Performance Evaluation of Amperometric Sensors for the Monitoring of O3 and NO2 in Ambient Air at ppb Level. (2015)
- Record Type:
- Journal Article
- Title:
- Performance Evaluation of Amperometric Sensors for the Monitoring of O3 and NO2 in Ambient Air at ppb Level. (2015)
- Main Title:
- Performance Evaluation of Amperometric Sensors for the Monitoring of O3 and NO2 in Ambient Air at ppb Level
- Authors:
- Spinelle, Laurent
Gerboles, Michel
Aleixandre, Manuel - Abstract:
- Abstract: Hereafter we present a laboratory evaluation of commercially available amperometric sensors for the monitoring of O3 and NO2 in ambient air at ppb level. The tests includes the determination of the response time of sensors, their calibration function, the evaluations of repeatability, short and long term drifts, the hysteresis effect and the matrix effect. Interferences from gaseous compounds, temperature and humidity parameters are also evaluated.
- Is Part Of:
- Procedia engineering. Volume 120(2015)
- Journal:
- Procedia engineering
- Issue:
- Volume 120(2015)
- Issue Display:
- Volume 120, Issue 2015 (2015)
- Year:
- 2015
- Volume:
- 120
- Issue:
- 2015
- Issue Sort Value:
- 2015-0120-2015-0000
- Page Start:
- 480
- Page End:
- 483
- Publication Date:
- 2015
- Subjects:
- Protocol of sensor evaluation -- measurement uncertainty -- low cost sensors -- validation
Engineering -- Congresses
Engineering -- Periodicals
Engineering
Conference proceedings
Periodicals
620.005 - Journal URLs:
- http://www.sciencedirect.com/science/journal/18777058 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.proeng.2015.08.676 ↗
- Languages:
- English
- ISSNs:
- 1877-7058
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 8186.xml