Commissioning an Elekta Versa HD linear accelerator. (8th January 2016)
- Record Type:
- Journal Article
- Title:
- Commissioning an Elekta Versa HD linear accelerator. (8th January 2016)
- Main Title:
- Commissioning an Elekta Versa HD linear accelerator
- Authors:
- Narayanasamy, Ganesh
Saenz, Daniel
Cruz, Wilbert
Ha, Chul S.
Papanikolaou, Niko
Stathakis, Sotirios - Abstract:
- Abstract : The purpose of this study is to report the dosimetric aspects of commissioning performed on an Elekta Versa HD linear accelerator (linac) with high‐dose‐rate flattening filter‐free (FFF) photon modes and electron modes. Acceptance and commissioning was performed on the Elekta Versa HD linac with five photon energies (6 MV, 10 MV, 18 MV, 6 MV FFF, 10 MV FFF), four electron energies (6 MeV, 9 MeV, 12 MeV, 15 MeV) and 160‐leaf (5 mm wide) multileaf collimators (MLCs). Mechanical and dosimetric data were measured and evaluated. The measurements include percent depth doses (PDDs), in‐plane and cross‐plane profiles, head scatter factor ( S c ), relative photon output factors ( S cp ), universal wedge transmission factor, MLC transmission factors, and electron cone factors. Gantry, collimator, and couch isocentricity measurements were within 1 mm, 0.7 mm, and 0.7 mm diameter, respectively. The PDDs of 6 MV FFF and 10 MV FFF beams show deeper d max and steeper falloff with depth than the corresponding flattened beams. While flatness values of 6 MV FFF and 10 MV FFF normalized profiles were expectedly higher than the corresponding flattened beams, the symmetry values were almost identical. The cross‐plane penumbra values were higher than the in‐plane penumbra values for all the energies. The MLC transmission values were 0.5%, 0.6%, and 0.6% for 6 MV, 10 MV, and 18 MV photon beams, respectively. The electron PDDs, profiles, and cone factors agree well with the literature.Abstract : The purpose of this study is to report the dosimetric aspects of commissioning performed on an Elekta Versa HD linear accelerator (linac) with high‐dose‐rate flattening filter‐free (FFF) photon modes and electron modes. Acceptance and commissioning was performed on the Elekta Versa HD linac with five photon energies (6 MV, 10 MV, 18 MV, 6 MV FFF, 10 MV FFF), four electron energies (6 MeV, 9 MeV, 12 MeV, 15 MeV) and 160‐leaf (5 mm wide) multileaf collimators (MLCs). Mechanical and dosimetric data were measured and evaluated. The measurements include percent depth doses (PDDs), in‐plane and cross‐plane profiles, head scatter factor ( S c ), relative photon output factors ( S cp ), universal wedge transmission factor, MLC transmission factors, and electron cone factors. Gantry, collimator, and couch isocentricity measurements were within 1 mm, 0.7 mm, and 0.7 mm diameter, respectively. The PDDs of 6 MV FFF and 10 MV FFF beams show deeper d max and steeper falloff with depth than the corresponding flattened beams. While flatness values of 6 MV FFF and 10 MV FFF normalized profiles were expectedly higher than the corresponding flattened beams, the symmetry values were almost identical. The cross‐plane penumbra values were higher than the in‐plane penumbra values for all the energies. The MLC transmission values were 0.5%, 0.6%, and 0.6% for 6 MV, 10 MV, and 18 MV photon beams, respectively. The electron PDDs, profiles, and cone factors agree well with the literature. The outcome of radiation treatment is directly related to the accuracy in the dose modeled in the treatment planning system, which is based on the commissioned data. Commissioning data provided us a valuable insight into the dosimetric characteristics of the beam. This set of commissioning data can provide comparison data to others performing Versa HD commissioning, thereby improving patient safety. PACS number(s): 87.56.bd … (more)
- Is Part Of:
- Journal of applied clinical medical physics. Volume 17:Number 1(2016)
- Journal:
- Journal of applied clinical medical physics
- Issue:
- Volume 17:Number 1(2016)
- Issue Display:
- Volume 17, Issue 1 (2016)
- Year:
- 2016
- Volume:
- 17
- Issue:
- 1
- Issue Sort Value:
- 2016-0017-0001-0000
- Page Start:
- 179
- Page End:
- 191
- Publication Date:
- 2016-01-08
- Subjects:
- commissioning -- Versa HD -- linac -- dosimetry -- flattening filter‐free (FFF)
Medical physics -- Periodicals
Clinical medicine -- Periodicals
Health Physics
Clinical Medicine
Electronic journals
Periodicals
Periodicals
Fulltext
Internet Resources
610.153 - Journal URLs:
- http://aapm.onlinelibrary.wiley.com/hub/journal/10.1002/(ISSN)1526-9914/ ↗
http://bibpurl.oclc.org/web/7294 ↗
http://www.jacmp.org/ ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1120/jacmp.v17i1.5799 ↗
- Languages:
- English
- ISSNs:
- 1526-9914
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 8295.xml