Quantification of residual dose estimation error on log file-based patient dose calculation. Issue 5 (May 2016)
- Record Type:
- Journal Article
- Title:
- Quantification of residual dose estimation error on log file-based patient dose calculation. Issue 5 (May 2016)
- Main Title:
- Quantification of residual dose estimation error on log file-based patient dose calculation
- Authors:
- Katsuta, Yoshiyuki
Kadoya, Noriyuki
Fujita, Yukio
Shimizu, Eiji
Matsunaga, Kenichi
Matsushita, Haruo
Majima, Kazuhiro
Jingu, Keiichi - Abstract:
- Highlights: We quantified residual dose estimation error for log file-based method. The residual error cannot be reflected on estimated dose. The determined residual error is 1.6% for target volume and 1.1 Gy for OARs. Abstract: Purpose: The log file-based patient dose estimation includes a residual dose estimation error caused by leaf miscalibration, which cannot be reflected on the estimated dose. The purpose of this study is to determine this residual dose estimation error. Methods and materials: Modified log files for seven head-and-neck and prostate volumetric modulated arc therapy (VMAT) plans simulating leaf miscalibration were generated by shifting both leaf banks (systematic leaf gap errors: ±2.0, ±1.0, and ±0.5 mm in opposite directions and systematic leaf shifts: ±1.0 mm in the same direction) using MATLAB-based (MathWorks, Natick, MA) in-house software. The generated modified and non-modified log files were imported back into the treatment planning system and recalculated. Subsequently, the generalized equivalent uniform dose (gEUD) was quantified for the definition of the planning target volume (PTV) and organs at risks. Results: For MLC leaves calibrated within ±0.5 mm, the quantified residual dose estimation errors that obtained from the slope of the linear regression of gEUD changes between non- and modified log file doses per leaf gap are in head-and-neck plans 1.32 ± 0.27% and 0.82 ± 0.17 Gy for PTV and spinal cord, respectively, and in prostate plansHighlights: We quantified residual dose estimation error for log file-based method. The residual error cannot be reflected on estimated dose. The determined residual error is 1.6% for target volume and 1.1 Gy for OARs. Abstract: Purpose: The log file-based patient dose estimation includes a residual dose estimation error caused by leaf miscalibration, which cannot be reflected on the estimated dose. The purpose of this study is to determine this residual dose estimation error. Methods and materials: Modified log files for seven head-and-neck and prostate volumetric modulated arc therapy (VMAT) plans simulating leaf miscalibration were generated by shifting both leaf banks (systematic leaf gap errors: ±2.0, ±1.0, and ±0.5 mm in opposite directions and systematic leaf shifts: ±1.0 mm in the same direction) using MATLAB-based (MathWorks, Natick, MA) in-house software. The generated modified and non-modified log files were imported back into the treatment planning system and recalculated. Subsequently, the generalized equivalent uniform dose (gEUD) was quantified for the definition of the planning target volume (PTV) and organs at risks. Results: For MLC leaves calibrated within ±0.5 mm, the quantified residual dose estimation errors that obtained from the slope of the linear regression of gEUD changes between non- and modified log file doses per leaf gap are in head-and-neck plans 1.32 ± 0.27% and 0.82 ± 0.17 Gy for PTV and spinal cord, respectively, and in prostate plans 1.22 ± 0.36%, 0.95 ± 0.14 Gy, and 0.45 ± 0.08 Gy for PTV, rectum, and bladder, respectively. Conclusions: In this work, we determine the residual dose estimation errors for VMAT delivery using the log file-based patient dose calculation according to the MLC calibration accuracy. … (more)
- Is Part Of:
- Physica medica. Volume 32:Issue 5(2016)
- Journal:
- Physica medica
- Issue:
- Volume 32:Issue 5(2016)
- Issue Display:
- Volume 32, Issue 5 (2016)
- Year:
- 2016
- Volume:
- 32
- Issue:
- 5
- Issue Sort Value:
- 2016-0032-0005-0000
- Page Start:
- 701
- Page End:
- 705
- Publication Date:
- 2016-05
- Subjects:
- Radiotherapy -- VMAT -- Log file -- Patient QA -- DVH
Medical physics -- Periodicals
Biophysics -- Periodicals
Biophysics -- Periodicals
Imagerie médicale -- Périodiques
Radiothérapie -- Périodiques
Rayons X -- Sécurité -- Mesures -- Périodiques
Physique -- Périodiques
Médecine -- Périodiques
610.153 - Journal URLs:
- http://www.sciencedirect.com/science/journal/11201797 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/11201797 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/11201797 ↗
http://www.elsevier.com/journals ↗
http://www.physicamedica.com ↗ - DOI:
- 10.1016/j.ejmp.2016.04.015 ↗
- Languages:
- English
- ISSNs:
- 1120-1797
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6475.070000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 7851.xml