The effect of a low radiation CT protocol on accuracy of CT guided implant migration measurement: A cadaver study. Issue 4 (25th November 2015)
- Record Type:
- Journal Article
- Title:
- The effect of a low radiation CT protocol on accuracy of CT guided implant migration measurement: A cadaver study. Issue 4 (25th November 2015)
- Main Title:
- The effect of a low radiation CT protocol on accuracy of CT guided implant migration measurement: A cadaver study
- Authors:
- Boettner, Friedrich
Sculco, Peter K.
Lipman, Joseph
Saboeiro, Gregory
Renner, Lisa
Faschingbauer, Martin - Abstract:
- ABSTRACT: The current study compared the impact of low radiation CT protocols on the accuracy, repeatability, and inter‐ and intra‐observer variability of implant migration studies in total hip arthroplasty. Two total hip replacements were performed in two human cadavers and six tantalum beads were inserted into the femur similar to radiostereometric analysis. Six different 28 mm heads (−3 mm, 0 mm, 2.5 mm, 5.0 mm, 7.5 mm, and 10 mm) were added to simulate five reproducible translations (maximum total point migration) of the center of the head. Three CT scans with varying levels of radiation were performed for each head position. The effective dose (mSv) was 3.8 mSv for Protocol A (standard protocol), 0.7 mSv for Protocol B and 1.6 mSv for Protocol C. Implant migration was measured in a 3‐D analysis software (Geomagic Studio 7). The accuracy was 0.16 mm for CT Protocol A, 0.13 mm for Protocol B and 0.14 mm for Protocol C; The repeatability was 0.22 mm for CT Protocol A, 0.18 mm for Protocol B and 0.20 mm for Protocol C; ICC for inter observer reliability was 0.89, intra observer reliability was 0.95. The difference in accuracy between standard protocol A and the two low radiation protocols (B, C) was less than 0.05 mm. The accuracy, inter‐ and intra‐observer reliability of all three CT protocols is comparable to radiostereometric analysis. Reducing the CT radiation exposure to numbers similar to an AP Pelvis radiograph (0.7 mSv protocol B) does not affect the accuracy ofABSTRACT: The current study compared the impact of low radiation CT protocols on the accuracy, repeatability, and inter‐ and intra‐observer variability of implant migration studies in total hip arthroplasty. Two total hip replacements were performed in two human cadavers and six tantalum beads were inserted into the femur similar to radiostereometric analysis. Six different 28 mm heads (−3 mm, 0 mm, 2.5 mm, 5.0 mm, 7.5 mm, and 10 mm) were added to simulate five reproducible translations (maximum total point migration) of the center of the head. Three CT scans with varying levels of radiation were performed for each head position. The effective dose (mSv) was 3.8 mSv for Protocol A (standard protocol), 0.7 mSv for Protocol B and 1.6 mSv for Protocol C. Implant migration was measured in a 3‐D analysis software (Geomagic Studio 7). The accuracy was 0.16 mm for CT Protocol A, 0.13 mm for Protocol B and 0.14 mm for Protocol C; The repeatability was 0.22 mm for CT Protocol A, 0.18 mm for Protocol B and 0.20 mm for Protocol C; ICC for inter observer reliability was 0.89, intra observer reliability was 0.95. The difference in accuracy between standard protocol A and the two low radiation protocols (B, C) was less than 0.05 mm. The accuracy, inter‐ and intra‐observer reliability of all three CT protocols is comparable to radiostereometric analysis. Reducing the CT radiation exposure to numbers similar to an AP Pelvis radiograph (0.7 mSv protocol B) does not affect the accuracy of implant migration measurements. Keywords: Radiation dose, CT, migration analysis, cadaver. © 2015 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 34:725–728, 2016. … (more)
- Is Part Of:
- Journal of orthopaedic research. Volume 34:Issue 4(2016)
- Journal:
- Journal of orthopaedic research
- Issue:
- Volume 34:Issue 4(2016)
- Issue Display:
- Volume 34, Issue 4 (2016)
- Year:
- 2016
- Volume:
- 34
- Issue:
- 4
- Issue Sort Value:
- 2016-0034-0004-0000
- Page Start:
- 725
- Page End:
- 728
- Publication Date:
- 2015-11-25
- Subjects:
- femoral migration -- CT protocol -- low radiation -- cadaver
Orthopedics -- Periodicals
Musculoskeletal system -- Periodicals
616.7 - Journal URLs:
- http://onlinelibrary.wiley.com/ ↗
- DOI:
- 10.1002/jor.23060 ↗
- Languages:
- English
- ISSNs:
- 0736-0266
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5027.665000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 8098.xml