite -- Polyethylene wear -- UHMWPE
Hip joint -- Pathophysiology -- Periodicals
Hip joint -- Diseases -- Periodicals
Hip joint -- Surgery -- Periodicals
617.581005 - Journal URLs:
- http://www.uk.sagepub.com/home.nav ↗
http://journals.sagepub.com/home/hpi ↗ - DOI:
- 10.1177/1120700018768615 ↗
- Languages:
- English
- ISSNs:
- 1120-7000
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 8328.xml