In-parallel connected intermittent hemodialysis through ECMO does not affect hemodynamic parameters derived from transpulmonary thermodilution. (November 2017)
- Record Type:
- Journal Article
- Title:
- In-parallel connected intermittent hemodialysis through ECMO does not affect hemodynamic parameters derived from transpulmonary thermodilution. (November 2017)
- Main Title:
- In-parallel connected intermittent hemodialysis through ECMO does not affect hemodynamic parameters derived from transpulmonary thermodilution
- Authors:
- Lahmer, Tobias
Mayr, Ulrich
Rasch, Sebastian
Batres Baires, Gonzalo
Schmid, Roland M.
Huber, Wolfgang - Abstract:
- Introduction: We report a case of renal replacement therapy (RRT) during extracorporeal membrane oxygenation (ECMO) via a single venous acce ss and analyze the feasibility of transpulmonary thermodilution (TPTD) for hemodynamic monitoring. Case report: ECMO and RRT connected into the ECMO-extracorporeal circuit were performed via a single venous access because of multiple venous thromboses. An indicator for TPTD and pulse contour analysis (PCA) was applied into the central venous catheter (CVC) placed in the right vena jugularis. TPTD and PCA demonstrated comparable cardiac index. Discussion: Congruent data for TPTD and PCA could be observed during TPTD and PCA measurements before ECMO, after ECMO and during ECMO and RRT. This might be explained by high blood flow having the lowest impact on TPTD by venous drainage in the femoral vein/distal vena cava and the TPTD indicator injection using the jugular CVC, as reported in our case. Conclusion: Hemodynamic monitoring using TPTD and PCA during ECMO/RRT is feasible and provides reliable results.
- Is Part Of:
- Perfusion. Volume 32:Number 8(2017)
- Journal:
- Perfusion
- Issue:
- Volume 32:Number 8(2017)
- Issue Display:
- Volume 32, Issue 8 (2017)
- Year:
- 2017
- Volume:
- 32
- Issue:
- 8
- Issue Sort Value:
- 2017-0032-0008-0000
- Page Start:
- 702
- Page End:
- 705
- Publication Date:
- 2017-11
- Subjects:
- ECMO -- RRT -- hemodynamic monitoring
Perfusion (Physiology) -- Periodicals
Blood -- Circulation, Artificial -- Periodicals
Heart -- Surgery -- Periodicals
Extracorporeal Circulation -- Periodicals
Perfusion -- Periodicals
Circulation extracorporelle -- Périodiques
Perfusion -- Périodiques
617.41 - Journal URLs:
- http://prf.sagepub.com/ ↗
http://www.uk.sagepub.com/home.nav ↗ - DOI:
- 10.1177/0267659117707816 ↗
- Languages:
- English
- ISSNs:
- 0267-6591
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 7970.xml