GafChromic® EBT3 films for patient specific IMRT QA using a multichannel approach. Issue 8 (December 2015)
- Record Type:
- Journal Article
- Title:
- GafChromic® EBT3 films for patient specific IMRT QA using a multichannel approach. Issue 8 (December 2015)
- Main Title:
- GafChromic® EBT3 films for patient specific IMRT QA using a multichannel approach
- Authors:
- Marrazzo, Livia
Zani, Margherita
Pallotta, Stefania
Arilli, Chiara
Casati, Marta
Compagnucci, Antonella
Talamonti, Cinzia
Bucciolini, Marta - Abstract:
- Highlights: EBT3 were evaluated for pre-treatment patient specific quality assurance. A calibration with 5 dose points was found to be adequate. Results are improved with the introduction of a reference for calibration rescaling. We demonstrated the multichannel approach is superior to the single channel dosimetry. EBT3 films turned out to be a reliable tool for IMRT patient specific QA. Graphical Abstract: Abstract: Purpose: To evaluate EBT3 for pre-treatment patient specific quality assurance (QA). The method we propose combines the experience gained in our center with the guidelines of the protocol proposed by Lewis et al. in 2012. To compare the multichannel approach with the single channel dosimetry. Methods: Gafchromic® EBT3 films were irradiated both at linac and TomoTherapy and calibration curves were obtained. A series of irradiations with simple fields (uniform dose distributions on regular shaped targets) was performed. In a second stage, films were exposed to full clinical plans at linac (step and shoot IMRT and VMAT). At TomoTherapy dose maps were obtained for a clinical plan in three different coronal planes. Films were digitized using an Epson 10000XL scanner and FilmQA™ Pro software was employed for the analysis. Results: The measured calibration curves suggest that, at least for the two beams taken into account (6 MV linac and TomoTherapy), a single calibration can be successfully adopted for each film lot. The application of the multichannel optimizationHighlights: EBT3 were evaluated for pre-treatment patient specific quality assurance. A calibration with 5 dose points was found to be adequate. Results are improved with the introduction of a reference for calibration rescaling. We demonstrated the multichannel approach is superior to the single channel dosimetry. EBT3 films turned out to be a reliable tool for IMRT patient specific QA. Graphical Abstract: Abstract: Purpose: To evaluate EBT3 for pre-treatment patient specific quality assurance (QA). The method we propose combines the experience gained in our center with the guidelines of the protocol proposed by Lewis et al. in 2012. To compare the multichannel approach with the single channel dosimetry. Methods: Gafchromic® EBT3 films were irradiated both at linac and TomoTherapy and calibration curves were obtained. A series of irradiations with simple fields (uniform dose distributions on regular shaped targets) was performed. In a second stage, films were exposed to full clinical plans at linac (step and shoot IMRT and VMAT). At TomoTherapy dose maps were obtained for a clinical plan in three different coronal planes. Films were digitized using an Epson 10000XL scanner and FilmQA™ Pro software was employed for the analysis. Results: The measured calibration curves suggest that, at least for the two beams taken into account (6 MV linac and TomoTherapy), a single calibration can be successfully adopted for each film lot. The application of the multichannel optimization method strongly improves the results in terms of gamma passing rates of the comparison between measured and calculated maps. Conclusions: Up to now EBT films, although attractive, were not preferred for routine patient specific QA due to their complex and time consuming processing and to the challenging work of characterization. The application of the mentioned protocol, together with some additional precautions, and the adoption of the multichannel optimization dosimetry, make this detector a handy and reliable tool for patient specific QA. … (more)
- Is Part Of:
- Physica medica. Volume 31:Issue 8(2015)
- Journal:
- Physica medica
- Issue:
- Volume 31:Issue 8(2015)
- Issue Display:
- Volume 31, Issue 8 (2015)
- Year:
- 2015
- Volume:
- 31
- Issue:
- 8
- Issue Sort Value:
- 2015-0031-0008-0000
- Page Start:
- 1035
- Page End:
- 1042
- Publication Date:
- 2015-12
- Subjects:
- Radiochromic film -- Patient specific quality assurance -- Multichannel dosimetry -- IMRT
Medical physics -- Periodicals
Biophysics -- Periodicals
Biophysics -- Periodicals
Imagerie médicale -- Périodiques
Radiothérapie -- Périodiques
Rayons X -- Sécurité -- Mesures -- Périodiques
Physique -- Périodiques
Médecine -- Périodiques
610.153 - Journal URLs:
- http://www.sciencedirect.com/science/journal/11201797 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/11201797 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/11201797 ↗
http://www.elsevier.com/journals ↗
http://www.physicamedica.com ↗ - DOI:
- 10.1016/j.ejmp.2015.08.010 ↗
- Languages:
- English
- ISSNs:
- 1120-1797
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6475.070000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 7901.xml