Validation of the G.LAB MD2200 wrist blood pressure monitor according to the European Society of Hypertension, the British Hypertension Society, and the International Organization for Standardization Protocols. Issue 2 (April 2017)
- Record Type:
- Journal Article
- Title:
- Validation of the G.LAB MD2200 wrist blood pressure monitor according to the European Society of Hypertension, the British Hypertension Society, and the International Organization for Standardization Protocols. Issue 2 (April 2017)
- Main Title:
- Validation of the G.LAB MD2200 wrist blood pressure monitor according to the European Society of Hypertension, the British Hypertension Society, and the International Organization for Standardization Protocols
- Authors:
- Liu, Ze-Yu
Zhang, Qing-Han
Ye, Xiao-Lei
Liu, Da-Peng
Cheng, Kang
Zhang, Chun-Hai
Wan, Yi - Abstract:
- Abstract : Objective: To validate the G.LAB MD2200 automated wrist blood pressure (BP) monitors according to the European Society of Hypertension International Protocol (ESH-IP) revision 2010, the British Hypertension Society (BHS), and the International Organization for Standardization (ISO) 81060-2:2013 protocols. Materials and methods: The device was assessed on 33 participants according to the ESH requirements and was then tested on 85 participants according to the BHS and ISO 81060-2:2013 criteria. The validation procedures and data analysis followed the protocols precisely. Results: The G.LAB MD2200 devices passed all parts of ESH-IP revision 2010 for both systolic and diastolic BP, with a device–observer difference of 2.15±5.51 and 1.51±5.16 mmHg, respectively. The device achieved A/A grading for the BHS protocol and it also fulfilled the criteria of ISO 81060-2:2013, with mean differences of systolic and diastolic BP between the device and the observer of 2.19±5.21 and 2.11±4.70 mmHg, respectively. Conclusion: The G.LAB MD2200 automated wrist BP monitor passed the ESH-IP revision 2010 and the ISO 81060-2:2013 protocol, and achieved the A/A grade of the BHS protocol, which can be recommended for self-measurement in the general population.
- Is Part Of:
- Blood pressure monitoring. Volume 22:Issue 2(2017)
- Journal:
- Blood pressure monitoring
- Issue:
- Volume 22:Issue 2(2017)
- Issue Display:
- Volume 22, Issue 2 (2017)
- Year:
- 2017
- Volume:
- 22
- Issue:
- 2
- Issue Sort Value:
- 2017-0022-0002-0000
- Page Start:
- Page End:
- Publication Date:
- 2017-04
- Subjects:
- blood pressure -- self-measurement -- validation
Blood pressure -- Measurement -- Periodicals
Blood Pressure Determination -- Periodicals
Blood Pressure Monitoring, Ambulatory -- Periodicals
Blood Pressure Monitors -- Periodicals
Hypertension -- Periodicals
612.140287 - Journal URLs:
- http://gateway.ovid.com/ovidweb.cgi?T=JS&MODE=ovid&PAGE=toc&D=ovft&AN=00126097-000000000-00000&NEWS=N ↗
http://ovidsp.ovid.com/ovidweb.cgi?T=JS&NEWS=n&CSC=Y&PAGE=toc&D=yrovft&AN=00126097-000000000-00000 ↗
http://www.bpmonitoring.com/ ↗
http://journals.lww.com/pages/default.aspx ↗
http://firstsearch.oclc.org ↗ - DOI:
- 10.1097/MBP.0000000000000234 ↗
- Languages:
- English
- ISSNs:
- 1359-5237
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 2113.035500
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 7903.xml