Remote beam output audits: A global assessment of results out of tolerance. (July 2018)
- Record Type:
- Journal Article
- Title:
- Remote beam output audits: A global assessment of results out of tolerance. (July 2018)
- Main Title:
- Remote beam output audits: A global assessment of results out of tolerance
- Authors:
- Kry, Stephen F.
Peterson, Christine B.
Howell, Rebecca M.
Izewska, Joanna
Lye, Jessica
Clark, Catharine H.
Nakamura, Mitsuhiro
Hurkmans, Coen
Alvarez, Paola
Alves, Andrew
Bokulic, Tomislav
Followill, David
Kazantsev, Pavel
Lowenstein, Jessica
Molineu, Andrea
Palmer, Jacob
Smith, Susan A.
Taylor, Paige
Wesolowska, Paulina
Williams, Ivan - Abstract:
- Abstract: Background and purpose: Remote beam output audits, which independently measure an institution's machine calibration, are a common component of independent radiotherapy peer review. This work reviews the results and trends of these audit results across several organisations and geographical regions. Materials and methods: Beam output audit results from the Australian Clinical Dosimetry Services, International Atomic Energy Agency, Imaging and Radiation Oncology Core, and Radiation Dosimetry Services were evaluated from 2010 to the present. The rate of audit results outside a ±5% tolerance was evaluated for photon and electron beams as a function of the year of irradiation and nominal beam energy. Additionally, examples of confirmed calibration errors were examined to provide guidance to clinical physicists and auditing bodies. Results: Of the 210, 167 audit results, 1323 (0.63%) were outside of tolerance. There was a clear trend of improved audit performance for more recent dates, and while all photon energies generally showed uniform rates of results out of tolerance, low (6 MeV) and high (≥18 MeV) energy electron beams showed significantly elevated rates. Twenty nine confirmed calibration errors were explored and attributed to a range of issues, such as equipment failures, errors in setup, and errors in performing the clinical reference calibration. Forty-two percent of these confirmed errors were detected during ongoing periodic monitoring, and not at the time ofAbstract: Background and purpose: Remote beam output audits, which independently measure an institution's machine calibration, are a common component of independent radiotherapy peer review. This work reviews the results and trends of these audit results across several organisations and geographical regions. Materials and methods: Beam output audit results from the Australian Clinical Dosimetry Services, International Atomic Energy Agency, Imaging and Radiation Oncology Core, and Radiation Dosimetry Services were evaluated from 2010 to the present. The rate of audit results outside a ±5% tolerance was evaluated for photon and electron beams as a function of the year of irradiation and nominal beam energy. Additionally, examples of confirmed calibration errors were examined to provide guidance to clinical physicists and auditing bodies. Results: Of the 210, 167 audit results, 1323 (0.63%) were outside of tolerance. There was a clear trend of improved audit performance for more recent dates, and while all photon energies generally showed uniform rates of results out of tolerance, low (6 MeV) and high (≥18 MeV) energy electron beams showed significantly elevated rates. Twenty nine confirmed calibration errors were explored and attributed to a range of issues, such as equipment failures, errors in setup, and errors in performing the clinical reference calibration. Forty-two percent of these confirmed errors were detected during ongoing periodic monitoring, and not at the time of the first audit of the machine. Conclusions: Remote beam output audits have identified, and continue to identify, numerous and often substantial beam calibration errors. … (more)
- Is Part Of:
- Physics and imaging in radiation oncology. Volume 7(2018)
- Journal:
- Physics and imaging in radiation oncology
- Issue:
- Volume 7(2018)
- Issue Display:
- Volume 7, Issue 2018 (2018)
- Year:
- 2018
- Volume:
- 7
- Issue:
- 2018
- Issue Sort Value:
- 2018-0007-2018-0000
- Page Start:
- 39
- Page End:
- 44
- Publication Date:
- 2018-07
- Subjects:
- Global harmonization group -- Remote beam output audit -- Dosimetry audit -- Calibration -- QA
Radiotherapy -- Periodicals
Radiation dosimetry -- Periodicals
Cancer -- Imaging -- Periodicals
Oncology -- Periodicals
615.842 - Journal URLs:
- http://www.sciencedirect.com/ ↗
https://www.journals.elsevier.com/physics-and-imaging-in-radiation-oncology/ ↗ - DOI:
- 10.1016/j.phro.2018.08.005 ↗
- Languages:
- English
- ISSNs:
- 2405-6316
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 7999.xml