Successful bony integration of a porous tantalum implant despite longlasting and ongoing infection: Histologic workup of an explanted shoulder prosthesis. Issue 8 (10th September 2018)
- Record Type:
- Journal Article
- Title:
- Successful bony integration of a porous tantalum implant despite longlasting and ongoing infection: Histologic workup of an explanted shoulder prosthesis. Issue 8 (10th September 2018)
- Main Title:
- Successful bony integration of a porous tantalum implant despite longlasting and ongoing infection: Histologic workup of an explanted shoulder prosthesis
- Authors:
- Wahl, Peter
Sprecher, Christoph M.
Brüning, Christian
Meier, Christoph
Milz, Stefan
Gautier, Emanuel
Fintan Moriarty, T - Abstract:
- ABSTRACT: Infection associated with an implant is a complication feared in surgery, as it leads to loosening and dysfunction. This report documents an unexpected good bony integration of a porous tantalum shoulder prosthesis despite infection. A shoulder prosthesis with a porous tantalum glenoidal base plate was retrieved after 3 years of ongoing infection with Staphylococcus spp . Methyl‐methacrylate embedded sections of the retrieved glenoidal component were analyzed by optical and scanning electron beam microscopy (SEM). Bone ongrowth and ingrowth were quantified. Bone had formed at the implant surface and within the open cell structure of the porous tantalum. The bone implant contact index was 32%. The bone ingrowth or relative bone area within the open structure was 8.2%, respectively 11.9% in the outer 50% of the thickness. Due to the section thickness, bone ongrowth could best be documented in SEM. Despite long‐lasting and ongoing infection, the glenoidal base plate of the prosthesis showed good bony integration upon removal. The bone ingrowth into the porous tantalum was comparable to the values previously reported for the undersurface of retrieved proximal humerus resurfacing implants. Good integration of the implant however did not solve the problem of infection, and related morbidity. © 2018 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 106B: 2924–2931, 2018.
- Is Part Of:
- Journal of biomedical materials research. Volume 106:Issue 8(2018)
- Journal:
- Journal of biomedical materials research
- Issue:
- Volume 106:Issue 8(2018)
- Issue Display:
- Volume 106, Issue 8 (2018)
- Year:
- 2018
- Volume:
- 106
- Issue:
- 8
- Issue Sort Value:
- 2018-0106-0008-0000
- Page Start:
- 2924
- Page End:
- 2931
- Publication Date:
- 2018-09-10
- Subjects:
- porous tantalum -- trabecular metal -- infection -- implant -- bone ingrowth
Biomedical materials -- Periodicals
610.28 - Journal URLs:
- http://onlinelibrary.wiley.com/ ↗
- DOI:
- 10.1002/jbm.b.34174 ↗
- Languages:
- English
- ISSNs:
- 1552-4973
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4953.725000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 8003.xml