Blood pressure cuff comparability study. Issue 6 (December 2016)
- Record Type:
- Journal Article
- Title:
- Blood pressure cuff comparability study. Issue 6 (December 2016)
- Main Title:
- Blood pressure cuff comparability study
- Authors:
- Ostchega, Yechiam
Nwankwo, Tatiana
Zhang, Guangyu
Chiappa, Michele - Abstract:
- Abstract : Background: Manufacturer-supplied blood pressure (BP) cuffs are part of the automatic oscillometric BP devices algorithm. Materials and methods: This study assessed the differences in BP values using the Omron HEM 907-XL (Omron) device with two types of cuffs: the Baum cuff (BC) and the supplied Omron cuff (OC). A sample of 102 adults participated in the study, 34 per cuff size (adult, large, and extra-large). After a 5-min resting period, three pairs of BP determinations (systolic and diastolic) were taken simultaneously on both arms. One arm was cuffed with a BC and the other arm was cuffed with an OC. The cuffs were switched to opposite arms after 5 min of rest. The order was decided randomly as to which cuff was applied to which arm first. Results: The BP readings were highly correlated between the cuffs (systolic BP, r =0.98; diastolic BP, r =0.98). The overall mean differences (BC−OC) were 2.66 mmHg (SD=3.9 mmHg) for systolic BP ( P <0.05) and 0.33 mmHg (SD=2.03 mmHg) for diastolic BP ( P >0.05). Increased cuff size corresponded to increased differences in systolic BP values (adult: 1.51 mmHg; large: 2.56 mmHg; and extra-large: 3.9 mmHg; P <0.05). For diastolic BP values, a statistically significant difference was observed only for adult cuff size (difference=1.31 mmHg, SD=1.34 mmHg, P <0.05). Conclusion: Using a BC with the Omron could result in higher systolic BP readings and higher diastolic BP readings with the adult cuff size.
- Is Part Of:
- Blood pressure monitoring. Volume 21:Issue 6(2016)
- Journal:
- Blood pressure monitoring
- Issue:
- Volume 21:Issue 6(2016)
- Issue Display:
- Volume 21, Issue 6 (2016)
- Year:
- 2016
- Volume:
- 21
- Issue:
- 6
- Issue Sort Value:
- 2016-0021-0006-0000
- Page Start:
- Page End:
- Publication Date:
- 2016-12
- Subjects:
- automatic oscillometric blood pressure device -- blood pressure cuffs -- comparison
Blood pressure -- Measurement -- Periodicals
Blood Pressure Determination -- Periodicals
Blood Pressure Monitoring, Ambulatory -- Periodicals
Blood Pressure Monitors -- Periodicals
Hypertension -- Periodicals
612.140287 - Journal URLs:
- http://gateway.ovid.com/ovidweb.cgi?T=JS&MODE=ovid&PAGE=toc&D=ovft&AN=00126097-000000000-00000&NEWS=N ↗
http://ovidsp.ovid.com/ovidweb.cgi?T=JS&NEWS=n&CSC=Y&PAGE=toc&D=yrovft&AN=00126097-000000000-00000 ↗
http://www.bpmonitoring.com/ ↗
http://journals.lww.com/pages/default.aspx ↗
http://firstsearch.oclc.org ↗ - DOI:
- 10.1097/MBP.0000000000000208 ↗
- Languages:
- English
- ISSNs:
- 1359-5237
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 2113.035500
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 8244.xml