Evaluation of residual abdominal tumour motion in carbon ion gated treatments through respiratory motion modelling. (February 2017)
- Record Type:
- Journal Article
- Title:
- Evaluation of residual abdominal tumour motion in carbon ion gated treatments through respiratory motion modelling. (February 2017)
- Main Title:
- Evaluation of residual abdominal tumour motion in carbon ion gated treatments through respiratory motion modelling
- Authors:
- Meschini, Giorgia
Seregni, Matteo
Pella, Andrea
Ciocca, Mario
Fossati, Piero
Valvo, Francesca
Riboldi, Marco
Baroni, Guido - Abstract:
- Highlights: A modelling approach is adopted for the analysis of abdominal respiratory motion. The estimated residual motion is within thresholds planned for gated treatments. The gating protocol guarantees the geometrical reproducibility of target position. A dosimetric evaluation of estimated residual motion is now feasible. Abstract: At the Italian National Centre for Oncologic Hadrontherapy (CNAO) patients with upper-abdominal tumours are being treated with carbon ion therapy, adopting the respiratory gating technique in combination with layered rescanning and abdominal compression to mitigate organ motion. Since online imaging of the irradiated volume is not feasible, this study proposes a modelling approach for the estimation of residual motion of the target within the gating window. The model extracts a priori respiratory motion information from the planning 4DCT using deformable image registration (DIR), then combines such information with the external surrogate signal recorded during dose delivery. This provides estimation of a CT volume corresponding to any given respiratory phase measured during treatment. The method was applied for the retrospective estimation of tumour residual motion during irradiation, considering 16 patients treated at CNAO with the respiratory gating protocol. The estimated tumour displacement, calculated with respect to the reference end-exhale position, was always limited (average displacement is 0.32 ± 0.65 mm over all patients) and belowHighlights: A modelling approach is adopted for the analysis of abdominal respiratory motion. The estimated residual motion is within thresholds planned for gated treatments. The gating protocol guarantees the geometrical reproducibility of target position. A dosimetric evaluation of estimated residual motion is now feasible. Abstract: At the Italian National Centre for Oncologic Hadrontherapy (CNAO) patients with upper-abdominal tumours are being treated with carbon ion therapy, adopting the respiratory gating technique in combination with layered rescanning and abdominal compression to mitigate organ motion. Since online imaging of the irradiated volume is not feasible, this study proposes a modelling approach for the estimation of residual motion of the target within the gating window. The model extracts a priori respiratory motion information from the planning 4DCT using deformable image registration (DIR), then combines such information with the external surrogate signal recorded during dose delivery. This provides estimation of a CT volume corresponding to any given respiratory phase measured during treatment. The method was applied for the retrospective estimation of tumour residual motion during irradiation, considering 16 patients treated at CNAO with the respiratory gating protocol. The estimated tumour displacement, calculated with respect to the reference end-exhale position, was always limited (average displacement is 0.32 ± 0.65 mm over all patients) and below the maximum motion defined in the treatment plan. This supports the hypothesis of target position reproducibility, which is the crucial assumption in the gating approach. We also demonstrated the use of the model as a simulation tool to establish a patient-specific relationship between residual motion and the width of the gating window. In conclusion, the implemented method yields an estimation of the repeatability of the internal anatomy configuration during gated treatments, which can be used for further studies concerning the dosimetric impact of the estimated residual organ motion. … (more)
- Is Part Of:
- Physica medica. Volume 34(2017)
- Journal:
- Physica medica
- Issue:
- Volume 34(2017)
- Issue Display:
- Volume 34, Issue 2017 (2017)
- Year:
- 2017
- Volume:
- 34
- Issue:
- 2017
- Issue Sort Value:
- 2017-0034-2017-0000
- Page Start:
- 28
- Page End:
- 37
- Publication Date:
- 2017-02
- Subjects:
- Motion modelling -- 4DCT -- Deformable image registration -- Gating -- Particle therapy
Medical physics -- Periodicals
Biophysics -- Periodicals
Biophysics -- Periodicals
Imagerie médicale -- Périodiques
Radiothérapie -- Périodiques
Rayons X -- Sécurité -- Mesures -- Périodiques
Physique -- Périodiques
Médecine -- Périodiques
610.153 - Journal URLs:
- http://www.sciencedirect.com/science/journal/11201797 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/11201797 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/11201797 ↗
http://www.elsevier.com/journals ↗
http://www.physicamedica.com ↗ - DOI:
- 10.1016/j.ejmp.2017.01.009 ↗
- Languages:
- English
- ISSNs:
- 1120-1797
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6475.070000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 7637.xml