Dosimetric effects of positioning shifts using 6D‐frameless stereotactic Brainlab system in hypofractionated intracranial radiotherapy. (8th January 2016)
- Record Type:
- Journal Article
- Title:
- Dosimetric effects of positioning shifts using 6D‐frameless stereotactic Brainlab system in hypofractionated intracranial radiotherapy. (8th January 2016)
- Main Title:
- Dosimetric effects of positioning shifts using 6D‐frameless stereotactic Brainlab system in hypofractionated intracranial radiotherapy
- Authors:
- Jin, Hosang
Keeling, Vance P.
Ali, Imad
Ahmad, Salahuddin - Abstract:
- Abstract : Dosimetric consequences of positional shifts were studied using frameless Brainlab ExacTrac X‐ray system for hypofractionated (3 or 5 fractions) intracranial stereotactic radiotherapy (SRT). SRT treatments of 17 patients with metastatic intracranial tumors using the stereotactic system were retrospectively investigated. The treatments were simulated in a treatment planning system by modifying planning parameters with a matrix conversion technique based on positional shifts for initial infrared (IR)‐based setup (XC: X‐ray correction) and post‐correction (XV: X‐ray verification). The simulation was implemented with (a) 3D translational shifts only and (b) 6D translational and rotational shifts for dosimetric effects of angular correction. Mean translations and rotations (± 1 SD) of 77 fractions based on the initial IR setup (XC) were 0.51 ± 0.86 mm (lateral), 0.30 ± 1.55 mm (longitudinal), and − 1.63 ± 1.00 mm (vertical); 0.53 ± 0.56 mm (pitch), 0.42 ± 0.60 mm (roll), and 0.44 ± 0.90 mm (yaw), respectively. These were − 0.07 ± 0.24 mm, − 0.07 ± 0.25 mm, 0.06 ± 0.21 mm, 0.04 ± 0.23 mm, 0.00 ± 0.30 mm, and 0.02 ± 0.22 mm, respectively, for the postcorrection (XV). Substantial degradation of the treatment plans was observed in D 95 of PTV ( 2.6 % ± 3.3 % ; simulated treatment versus treatment planning), D min of PTV ( 13.4 % ± 11.6 % ), and D min of CTV ( 2.8 % ± 3.8 %, with the maximum error of 10.0%) from XC, while dosimetrically negligibleAbstract : Dosimetric consequences of positional shifts were studied using frameless Brainlab ExacTrac X‐ray system for hypofractionated (3 or 5 fractions) intracranial stereotactic radiotherapy (SRT). SRT treatments of 17 patients with metastatic intracranial tumors using the stereotactic system were retrospectively investigated. The treatments were simulated in a treatment planning system by modifying planning parameters with a matrix conversion technique based on positional shifts for initial infrared (IR)‐based setup (XC: X‐ray correction) and post‐correction (XV: X‐ray verification). The simulation was implemented with (a) 3D translational shifts only and (b) 6D translational and rotational shifts for dosimetric effects of angular correction. Mean translations and rotations (± 1 SD) of 77 fractions based on the initial IR setup (XC) were 0.51 ± 0.86 mm (lateral), 0.30 ± 1.55 mm (longitudinal), and − 1.63 ± 1.00 mm (vertical); 0.53 ± 0.56 mm (pitch), 0.42 ± 0.60 mm (roll), and 0.44 ± 0.90 mm (yaw), respectively. These were − 0.07 ± 0.24 mm, − 0.07 ± 0.25 mm, 0.06 ± 0.21 mm, 0.04 ± 0.23 mm, 0.00 ± 0.30 mm, and 0.02 ± 0.22 mm, respectively, for the postcorrection (XV). Substantial degradation of the treatment plans was observed in D 95 of PTV ( 2.6 % ± 3.3 % ; simulated treatment versus treatment planning), D min of PTV ( 13.4 % ± 11.6 % ), and D min of CTV ( 2.8 % ± 3.8 %, with the maximum error of 10.0%) from XC, while dosimetrically negligible changes ( < 0.1 % ) were detected for both CTV and PTV from XV simulation. 3D angular correction significantly improved CTV dose coverage when the total angular shifts ( | pitch | + | roll | + | yaw | ) were greater than 2°. With the 6D stereoscopic X‐ray verification imaging and frameless immobilization, submillimeter and subdegree accuracy is achieved with negligible dosimetric deviations. 3D angular correction is required when the angular deviation is substantial. A CTV‐to‐PTV safety margin of 2 mm is large enough to prevent deterioration of CTV coverage. PACS number: 87.55.dk … (more)
- Is Part Of:
- Journal of applied clinical medical physics. Volume 17:Number 1(2016)
- Journal:
- Journal of applied clinical medical physics
- Issue:
- Volume 17:Number 1(2016)
- Issue Display:
- Volume 17, Issue 1 (2016)
- Year:
- 2016
- Volume:
- 17
- Issue:
- 1
- Issue Sort Value:
- 2016-0017-0001-0000
- Page Start:
- 102
- Page End:
- 111
- Publication Date:
- 2016-01-08
- Subjects:
- frameless stereotactic radiotherapy -- dosimetric effects -- translational and rotational shifts -- PTV margin -- hypofractionation
Medical physics -- Periodicals
Clinical medicine -- Periodicals
Health Physics
Clinical Medicine
Electronic journals
Periodicals
Periodicals
Fulltext
Internet Resources
610.153 - Journal URLs:
- http://aapm.onlinelibrary.wiley.com/hub/journal/10.1002/(ISSN)1526-9914/ ↗
http://bibpurl.oclc.org/web/7294 ↗
http://www.jacmp.org/ ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1120/jacmp.v17i1.5682 ↗
- Languages:
- English
- ISSNs:
- 1526-9914
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 8320.xml