A pressure and shear sensor system for stress measurement at lower limb residuum/socket interface. Issue 7 (July 2016)
- Record Type:
- Journal Article
- Title:
- A pressure and shear sensor system for stress measurement at lower limb residuum/socket interface. Issue 7 (July 2016)
- Main Title:
- A pressure and shear sensor system for stress measurement at lower limb residuum/socket interface
- Authors:
- Laszczak, P.
McGrath, M.
Tang, J.
Gao, J.
Jiang, L.
Bader, D.L.
Moser, D.
Zahedi, S. - Abstract:
- Highlights: A sensor system is reported to measure pressure and shear in lower limb prosthetics. Static and dynamic performance was calibrated using mechanical test machine. Pilot amputee test was conducted to measure stresses at residuum/socket interface. Amputee test results and analysis preliminarily validate sensor system performance. Abstract: A sensor system for measurement of pressure and shear at the lower limb residuum/socket interface is described. The system comprises of a flexible sensor unit and a data acquisition unit with wireless data transmission capability. Static and dynamic performance of the sensor system was characterised using a mechanical test machine. The static calibration results suggest that the developed sensor system presents high linearity (linearity error ≤ 3.8%) and resolution (0.9 kPa for pressure and 0.2 kPa for shear). Dynamic characterisation of the sensor system shows hysteresis error of approximately 15% for pressure and 8% for shear. Subsequently, a pilot amputee walking test was conducted. Three sensors were placed at the residuum/socket interface of a knee disarticulation amputee and simultaneous measurements were obtained during pilot amputee walking test. The pressure and shear peak values as well as their temporal profiles are presented and discussed. In particular, peak pressure and shear of approximately 58 kPa and 27 kPa, respectively, were recorded. Their temporal profiles also provide dynamic coupling information at thisHighlights: A sensor system is reported to measure pressure and shear in lower limb prosthetics. Static and dynamic performance was calibrated using mechanical test machine. Pilot amputee test was conducted to measure stresses at residuum/socket interface. Amputee test results and analysis preliminarily validate sensor system performance. Abstract: A sensor system for measurement of pressure and shear at the lower limb residuum/socket interface is described. The system comprises of a flexible sensor unit and a data acquisition unit with wireless data transmission capability. Static and dynamic performance of the sensor system was characterised using a mechanical test machine. The static calibration results suggest that the developed sensor system presents high linearity (linearity error ≤ 3.8%) and resolution (0.9 kPa for pressure and 0.2 kPa for shear). Dynamic characterisation of the sensor system shows hysteresis error of approximately 15% for pressure and 8% for shear. Subsequently, a pilot amputee walking test was conducted. Three sensors were placed at the residuum/socket interface of a knee disarticulation amputee and simultaneous measurements were obtained during pilot amputee walking test. The pressure and shear peak values as well as their temporal profiles are presented and discussed. In particular, peak pressure and shear of approximately 58 kPa and 27 kPa, respectively, were recorded. Their temporal profiles also provide dynamic coupling information at this critical residuum/socket interface. These preliminary amputee test results suggest strong potential of the developed sensor system for exploitation as an assistive technology to facilitate socket design, socket fit and effective monitoring of lower limb residuum health. … (more)
- Is Part Of:
- Medical engineering & physics. Volume 38:Issue 7(2016:Jul.)
- Journal:
- Medical engineering & physics
- Issue:
- Volume 38:Issue 7(2016:Jul.)
- Issue Display:
- Volume 38, Issue 7 (2016)
- Year:
- 2016
- Volume:
- 38
- Issue:
- 7
- Issue Sort Value:
- 2016-0038-0007-0000
- Page Start:
- 695
- Page End:
- 700
- Publication Date:
- 2016-07
- Subjects:
- Pressure and shear sensor -- Residuum/socket interface -- Lower-limb amputees
Biomedical engineering -- Periodicals
Biomedical Engineering -- Periodicals
Physics -- Periodicals
Génie biomédical -- Périodiques
Biomedical engineering
Electronic journals
Periodicals
610.28 - Journal URLs:
- http://www.medengphys.com ↗
http://www.sciencedirect.com/science/journal/13504533 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/13504533 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/13504533 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.medengphy.2016.04.007 ↗
- Languages:
- English
- ISSNs:
- 1350-4533
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5527.323000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 7932.xml